从足球头部运动学角度对口罩和头盔传感器数据进行现场评估。

IF 3 2区 医学 Q3 ENGINEERING, BIOMEDICAL Annals of Biomedical Engineering Pub Date : 2024-07-25 DOI:10.1007/s10439-024-03583-0
Ty D. Holcomb, Madison E. Marks, N. Stewart Pritchard, Logan E. Miller, Steve Rowson, Garrett S. Bullock, Jillian E. Urban, Joel D. Stitzel
{"title":"从足球头部运动学角度对口罩和头盔传感器数据进行现场评估。","authors":"Ty D. Holcomb,&nbsp;Madison E. Marks,&nbsp;N. Stewart Pritchard,&nbsp;Logan E. Miller,&nbsp;Steve Rowson,&nbsp;Garrett S. Bullock,&nbsp;Jillian E. Urban,&nbsp;Joel D. Stitzel","doi":"10.1007/s10439-024-03583-0","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><p>Wearable sensors are used to measure head impact exposure in sports. The Head Impact Telemetry (HIT) System is a helmet-mounted system that has been commonly utilized to measure head impacts in American football. Advancements in sensor technology have fueled the development of alternative sensor methods such as instrumented mouthguards. The objective of this study was to compare peak magnitude measured from high school football athletes dually instrumented with the HIT System and a mouthpiece-based sensor system.</p><h3>Methods</h3><p>Data was collected at all contact practices and competitions over a single season of spring football. Recorded events were observed and identified on video and paired using event timestamps. Paired events were further stratified by removing mouthpiece events with peak resultant linear acceleration below 10 g and events with contact to the facemask or body of athletes.</p><h3>Results</h3><p>A total of 133 paired events were analyzed in the results. There was a median difference (mouthpiece subtracted from HIT System) in peak resultant linear and rotational acceleration for concurrently measured events of 7.3 g and 189 rad/s<sup>2</sup>. Greater magnitude events resulted in larger kinematic differences between sensors and a Bland Altman analysis found a mean bias of 8.8 g and 104 rad/s<sup>2</sup>, respectively.</p><h3>Conclusion</h3><p>If the mouthpiece-based sensor is considered close to truth, the results of this study are consistent with previous HIT System validation studies indicating low error on average but high scatter across individual events. Future researchers should be mindful of sensor limitations when comparing results collected using varying sensor technologies.</p></div>","PeriodicalId":7986,"journal":{"name":"Annals of Biomedical Engineering","volume":null,"pages":null},"PeriodicalIF":3.0000,"publicationDate":"2024-07-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11402845/pdf/","citationCount":"0","resultStr":"{\"title\":\"On-Field Evaluation of Mouthpiece-and-Helmet-Mounted Sensor Data from Head Kinematics in Football\",\"authors\":\"Ty D. Holcomb,&nbsp;Madison E. Marks,&nbsp;N. Stewart Pritchard,&nbsp;Logan E. Miller,&nbsp;Steve Rowson,&nbsp;Garrett S. Bullock,&nbsp;Jillian E. Urban,&nbsp;Joel D. Stitzel\",\"doi\":\"10.1007/s10439-024-03583-0\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><p>Wearable sensors are used to measure head impact exposure in sports. The Head Impact Telemetry (HIT) System is a helmet-mounted system that has been commonly utilized to measure head impacts in American football. Advancements in sensor technology have fueled the development of alternative sensor methods such as instrumented mouthguards. The objective of this study was to compare peak magnitude measured from high school football athletes dually instrumented with the HIT System and a mouthpiece-based sensor system.</p><h3>Methods</h3><p>Data was collected at all contact practices and competitions over a single season of spring football. Recorded events were observed and identified on video and paired using event timestamps. Paired events were further stratified by removing mouthpiece events with peak resultant linear acceleration below 10 g and events with contact to the facemask or body of athletes.</p><h3>Results</h3><p>A total of 133 paired events were analyzed in the results. There was a median difference (mouthpiece subtracted from HIT System) in peak resultant linear and rotational acceleration for concurrently measured events of 7.3 g and 189 rad/s<sup>2</sup>. Greater magnitude events resulted in larger kinematic differences between sensors and a Bland Altman analysis found a mean bias of 8.8 g and 104 rad/s<sup>2</sup>, respectively.</p><h3>Conclusion</h3><p>If the mouthpiece-based sensor is considered close to truth, the results of this study are consistent with previous HIT System validation studies indicating low error on average but high scatter across individual events. Future researchers should be mindful of sensor limitations when comparing results collected using varying sensor technologies.</p></div>\",\"PeriodicalId\":7986,\"journal\":{\"name\":\"Annals of Biomedical Engineering\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":3.0000,\"publicationDate\":\"2024-07-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11402845/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Annals of Biomedical Engineering\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://link.springer.com/article/10.1007/s10439-024-03583-0\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Annals of Biomedical Engineering","FirstCategoryId":"5","ListUrlMain":"https://link.springer.com/article/10.1007/s10439-024-03583-0","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

目的:可穿戴传感器用于测量运动中头部受到的冲击。头部撞击遥测(HIT)系统是一种安装在头盔上的系统,通常用于测量美式橄榄球运动中的头部撞击。传感器技术的进步推动了其他传感器方法的发展,例如带仪器的护齿。本研究的目的是比较使用 HIT 系统和口罩式传感器系统测量的高中橄榄球运动员的峰值幅度:方法:在春季橄榄球赛的一个赛季中,在所有接触性训练和比赛中收集数据。通过视频观察和识别记录的事件,并使用事件时间戳进行配对。通过剔除峰值线性加速度低于 10 g 的口罩事件和运动员面罩或身体接触事件,对配对事件进行进一步分层:结果:共分析了 133 个配对事件。在同时测量的事件中,线性加速度和旋转加速度峰值的中位数差异(口罩减去 HIT 系统)分别为 7.3 g 和 189 rad/s2。幅度更大的事件导致传感器之间的运动学差异更大,布兰德-阿尔特曼分析发现平均偏差分别为 8.8 克和 104 拉德/秒2:如果将口罩式传感器视为接近真实的传感器,那么本研究的结果与之前的 HIT 系统验证研究一致,表明平均误差较低,但单个事件之间的误差较大。未来的研究人员在比较使用不同传感器技术收集的结果时,应注意传感器的局限性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
On-Field Evaluation of Mouthpiece-and-Helmet-Mounted Sensor Data from Head Kinematics in Football

Purpose

Wearable sensors are used to measure head impact exposure in sports. The Head Impact Telemetry (HIT) System is a helmet-mounted system that has been commonly utilized to measure head impacts in American football. Advancements in sensor technology have fueled the development of alternative sensor methods such as instrumented mouthguards. The objective of this study was to compare peak magnitude measured from high school football athletes dually instrumented with the HIT System and a mouthpiece-based sensor system.

Methods

Data was collected at all contact practices and competitions over a single season of spring football. Recorded events were observed and identified on video and paired using event timestamps. Paired events were further stratified by removing mouthpiece events with peak resultant linear acceleration below 10 g and events with contact to the facemask or body of athletes.

Results

A total of 133 paired events were analyzed in the results. There was a median difference (mouthpiece subtracted from HIT System) in peak resultant linear and rotational acceleration for concurrently measured events of 7.3 g and 189 rad/s2. Greater magnitude events resulted in larger kinematic differences between sensors and a Bland Altman analysis found a mean bias of 8.8 g and 104 rad/s2, respectively.

Conclusion

If the mouthpiece-based sensor is considered close to truth, the results of this study are consistent with previous HIT System validation studies indicating low error on average but high scatter across individual events. Future researchers should be mindful of sensor limitations when comparing results collected using varying sensor technologies.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Annals of Biomedical Engineering
Annals of Biomedical Engineering 工程技术-工程:生物医学
CiteScore
7.50
自引率
15.80%
发文量
212
审稿时长
3 months
期刊介绍: Annals of Biomedical Engineering is an official journal of the Biomedical Engineering Society, publishing original articles in the major fields of bioengineering and biomedical engineering. The Annals is an interdisciplinary and international journal with the aim to highlight integrated approaches to the solutions of biological and biomedical problems.
期刊最新文献
The Potential of Sound Analysis to Reveal Hemodynamic Conditions of Arteriovenous Fistulae for Hemodialysis. Active Machine Learning for Pre-procedural Prediction of Time-Varying Boundary Condition After Fontan Procedure Using Generative Adversarial Networks. Simulating Pelvis Kinematics from Belt and Seat Loading in Frontal Car Crash Scenarios: Important Boundary Conditions that Influence the Outcome. Lumbar Spine Orientation Affects Compressive Fracture Outcome. An In vivo Pilot Study to Estimate the Swelling of the Aneurysm Wall Rabbit Model Generated with Pulsed Fluid Against the Aneurysm Wall.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1