Jonas Wörmann, Marc Strik, Stjepan Jurisic, Kara Stout, Mohamed Elrefai, Nina Becher, Beat Schaer, Antonius van Stipdonk, Neil T Srinivasan, Sylvain Ploux, Alexander Breitenstein, Jordana Kron, Paul R Roberts, Tobias Toennis, Dominik Linz, Rajdip Dulai, Julia Hermes-Laufer, Jay Koneru, Ömer Erküner, Sebastian Dittrich, Jan-Hendrik van den Bruck, Jan-Hendrik Schipper, Arian Sultan, Kerstin D Rosenberger, Daniel Steven, Jakob Lüker
{"title":"皮下心律转复除颤器患者大型多中心登记中电池过早耗尽的发生率。","authors":"Jonas Wörmann, Marc Strik, Stjepan Jurisic, Kara Stout, Mohamed Elrefai, Nina Becher, Beat Schaer, Antonius van Stipdonk, Neil T Srinivasan, Sylvain Ploux, Alexander Breitenstein, Jordana Kron, Paul R Roberts, Tobias Toennis, Dominik Linz, Rajdip Dulai, Julia Hermes-Laufer, Jay Koneru, Ömer Erküner, Sebastian Dittrich, Jan-Hendrik van den Bruck, Jan-Hendrik Schipper, Arian Sultan, Kerstin D Rosenberger, Daniel Steven, Jakob Lüker","doi":"10.1093/europace/euae170","DOIUrl":null,"url":null,"abstract":"<p><strong>Aims: </strong>Subcutaneous implantable cardioverter-defibrillators (S-ICDs) have become established in preventing sudden cardiac death, with some advantages over transvenous defibrillator systems, including a lower incidence of lead failures. Despite technological advancements, S-ICD carriers may suffer from significant complications, such as premature battery depletion (PBD), that led to an advisory for nearly 40 000 patients. This multicentre study evaluated the incidence of PBD in a large set of S-ICD patients.</p><p><strong>Methods and results: </strong>Data from patients implanted with S-ICD models A209 and A219 between October 2012 and July 2023 across nine centres in Europe and the USA were reviewed. Incidence and implications of PBD, defined as clinically observed sudden drop in battery longevity, were analysed and compared to PBD with the definition of battery depletion within 60 months. Prospectively collected clinical data were obtained retrospectively from medical records, device telemetry, and manufacturer reports. This registry is listed on ClinicalTrials.gov (NCT05713708). Of the 1112 S-ICD devices analysed, 547 (49.2%) were equipped with a potentially affected capacitor linked to PBD occurrence, currently under Food and Drug Administration advisory. The median follow-up time for all patients was 46 [inter-quartile range (IQR) 24-63] months. Clinically suspected PBD was observed in 159 (29.1%) of cases, with a median time to generator removal or replacement of 65 (IQR 55-72) months, indicative of significant deviations from expected battery lifespan. Manufacturer confirmation of PBD was made in 91.7% of devices returned for analysis. No cases of PBD were observed in devices that were not under advisory.</p><p><strong>Conclusion: </strong>This manufacturer-independent analysis highlights a notable incidence of PBD in patients equipped with S-ICD models under advisory, and the rate of PBD in this study corresponds to the rate currently estimated by the manufacturer. To the best of our knowledge, this provides the largest contemporary peer-reviewed study cohort investigating the actual incidence of PBD in S-ICD patients. These findings emphasize the importance of post-market registries in collaboration between clinicians and the manufacturer to optimize safety and efficacy in S-ICD treatment.</p>","PeriodicalId":11981,"journal":{"name":"Europace","volume":" ","pages":""},"PeriodicalIF":7.9000,"publicationDate":"2024-07-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11218560/pdf/","citationCount":"0","resultStr":"{\"title\":\"Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients.\",\"authors\":\"Jonas Wörmann, Marc Strik, Stjepan Jurisic, Kara Stout, Mohamed Elrefai, Nina Becher, Beat Schaer, Antonius van Stipdonk, Neil T Srinivasan, Sylvain Ploux, Alexander Breitenstein, Jordana Kron, Paul R Roberts, Tobias Toennis, Dominik Linz, Rajdip Dulai, Julia Hermes-Laufer, Jay Koneru, Ömer Erküner, Sebastian Dittrich, Jan-Hendrik van den Bruck, Jan-Hendrik Schipper, Arian Sultan, Kerstin D Rosenberger, Daniel Steven, Jakob Lüker\",\"doi\":\"10.1093/europace/euae170\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Aims: </strong>Subcutaneous implantable cardioverter-defibrillators (S-ICDs) have become established in preventing sudden cardiac death, with some advantages over transvenous defibrillator systems, including a lower incidence of lead failures. Despite technological advancements, S-ICD carriers may suffer from significant complications, such as premature battery depletion (PBD), that led to an advisory for nearly 40 000 patients. This multicentre study evaluated the incidence of PBD in a large set of S-ICD patients.</p><p><strong>Methods and results: </strong>Data from patients implanted with S-ICD models A209 and A219 between October 2012 and July 2023 across nine centres in Europe and the USA were reviewed. Incidence and implications of PBD, defined as clinically observed sudden drop in battery longevity, were analysed and compared to PBD with the definition of battery depletion within 60 months. Prospectively collected clinical data were obtained retrospectively from medical records, device telemetry, and manufacturer reports. This registry is listed on ClinicalTrials.gov (NCT05713708). Of the 1112 S-ICD devices analysed, 547 (49.2%) were equipped with a potentially affected capacitor linked to PBD occurrence, currently under Food and Drug Administration advisory. The median follow-up time for all patients was 46 [inter-quartile range (IQR) 24-63] months. Clinically suspected PBD was observed in 159 (29.1%) of cases, with a median time to generator removal or replacement of 65 (IQR 55-72) months, indicative of significant deviations from expected battery lifespan. Manufacturer confirmation of PBD was made in 91.7% of devices returned for analysis. No cases of PBD were observed in devices that were not under advisory.</p><p><strong>Conclusion: </strong>This manufacturer-independent analysis highlights a notable incidence of PBD in patients equipped with S-ICD models under advisory, and the rate of PBD in this study corresponds to the rate currently estimated by the manufacturer. To the best of our knowledge, this provides the largest contemporary peer-reviewed study cohort investigating the actual incidence of PBD in S-ICD patients. These findings emphasize the importance of post-market registries in collaboration between clinicians and the manufacturer to optimize safety and efficacy in S-ICD treatment.</p>\",\"PeriodicalId\":11981,\"journal\":{\"name\":\"Europace\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":7.9000,\"publicationDate\":\"2024-07-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11218560/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Europace\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/europace/euae170\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Europace","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/europace/euae170","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Occurrence of premature battery depletion in a large multicentre registry of subcutaneous cardioverter-defibrillator patients.
Aims: Subcutaneous implantable cardioverter-defibrillators (S-ICDs) have become established in preventing sudden cardiac death, with some advantages over transvenous defibrillator systems, including a lower incidence of lead failures. Despite technological advancements, S-ICD carriers may suffer from significant complications, such as premature battery depletion (PBD), that led to an advisory for nearly 40 000 patients. This multicentre study evaluated the incidence of PBD in a large set of S-ICD patients.
Methods and results: Data from patients implanted with S-ICD models A209 and A219 between October 2012 and July 2023 across nine centres in Europe and the USA were reviewed. Incidence and implications of PBD, defined as clinically observed sudden drop in battery longevity, were analysed and compared to PBD with the definition of battery depletion within 60 months. Prospectively collected clinical data were obtained retrospectively from medical records, device telemetry, and manufacturer reports. This registry is listed on ClinicalTrials.gov (NCT05713708). Of the 1112 S-ICD devices analysed, 547 (49.2%) were equipped with a potentially affected capacitor linked to PBD occurrence, currently under Food and Drug Administration advisory. The median follow-up time for all patients was 46 [inter-quartile range (IQR) 24-63] months. Clinically suspected PBD was observed in 159 (29.1%) of cases, with a median time to generator removal or replacement of 65 (IQR 55-72) months, indicative of significant deviations from expected battery lifespan. Manufacturer confirmation of PBD was made in 91.7% of devices returned for analysis. No cases of PBD were observed in devices that were not under advisory.
Conclusion: This manufacturer-independent analysis highlights a notable incidence of PBD in patients equipped with S-ICD models under advisory, and the rate of PBD in this study corresponds to the rate currently estimated by the manufacturer. To the best of our knowledge, this provides the largest contemporary peer-reviewed study cohort investigating the actual incidence of PBD in S-ICD patients. These findings emphasize the importance of post-market registries in collaboration between clinicians and the manufacturer to optimize safety and efficacy in S-ICD treatment.
期刊介绍:
EP - Europace - European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology. The journal aims to provide an avenue of communication of top quality European and international original scientific work and reviews in the fields of Arrhythmias, Pacing and Cellular Electrophysiology. The Journal offers the reader a collection of contemporary original peer-reviewed papers, invited papers and editorial comments together with book reviews and correspondence.