{"title":"Magtrace 的存在是否排除了在 MR-Linac 上进行自适应乳腺放射治疗的可能性?","authors":"","doi":"10.1016/j.jmir.2024.101716","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>This work reports on an unusual finding observed during image quality assessment in the preparation for the clinical implementation of breast magnetic resonance image-guided radiotherapy (MRIgRT) on a 1.5 Tesla (T) magnetic resonance linear accelerator (MR-Linac) (Elekta AB, Stockholm, Sweden).</p></div><div><h3>Case and outcomes</h3><p>A patient with T2 N0 M0 right breast invasive ductal carcinoma, receiving adjuvant radiotherapy, underwent two imaging sessions on the MR-Linac. The imaging protocol included T1- and T2-weighted (W) turbo spin echo (TSE) sequences, a T1W mDixon, and a T2W TSE navigated sequence acquired on end-expiration. All images were reconstructed in the axial plane.</p><p>Images were assessed for image quality and appropriateness for use within the treatment pathway using visual grading analysis (VGA). An artefact in the right breast was noted independently by all observers. The patient's skin and medical notes were reviewed for possible explanation. The findings were discussed with the patient's responsible clinician, and subsequent referral to the local multi-disciplinary team (MDT) for radiologist review was made.</p><p>On further investigation, the patient's images demonstrated a signal void in the subareolar region of the right breast coinciding with the surgical site. This was distal from the tumour bed and deemed unlikely to be related to a Magseed marker or intraoperative clips. The patient reported no history of nipple tattoo or piercing. There was nothing on clothing that this could be attributed to.</p></div><div><h3>Discussion</h3><p>Following MDT review, where all potential sources of signal void were considered, it was concluded that the cause was Magtrace, a superparamagnetic iron oxide tracer, recommended for sentinel lymph node localisation in patients with breast cancer in the United Kingdom. The artefact was characteristic of a magnetic susceptibility artefact. These can arise from local magnetic field inhomogeneities caused by the presence of the metal compounds in MagTrace. For breast MRIgRT on the MR-Linac, treatment verification and the possibility of real-time replanning is a critical aspect. The magnetic susceptibility artefact significantly inhibited plan adaption and confidence in the online image registration process making the patient ineligible for treatment on the MR-Linac.</p></div><div><h3>Conclusion</h3><p>As part of ongoing work-up for breast MRIgRT, the screening of patients for Magtrace is now included. Optimisation of MR imaging sequences for radiotherapy planning and image review to minimise distortion are being developed.</p></div>","PeriodicalId":46420,"journal":{"name":"Journal of Medical Imaging and Radiation Sciences","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-07-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Does the presence of Magtrace preclude adaptive breast radiotherapy on an MR-Linac?\",\"authors\":\"\",\"doi\":\"10.1016/j.jmir.2024.101716\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>This work reports on an unusual finding observed during image quality assessment in the preparation for the clinical implementation of breast magnetic resonance image-guided radiotherapy (MRIgRT) on a 1.5 Tesla (T) magnetic resonance linear accelerator (MR-Linac) (Elekta AB, Stockholm, Sweden).</p></div><div><h3>Case and outcomes</h3><p>A patient with T2 N0 M0 right breast invasive ductal carcinoma, receiving adjuvant radiotherapy, underwent two imaging sessions on the MR-Linac. The imaging protocol included T1- and T2-weighted (W) turbo spin echo (TSE) sequences, a T1W mDixon, and a T2W TSE navigated sequence acquired on end-expiration. All images were reconstructed in the axial plane.</p><p>Images were assessed for image quality and appropriateness for use within the treatment pathway using visual grading analysis (VGA). An artefact in the right breast was noted independently by all observers. The patient's skin and medical notes were reviewed for possible explanation. The findings were discussed with the patient's responsible clinician, and subsequent referral to the local multi-disciplinary team (MDT) for radiologist review was made.</p><p>On further investigation, the patient's images demonstrated a signal void in the subareolar region of the right breast coinciding with the surgical site. This was distal from the tumour bed and deemed unlikely to be related to a Magseed marker or intraoperative clips. The patient reported no history of nipple tattoo or piercing. There was nothing on clothing that this could be attributed to.</p></div><div><h3>Discussion</h3><p>Following MDT review, where all potential sources of signal void were considered, it was concluded that the cause was Magtrace, a superparamagnetic iron oxide tracer, recommended for sentinel lymph node localisation in patients with breast cancer in the United Kingdom. The artefact was characteristic of a magnetic susceptibility artefact. These can arise from local magnetic field inhomogeneities caused by the presence of the metal compounds in MagTrace. For breast MRIgRT on the MR-Linac, treatment verification and the possibility of real-time replanning is a critical aspect. The magnetic susceptibility artefact significantly inhibited plan adaption and confidence in the online image registration process making the patient ineligible for treatment on the MR-Linac.</p></div><div><h3>Conclusion</h3><p>As part of ongoing work-up for breast MRIgRT, the screening of patients for Magtrace is now included. Optimisation of MR imaging sequences for radiotherapy planning and image review to minimise distortion are being developed.</p></div>\",\"PeriodicalId\":46420,\"journal\":{\"name\":\"Journal of Medical Imaging and Radiation Sciences\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-07-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Imaging and Radiation Sciences\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1939865424004478\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Imaging and Radiation Sciences","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1939865424004478","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

导读:本文报告了在 1.5 特斯拉(T)磁共振直线加速器(MR-Linac)(瑞典斯德哥尔摩 Elekta AB 公司)上准备临床实施乳腺磁共振图像引导放疗(MRIgRT)的图像质量评估过程中观察到的异常发现。成像方案包括 T1 和 T2 加权(W)涡轮自旋回波(TSE)序列、T1W mDixon 和在呼气末采集的 T2W TSE 导航序列。所有图像均在轴向平面进行重建。采用视觉分级分析(VGA)评估图像质量和是否适合在治疗路径中使用。所有观察者均注意到右侧乳房存在伪影。他们查看了患者的皮肤和医疗记录,以寻找可能的解释。在进一步检查中,患者的图像显示右侧乳房乳晕下区域与手术部位重合处出现信号缺失。这离肿瘤床较远,被认为不太可能与 Magseed 标记或术中夹片有关。患者称没有乳头纹身或穿孔史。讨论经过多学科专家小组(MDT)的审查,考虑了信号无效的所有潜在来源,得出的结论是Magtrace是一种超顺磁性氧化铁示踪剂,推荐用于英国乳腺癌患者的前哨淋巴结定位。这种伪影具有磁感应伪影的特征。这可能是 MagTrace 中的金属化合物造成的局部磁场不均匀引起的。对于 MR-Linac 上的乳腺 MRIgRT 来说,治疗验证和实时重新扫描的可能性是至关重要的。磁感应伪影严重影响了计划的适应性和对在线图像注册过程的信心,使患者不符合在 MR-Linac 上进行治疗的条件。目前正在开发用于放疗计划和图像审查的磁共振成像序列优化技术,以最大限度地减少失真。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Does the presence of Magtrace preclude adaptive breast radiotherapy on an MR-Linac?

Introduction

This work reports on an unusual finding observed during image quality assessment in the preparation for the clinical implementation of breast magnetic resonance image-guided radiotherapy (MRIgRT) on a 1.5 Tesla (T) magnetic resonance linear accelerator (MR-Linac) (Elekta AB, Stockholm, Sweden).

Case and outcomes

A patient with T2 N0 M0 right breast invasive ductal carcinoma, receiving adjuvant radiotherapy, underwent two imaging sessions on the MR-Linac. The imaging protocol included T1- and T2-weighted (W) turbo spin echo (TSE) sequences, a T1W mDixon, and a T2W TSE navigated sequence acquired on end-expiration. All images were reconstructed in the axial plane.

Images were assessed for image quality and appropriateness for use within the treatment pathway using visual grading analysis (VGA). An artefact in the right breast was noted independently by all observers. The patient's skin and medical notes were reviewed for possible explanation. The findings were discussed with the patient's responsible clinician, and subsequent referral to the local multi-disciplinary team (MDT) for radiologist review was made.

On further investigation, the patient's images demonstrated a signal void in the subareolar region of the right breast coinciding with the surgical site. This was distal from the tumour bed and deemed unlikely to be related to a Magseed marker or intraoperative clips. The patient reported no history of nipple tattoo or piercing. There was nothing on clothing that this could be attributed to.

Discussion

Following MDT review, where all potential sources of signal void were considered, it was concluded that the cause was Magtrace, a superparamagnetic iron oxide tracer, recommended for sentinel lymph node localisation in patients with breast cancer in the United Kingdom. The artefact was characteristic of a magnetic susceptibility artefact. These can arise from local magnetic field inhomogeneities caused by the presence of the metal compounds in MagTrace. For breast MRIgRT on the MR-Linac, treatment verification and the possibility of real-time replanning is a critical aspect. The magnetic susceptibility artefact significantly inhibited plan adaption and confidence in the online image registration process making the patient ineligible for treatment on the MR-Linac.

Conclusion

As part of ongoing work-up for breast MRIgRT, the screening of patients for Magtrace is now included. Optimisation of MR imaging sequences for radiotherapy planning and image review to minimise distortion are being developed.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Imaging and Radiation Sciences
Journal of Medical Imaging and Radiation Sciences RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
2.30
自引率
11.10%
发文量
231
审稿时长
53 days
期刊介绍: Journal of Medical Imaging and Radiation Sciences is the official peer-reviewed journal of the Canadian Association of Medical Radiation Technologists. This journal is published four times a year and is circulated to approximately 11,000 medical radiation technologists, libraries and radiology departments throughout Canada, the United States and overseas. The Journal publishes articles on recent research, new technology and techniques, professional practices, technologists viewpoints as well as relevant book reviews.
期刊最新文献
A bibliometric analysis on research authorship and collaboration patterns in radiography professional journals: A 10-year review The design and construct a website for collection and report diagnostic reference levels (DRLs) in diagnostic radiography Beyond the Machine: A Radiographer's Tale Innovative Learning Activities to Prepare Radiography Students for Final-Year Clinical Placements: An Educational Perspective Enhancing radiation therapy student's cultural competency and safety of Canadian Indigenous populations using cultural immersion
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1