{"title":"系统设计参数对广角数字乳腺断层合成中微小钙化检测的影响的模拟研究。","authors":"Xiaoyu Duan, Hailiang Huang, Wei Zhao","doi":"10.1117/1.JMI.12.S1.S13002","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Accurate detection of microcalcifications ( <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> ) is crucial for the early detection of breast cancer. Some clinical studies have indicated that digital breast tomosynthesis (DBT) systems with a wide angular range have inferior <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability compared with those with a narrow angular range. This study aims to (1) provide guidance for optimizing wide-angle (WA) DBT for improving <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> detectability and (2) prioritize key optimization factors.</p><p><strong>Approach: </strong>An <i>in-silico</i> DBT pipeline was constructed to evaluate <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability of a WA DBT system under various imaging conditions: focal spot motion (FSM), angular dose distribution (ADS), detector pixel pitch, and detector electronic noise (EN). Images were simulated using a digital anthropomorphic breast phantom inserted with <math><mrow><mn>120</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> clusters. Evaluation metrics included the signal-to-noise ratio (SNR) of the filtered channel observer and the area under the receiver operator curve (AUC) of multiple-reader multiple-case analysis.</p><p><strong>Results: </strong>Results showed that FSM degraded <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> sharpness and decreased the SNR and AUC by 5.2% and 1.8%, respectively. Non-uniform ADS increased the SNR by 62.8% and the AUC by 10.2% for filtered backprojection reconstruction with a typical clinical filter setting. When EN decreased from 2000 to 200 electrons, the SNR and AUC increased by 21.6% and 5.0%, respectively. Decreasing the detector pixel pitch from 85 to <math><mrow><mn>50</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> improved the SNR and AUC by 55.6% and 7.5%, respectively. The combined improvement of a <math><mrow><mn>50</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> pixel pitch and EN200 was 89.2% in the SNR and 12.8% in the AUC.</p><p><strong>Conclusions: </strong>Based on the magnitude of impact, the priority for enhancing <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability in WA DBT is as follows: (1) utilizing detectors with a small pixel pitch and low EN level, (2) allocating a higher dose to central projections, and (3) reducing FSM. The results from this study can potentially provide guidance for DBT system optimization in the future.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11266813/pdf/","citationCount":"0","resultStr":"{\"title\":\"<i>In-silico</i> study of the impact of system design parameters on microcalcification detection in wide-angle digital breast tomosynthesis.\",\"authors\":\"Xiaoyu Duan, Hailiang Huang, Wei Zhao\",\"doi\":\"10.1117/1.JMI.12.S1.S13002\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Accurate detection of microcalcifications ( <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> ) is crucial for the early detection of breast cancer. Some clinical studies have indicated that digital breast tomosynthesis (DBT) systems with a wide angular range have inferior <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability compared with those with a narrow angular range. This study aims to (1) provide guidance for optimizing wide-angle (WA) DBT for improving <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> detectability and (2) prioritize key optimization factors.</p><p><strong>Approach: </strong>An <i>in-silico</i> DBT pipeline was constructed to evaluate <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability of a WA DBT system under various imaging conditions: focal spot motion (FSM), angular dose distribution (ADS), detector pixel pitch, and detector electronic noise (EN). Images were simulated using a digital anthropomorphic breast phantom inserted with <math><mrow><mn>120</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> clusters. Evaluation metrics included the signal-to-noise ratio (SNR) of the filtered channel observer and the area under the receiver operator curve (AUC) of multiple-reader multiple-case analysis.</p><p><strong>Results: </strong>Results showed that FSM degraded <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> sharpness and decreased the SNR and AUC by 5.2% and 1.8%, respectively. Non-uniform ADS increased the SNR by 62.8% and the AUC by 10.2% for filtered backprojection reconstruction with a typical clinical filter setting. When EN decreased from 2000 to 200 electrons, the SNR and AUC increased by 21.6% and 5.0%, respectively. Decreasing the detector pixel pitch from 85 to <math><mrow><mn>50</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> improved the SNR and AUC by 55.6% and 7.5%, respectively. The combined improvement of a <math><mrow><mn>50</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> pixel pitch and EN200 was 89.2% in the SNR and 12.8% in the AUC.</p><p><strong>Conclusions: </strong>Based on the magnitude of impact, the priority for enhancing <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability in WA DBT is as follows: (1) utilizing detectors with a small pixel pitch and low EN level, (2) allocating a higher dose to central projections, and (3) reducing FSM. The results from this study can potentially provide guidance for DBT system optimization in the future.</p>\",\"PeriodicalId\":1,\"journal\":{\"name\":\"Accounts of Chemical Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":16.4000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11266813/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Accounts of Chemical Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1117/1.JMI.12.S1.S13002\",\"RegionNum\":1,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/7/24 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1117/1.JMI.12.S1.S13002","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/24 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
In-silico study of the impact of system design parameters on microcalcification detection in wide-angle digital breast tomosynthesis.
Purpose: Accurate detection of microcalcifications ( ) is crucial for the early detection of breast cancer. Some clinical studies have indicated that digital breast tomosynthesis (DBT) systems with a wide angular range have inferior detectability compared with those with a narrow angular range. This study aims to (1) provide guidance for optimizing wide-angle (WA) DBT for improving detectability and (2) prioritize key optimization factors.
Approach: An in-silico DBT pipeline was constructed to evaluate detectability of a WA DBT system under various imaging conditions: focal spot motion (FSM), angular dose distribution (ADS), detector pixel pitch, and detector electronic noise (EN). Images were simulated using a digital anthropomorphic breast phantom inserted with clusters. Evaluation metrics included the signal-to-noise ratio (SNR) of the filtered channel observer and the area under the receiver operator curve (AUC) of multiple-reader multiple-case analysis.
Results: Results showed that FSM degraded sharpness and decreased the SNR and AUC by 5.2% and 1.8%, respectively. Non-uniform ADS increased the SNR by 62.8% and the AUC by 10.2% for filtered backprojection reconstruction with a typical clinical filter setting. When EN decreased from 2000 to 200 electrons, the SNR and AUC increased by 21.6% and 5.0%, respectively. Decreasing the detector pixel pitch from 85 to improved the SNR and AUC by 55.6% and 7.5%, respectively. The combined improvement of a pixel pitch and EN200 was 89.2% in the SNR and 12.8% in the AUC.
Conclusions: Based on the magnitude of impact, the priority for enhancing detectability in WA DBT is as follows: (1) utilizing detectors with a small pixel pitch and low EN level, (2) allocating a higher dose to central projections, and (3) reducing FSM. The results from this study can potentially provide guidance for DBT system optimization in the future.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.