Victor Dias Bergamasco, Vagner Rogério Dos Santos, Bernardo Kaplan Moscovici, Mauro Campos
{"title":"用于矫正无囊支持性无晶体眼的新设备。","authors":"Victor Dias Bergamasco, Vagner Rogério Dos Santos, Bernardo Kaplan Moscovici, Mauro Campos","doi":"10.1080/02713683.2024.2362839","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study describes a prototype developed for aphakia without capsular support (AWCS) and its proof of concept.</p><p><strong>Methods: </strong>This descriptive study used a 3D software to create and analyze virtual prototypes before manufacturing. A nylon-6/nylon-6.6 copolymer filament and a 3D printer were used for prototyping. A device implantation technique was developed using a 23-gauge hypodermic needle. Two opposing markings, 2 mm posterior to the limbus, were made to determine the location of the scleral punctures and the final position of the device. After adequate centralization and positioning of the device, its haptics were cut and cauterized to generate thermal modeling of the extremity and allow the thickening of the tips (flange), serving as an anchoring mechanism to the sclera. The efficacy and adequacy of the technique and device were then evaluated.</p><p><strong>Results: </strong>Vitreous tissue extrusion was not observed during the sclerotomy. The device was well fixed to the sclera; however, adequate IOL stability and centralization still needed to be achieved. The surgeon evaluated the adequacy of all the other devices' characteristics.</p><p><strong>Conclusions: </strong>The development of a technology prototype for correcting AWCS was possible. Although the proposed prototype met most of the established concept guidelines, the stability of the IOL position remains challenging.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2024-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"New Device for the Correction of Aphakia Without Capsular Support.\",\"authors\":\"Victor Dias Bergamasco, Vagner Rogério Dos Santos, Bernardo Kaplan Moscovici, Mauro Campos\",\"doi\":\"10.1080/02713683.2024.2362839\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>This study describes a prototype developed for aphakia without capsular support (AWCS) and its proof of concept.</p><p><strong>Methods: </strong>This descriptive study used a 3D software to create and analyze virtual prototypes before manufacturing. A nylon-6/nylon-6.6 copolymer filament and a 3D printer were used for prototyping. A device implantation technique was developed using a 23-gauge hypodermic needle. Two opposing markings, 2 mm posterior to the limbus, were made to determine the location of the scleral punctures and the final position of the device. After adequate centralization and positioning of the device, its haptics were cut and cauterized to generate thermal modeling of the extremity and allow the thickening of the tips (flange), serving as an anchoring mechanism to the sclera. The efficacy and adequacy of the technique and device were then evaluated.</p><p><strong>Results: </strong>Vitreous tissue extrusion was not observed during the sclerotomy. The device was well fixed to the sclera; however, adequate IOL stability and centralization still needed to be achieved. The surgeon evaluated the adequacy of all the other devices' characteristics.</p><p><strong>Conclusions: </strong>The development of a technology prototype for correcting AWCS was possible. Although the proposed prototype met most of the established concept guidelines, the stability of the IOL position remains challenging.</p>\",\"PeriodicalId\":1,\"journal\":{\"name\":\"Accounts of Chemical Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":16.4000,\"publicationDate\":\"2024-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Accounts of Chemical Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1080/02713683.2024.2362839\",\"RegionNum\":1,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/6/10 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1080/02713683.2024.2362839","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/10 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
New Device for the Correction of Aphakia Without Capsular Support.
Purpose: This study describes a prototype developed for aphakia without capsular support (AWCS) and its proof of concept.
Methods: This descriptive study used a 3D software to create and analyze virtual prototypes before manufacturing. A nylon-6/nylon-6.6 copolymer filament and a 3D printer were used for prototyping. A device implantation technique was developed using a 23-gauge hypodermic needle. Two opposing markings, 2 mm posterior to the limbus, were made to determine the location of the scleral punctures and the final position of the device. After adequate centralization and positioning of the device, its haptics were cut and cauterized to generate thermal modeling of the extremity and allow the thickening of the tips (flange), serving as an anchoring mechanism to the sclera. The efficacy and adequacy of the technique and device were then evaluated.
Results: Vitreous tissue extrusion was not observed during the sclerotomy. The device was well fixed to the sclera; however, adequate IOL stability and centralization still needed to be achieved. The surgeon evaluated the adequacy of all the other devices' characteristics.
Conclusions: The development of a technology prototype for correcting AWCS was possible. Although the proposed prototype met most of the established concept guidelines, the stability of the IOL position remains challenging.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.