{"title":"使用新型 SIROS D 瓶输送套件进行 90Y SIR-Spheres 活性测量。","authors":"Benjamin P Lopez, S Cheenu Kappadath","doi":"10.2967/jnmt.124.267413","DOIUrl":null,"url":null,"abstract":"<p><p>A new <sup>90</sup>Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate-to-activity conversion factor to assay <sup>90</sup>Y SIR-Spheres activity in the new SIROS kit. <b>Methods:</b> Eight D-vials with initial <sup>90</sup>Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. <b>Results:</b> SIROS D-vial <sup>90</sup>Y activity can be accurately and reliably estimated directly using setting 51 × 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within ±0.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 ± 0.003 GBq/(mR/h) (<i>R</i> <sup>2</sup> = 0.985). <b>Conclusion:</b> Dose calibrator settings and exposure-rate-to-activity conversion factor for <sup>90</sup>Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.</p>","PeriodicalId":16548,"journal":{"name":"Journal of nuclear medicine technology","volume":" ","pages":"256-260"},"PeriodicalIF":1.3000,"publicationDate":"2024-09-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"<sup>90</sup>Y SIR-Spheres Activity Measurement with New SIROS D-Vial Delivery Kit.\",\"authors\":\"Benjamin P Lopez, S Cheenu Kappadath\",\"doi\":\"10.2967/jnmt.124.267413\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A new <sup>90</sup>Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate-to-activity conversion factor to assay <sup>90</sup>Y SIR-Spheres activity in the new SIROS kit. <b>Methods:</b> Eight D-vials with initial <sup>90</sup>Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. <b>Results:</b> SIROS D-vial <sup>90</sup>Y activity can be accurately and reliably estimated directly using setting 51 × 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within ±0.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 ± 0.003 GBq/(mR/h) (<i>R</i> <sup>2</sup> = 0.985). <b>Conclusion:</b> Dose calibrator settings and exposure-rate-to-activity conversion factor for <sup>90</sup>Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.</p>\",\"PeriodicalId\":16548,\"journal\":{\"name\":\"Journal of nuclear medicine technology\",\"volume\":\" \",\"pages\":\"256-260\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-09-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of nuclear medicine technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2967/jnmt.124.267413\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of nuclear medicine technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2967/jnmt.124.267413","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

新推出的 90Y SIR-Spheres 运载试剂盒(SIROS D 瓶和防护罩)采用了不同于传统 V 瓶试剂盒的物理形式。在此,我们确定了剂量校准器的设置和曝光率与活性的转换系数,以便在新的 SIROS 套件中检测 90Y SIR-Spheres 活性。方法:使用剂量校准器和曝光率测量仪对丙烯酸防护罩内初始 90Y 放射性活度为 1.2 至 6.6 GBq 的 8 个 D-椭球体进行检测,直到放射性活度衰减到大约 0.1 GBq。确定了能使放射性活度误差中位数最小的剂量校准器设置,以及辐照速率与放射性活度的最佳拟合斜率。结果:使用 CRC-15R 和 CRC-55tR 剂量校准器上的 51 × 10 设置直接估算 SIROS D-vial 90Y 放射性活度(误差在 ±0.5% 以内),使用转换系数 0.664 ± 0.003 GBq/(mR/h) 间接估算 30 厘米处的暴露率读数(R 2 = 0.985)。结论使用新的 SIROS 套件进行 90Y 活性测定时,剂量校准器的设置和曝光率与活性的换算系数应根据原有的 V-Vial 参数进行更新,以避免约 10%的低估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
90Y SIR-Spheres Activity Measurement with New SIROS D-Vial Delivery Kit.

A new 90Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate-to-activity conversion factor to assay 90Y SIR-Spheres activity in the new SIROS kit. Methods: Eight D-vials with initial 90Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. Results: SIROS D-vial 90Y activity can be accurately and reliably estimated directly using setting 51 × 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within ±0.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 ± 0.003 GBq/(mR/h) (R 2 = 0.985). Conclusion: Dose calibrator settings and exposure-rate-to-activity conversion factor for 90Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of nuclear medicine technology
Journal of nuclear medicine technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.90
自引率
15.40%
发文量
57
期刊最新文献
Enhancing USP <825> Compliance in Transarterial Radioembolization Using 90Y Resin Microspheres. Evaluation of AI-Driven Automatic Segmentation Algorithms for Lung Shunt Fraction Estimation in 90Y-Labeled Microsphere Therapy. Cold Carbonated Water Reduces Interference from Extracardiac Gastric Activity During 82Rb PET in Patients. Follicular Variant of Papillary Thyroid Carcinoma with Extensive Tumor Venous Thrombosis and Spectrum of Unusual Metastases: Initial Radioiodine Response Followed by TENIS with NRAS Mutation. Practical Considerations for Prostate Cancer Imaging and Radiopharmaceutical Therapy.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1