使用新型 SIROS D 瓶输送套件进行 90Y SIR-Spheres 活性测量。

IF 1 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Journal of nuclear medicine technology Pub Date : 2024-09-05 DOI:10.2967/jnmt.124.267413
Benjamin P Lopez, S Cheenu Kappadath
{"title":"使用新型 SIROS D 瓶输送套件进行 90Y SIR-Spheres 活性测量。","authors":"Benjamin P Lopez, S Cheenu Kappadath","doi":"10.2967/jnmt.124.267413","DOIUrl":null,"url":null,"abstract":"<p><p>A new <sup>90</sup>Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate-to-activity conversion factor to assay <sup>90</sup>Y SIR-Spheres activity in the new SIROS kit. <b>Methods:</b> Eight D-vials with initial <sup>90</sup>Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. <b>Results:</b> SIROS D-vial <sup>90</sup>Y activity can be accurately and reliably estimated directly using setting 51 × 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within ±0.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 ± 0.003 GBq/(mR/h) (<i>R</i> <sup>2</sup> = 0.985). <b>Conclusion:</b> Dose calibrator settings and exposure-rate-to-activity conversion factor for <sup>90</sup>Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.</p>","PeriodicalId":16548,"journal":{"name":"Journal of nuclear medicine technology","volume":" ","pages":"256-260"},"PeriodicalIF":1.0000,"publicationDate":"2024-09-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"<sup>90</sup>Y SIR-Spheres Activity Measurement with New SIROS D-Vial Delivery Kit.\",\"authors\":\"Benjamin P Lopez, S Cheenu Kappadath\",\"doi\":\"10.2967/jnmt.124.267413\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A new <sup>90</sup>Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate-to-activity conversion factor to assay <sup>90</sup>Y SIR-Spheres activity in the new SIROS kit. <b>Methods:</b> Eight D-vials with initial <sup>90</sup>Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. <b>Results:</b> SIROS D-vial <sup>90</sup>Y activity can be accurately and reliably estimated directly using setting 51 × 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within ±0.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 ± 0.003 GBq/(mR/h) (<i>R</i> <sup>2</sup> = 0.985). <b>Conclusion:</b> Dose calibrator settings and exposure-rate-to-activity conversion factor for <sup>90</sup>Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.</p>\",\"PeriodicalId\":16548,\"journal\":{\"name\":\"Journal of nuclear medicine technology\",\"volume\":\" \",\"pages\":\"256-260\"},\"PeriodicalIF\":1.0000,\"publicationDate\":\"2024-09-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of nuclear medicine technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2967/jnmt.124.267413\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of nuclear medicine technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2967/jnmt.124.267413","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

新推出的 90Y SIR-Spheres 运载试剂盒(SIROS D 瓶和防护罩)采用了不同于传统 V 瓶试剂盒的物理形式。在此,我们确定了剂量校准器的设置和曝光率与活性的转换系数,以便在新的 SIROS 套件中检测 90Y SIR-Spheres 活性。方法:使用剂量校准器和曝光率测量仪对丙烯酸防护罩内初始 90Y 放射性活度为 1.2 至 6.6 GBq 的 8 个 D-椭球体进行检测,直到放射性活度衰减到大约 0.1 GBq。确定了能使放射性活度误差中位数最小的剂量校准器设置,以及辐照速率与放射性活度的最佳拟合斜率。结果:使用 CRC-15R 和 CRC-55tR 剂量校准器上的 51 × 10 设置直接估算 SIROS D-vial 90Y 放射性活度(误差在 ±0.5% 以内),使用转换系数 0.664 ± 0.003 GBq/(mR/h) 间接估算 30 厘米处的暴露率读数(R 2 = 0.985)。结论使用新的 SIROS 套件进行 90Y 活性测定时,剂量校准器的设置和曝光率与活性的换算系数应根据原有的 V-Vial 参数进行更新,以避免约 10%的低估。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
90Y SIR-Spheres Activity Measurement with New SIROS D-Vial Delivery Kit.

A new 90Y SIR-Spheres delivery kit (SIROS D-vial and shield) has been introduced with a different physical form from the legacy V-Vial kit. Here, we establish the dose calibrator settings and exposure-rate-to-activity conversion factor to assay 90Y SIR-Spheres activity in the new SIROS kit. Methods: Eight D-vials with initial 90Y activities from 1.2 to 6.6 GBq within acrylic shields were assayed with dose calibrators and exposure-rate meters until activities decayed to approximately 0.1 GBq. The dose calibrator settings resulting in the lowest median activity errors and the best-fit slope of exposure rate versus activity were identified. Results: SIROS D-vial 90Y activity can be accurately and reliably estimated directly using setting 51 × 10 on both the CRC-15R and the CRC-55tR dose calibrators (errors within ±0.5%) and indirectly with an exposure-rate reading at 30 cm using conversion factor 0.664 ± 0.003 GBq/(mR/h) (R 2 = 0.985). Conclusion: Dose calibrator settings and exposure-rate-to-activity conversion factor for 90Y activity assays with new SIROS kit should be updated from legacy V-Vial parameters to avoid an approximately 10% underestimation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of nuclear medicine technology
Journal of nuclear medicine technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.90
自引率
15.40%
发文量
57
期刊最新文献
Amyloid Imaging Update: How the Amyloid Landscape Is Changing in Light of the Recent Food and Drug Administration Approval of Antiamyloid Therapeutics. Brain Imaging-PET: Tau. Decoding the Jargon: Understanding the Nomenclature of Clinical Education. Delivery Methods of Radiopharmaceuticals: Exploring Global Strategies to Minimize Occupational Radiation Exposure. SPECT Views for Cardiac Amyloidosis Imaging.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1