Matthias Wetzl, Theresa Heilingbrunner, Felix Heindl, Evelyn Wenkel, Michael Uder, Sabine Ohlmeyer
{"title":"专用乳腺 CT 与乳腺 X 射线照相术对乳腺癌的检测能力取决于乳腺密度。","authors":"Matthias Wetzl, Theresa Heilingbrunner, Felix Heindl, Evelyn Wenkel, Michael Uder, Sabine Ohlmeyer","doi":"10.1097/RLI.0000000000001105","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>To evaluate the detectability of non-contrast-enhanced and contrast-enhanced spiral breast computed tomography ([non]-CE-SBCT) compared with mammography. Secondary objectives are to determine detectability depending on breast density and to evaluate appearance of breast malignancies according to BI-RADS descriptors.</p><p><strong>Methods: </strong>This retrospective institutional review board-approved study included 90 women with 105 biopsy-proven malignant breast lesions. Breast density, BI-RADS descriptors, and detectability were evaluated by 2 independent readers. Diagnostic confidence was rated on a 4-point Likert scale.</p><p><strong>Results: </strong>For readers 1 and 2, detectability was 83.8% and 80.0% for mammography, 99.1% and 99.1% for CE-SBCT ( P < 0.05), and 66.7% and 61.9% for non-CE-SBCT ( P < 0.05). With both readers, detectability in CE-SBCT was high for density A/B/C/D (both 100%/100%/100%/87.5%). Detectability of readers declined with increasing density for mammography (density A = 100%, B = 89.1% and 95.1%, C = 73.1%, D = 50.0% and 71.4%; P < 0.05) and for non-CE-SBCT (density A = 87.5% and 90.7%, B = 65.5% and 69.1%, C = 54.8% and 60.0%, D = 37.5%; P < 0.05). Mass lesions were detected with CT as often as with mammography, whereas architectural distortions and microcalcifications were detected less often with SBCT. Diagnostic confidence was very high or high in 97.2% for CE-SBCT, in 74.1% for non-CE-SBCT, and in 81.4% for mammography.</p><p><strong>Conclusions: </strong>Detectability and diagnostic confidence were very high in CE-SBCT, regardless of breast density. The detectability of non-CE-SBCT was lower than that of mammography and declined with increasing breast density.</p>","PeriodicalId":14486,"journal":{"name":"Investigative Radiology","volume":" ","pages":"861-865"},"PeriodicalIF":7.0000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Detectability of Breast Cancer in Dedicated Breast CT Compared With Mammography Dependent on Breast Density.\",\"authors\":\"Matthias Wetzl, Theresa Heilingbrunner, Felix Heindl, Evelyn Wenkel, Michael Uder, Sabine Ohlmeyer\",\"doi\":\"10.1097/RLI.0000000000001105\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objectives: </strong>To evaluate the detectability of non-contrast-enhanced and contrast-enhanced spiral breast computed tomography ([non]-CE-SBCT) compared with mammography. Secondary objectives are to determine detectability depending on breast density and to evaluate appearance of breast malignancies according to BI-RADS descriptors.</p><p><strong>Methods: </strong>This retrospective institutional review board-approved study included 90 women with 105 biopsy-proven malignant breast lesions. Breast density, BI-RADS descriptors, and detectability were evaluated by 2 independent readers. Diagnostic confidence was rated on a 4-point Likert scale.</p><p><strong>Results: </strong>For readers 1 and 2, detectability was 83.8% and 80.0% for mammography, 99.1% and 99.1% for CE-SBCT ( P < 0.05), and 66.7% and 61.9% for non-CE-SBCT ( P < 0.05). With both readers, detectability in CE-SBCT was high for density A/B/C/D (both 100%/100%/100%/87.5%). Detectability of readers declined with increasing density for mammography (density A = 100%, B = 89.1% and 95.1%, C = 73.1%, D = 50.0% and 71.4%; P < 0.05) and for non-CE-SBCT (density A = 87.5% and 90.7%, B = 65.5% and 69.1%, C = 54.8% and 60.0%, D = 37.5%; P < 0.05). Mass lesions were detected with CT as often as with mammography, whereas architectural distortions and microcalcifications were detected less often with SBCT. Diagnostic confidence was very high or high in 97.2% for CE-SBCT, in 74.1% for non-CE-SBCT, and in 81.4% for mammography.</p><p><strong>Conclusions: </strong>Detectability and diagnostic confidence were very high in CE-SBCT, regardless of breast density. The detectability of non-CE-SBCT was lower than that of mammography and declined with increasing breast density.</p>\",\"PeriodicalId\":14486,\"journal\":{\"name\":\"Investigative Radiology\",\"volume\":\" \",\"pages\":\"861-865\"},\"PeriodicalIF\":7.0000,\"publicationDate\":\"2024-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Investigative Radiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/RLI.0000000000001105\",\"RegionNum\":1,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/7/5 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Investigative Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/RLI.0000000000001105","RegionNum":1,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/5 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Detectability of Breast Cancer in Dedicated Breast CT Compared With Mammography Dependent on Breast Density.
Objectives: To evaluate the detectability of non-contrast-enhanced and contrast-enhanced spiral breast computed tomography ([non]-CE-SBCT) compared with mammography. Secondary objectives are to determine detectability depending on breast density and to evaluate appearance of breast malignancies according to BI-RADS descriptors.
Methods: This retrospective institutional review board-approved study included 90 women with 105 biopsy-proven malignant breast lesions. Breast density, BI-RADS descriptors, and detectability were evaluated by 2 independent readers. Diagnostic confidence was rated on a 4-point Likert scale.
Results: For readers 1 and 2, detectability was 83.8% and 80.0% for mammography, 99.1% and 99.1% for CE-SBCT ( P < 0.05), and 66.7% and 61.9% for non-CE-SBCT ( P < 0.05). With both readers, detectability in CE-SBCT was high for density A/B/C/D (both 100%/100%/100%/87.5%). Detectability of readers declined with increasing density for mammography (density A = 100%, B = 89.1% and 95.1%, C = 73.1%, D = 50.0% and 71.4%; P < 0.05) and for non-CE-SBCT (density A = 87.5% and 90.7%, B = 65.5% and 69.1%, C = 54.8% and 60.0%, D = 37.5%; P < 0.05). Mass lesions were detected with CT as often as with mammography, whereas architectural distortions and microcalcifications were detected less often with SBCT. Diagnostic confidence was very high or high in 97.2% for CE-SBCT, in 74.1% for non-CE-SBCT, and in 81.4% for mammography.
Conclusions: Detectability and diagnostic confidence were very high in CE-SBCT, regardless of breast density. The detectability of non-CE-SBCT was lower than that of mammography and declined with increasing breast density.
期刊介绍:
Investigative Radiology publishes original, peer-reviewed reports on clinical and laboratory investigations in diagnostic imaging, the diagnostic use of radioactive isotopes, computed tomography, positron emission tomography, magnetic resonance imaging, ultrasound, digital subtraction angiography, and related modalities. Emphasis is on early and timely publication. Primarily research-oriented, the journal also includes a wide variety of features of interest to clinical radiologists.