{"title":"系统设计参数对广角数字乳腺断层合成中微小钙化检测的影响的模拟研究。","authors":"Xiaoyu Duan, Hailiang Huang, Wei Zhao","doi":"10.1117/1.JMI.12.S1.S13002","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Accurate detection of microcalcifications ( <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> ) is crucial for the early detection of breast cancer. Some clinical studies have indicated that digital breast tomosynthesis (DBT) systems with a wide angular range have inferior <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability compared with those with a narrow angular range. This study aims to (1) provide guidance for optimizing wide-angle (WA) DBT for improving <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> detectability and (2) prioritize key optimization factors.</p><p><strong>Approach: </strong>An <i>in-silico</i> DBT pipeline was constructed to evaluate <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability of a WA DBT system under various imaging conditions: focal spot motion (FSM), angular dose distribution (ADS), detector pixel pitch, and detector electronic noise (EN). Images were simulated using a digital anthropomorphic breast phantom inserted with <math><mrow><mn>120</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> clusters. Evaluation metrics included the signal-to-noise ratio (SNR) of the filtered channel observer and the area under the receiver operator curve (AUC) of multiple-reader multiple-case analysis.</p><p><strong>Results: </strong>Results showed that FSM degraded <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> sharpness and decreased the SNR and AUC by 5.2% and 1.8%, respectively. Non-uniform ADS increased the SNR by 62.8% and the AUC by 10.2% for filtered backprojection reconstruction with a typical clinical filter setting. When EN decreased from 2000 to 200 electrons, the SNR and AUC increased by 21.6% and 5.0%, respectively. Decreasing the detector pixel pitch from 85 to <math><mrow><mn>50</mn> <mtext>  </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> improved the SNR and AUC by 55.6% and 7.5%, respectively. The combined improvement of a <math><mrow><mn>50</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> pixel pitch and EN200 was 89.2% in the SNR and 12.8% in the AUC.</p><p><strong>Conclusions: </strong>Based on the magnitude of impact, the priority for enhancing <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability in WA DBT is as follows: (1) utilizing detectors with a small pixel pitch and low EN level, (2) allocating a higher dose to central projections, and (3) reducing FSM. The results from this study can potentially provide guidance for DBT system optimization in the future.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11266813/pdf/","citationCount":"0","resultStr":"{\"title\":\"<i>In-silico</i> study of the impact of system design parameters on microcalcification detection in wide-angle digital breast tomosynthesis.\",\"authors\":\"Xiaoyu Duan, Hailiang Huang, Wei Zhao\",\"doi\":\"10.1117/1.JMI.12.S1.S13002\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Accurate detection of microcalcifications ( <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> ) is crucial for the early detection of breast cancer. Some clinical studies have indicated that digital breast tomosynthesis (DBT) systems with a wide angular range have inferior <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability compared with those with a narrow angular range. This study aims to (1) provide guidance for optimizing wide-angle (WA) DBT for improving <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> detectability and (2) prioritize key optimization factors.</p><p><strong>Approach: </strong>An <i>in-silico</i> DBT pipeline was constructed to evaluate <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability of a WA DBT system under various imaging conditions: focal spot motion (FSM), angular dose distribution (ADS), detector pixel pitch, and detector electronic noise (EN). Images were simulated using a digital anthropomorphic breast phantom inserted with <math><mrow><mn>120</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> clusters. Evaluation metrics included the signal-to-noise ratio (SNR) of the filtered channel observer and the area under the receiver operator curve (AUC) of multiple-reader multiple-case analysis.</p><p><strong>Results: </strong>Results showed that FSM degraded <math><mrow><mi>μ</mi> <mi>Calcs</mi></mrow> </math> sharpness and decreased the SNR and AUC by 5.2% and 1.8%, respectively. Non-uniform ADS increased the SNR by 62.8% and the AUC by 10.2% for filtered backprojection reconstruction with a typical clinical filter setting. When EN decreased from 2000 to 200 electrons, the SNR and AUC increased by 21.6% and 5.0%, respectively. Decreasing the detector pixel pitch from 85 to <math><mrow><mn>50</mn> <mtext>  </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> improved the SNR and AUC by 55.6% and 7.5%, respectively. The combined improvement of a <math><mrow><mn>50</mn> <mtext> </mtext> <mi>μ</mi> <mi>m</mi></mrow> </math> pixel pitch and EN200 was 89.2% in the SNR and 12.8% in the AUC.</p><p><strong>Conclusions: </strong>Based on the magnitude of impact, the priority for enhancing <math><mrow><mi>μ</mi> <mi>Calc</mi></mrow> </math> detectability in WA DBT is as follows: (1) utilizing detectors with a small pixel pitch and low EN level, (2) allocating a higher dose to central projections, and (3) reducing FSM. The results from this study can potentially provide guidance for DBT system optimization in the future.</p>\",\"PeriodicalId\":1,\"journal\":{\"name\":\"Accounts of Chemical Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":16.4000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11266813/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Accounts of Chemical Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1117/1.JMI.12.S1.S13002\",\"RegionNum\":1,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/7/24 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1117/1.JMI.12.S1.S13002","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/24 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0

摘要

目的:准确检测微钙化(μ Calcs)对早期发现乳腺癌至关重要。一些临床研究表明,与窄角度范围的数字乳腺断层合成(DBT)系统相比,宽角度范围的数字乳腺断层合成(DBT)系统对微钙化的检测能力较差。本研究旨在:(1) 为优化广角 (WA) DBT 以提高 μ Calc 检测能力提供指导;(2) 优先考虑关键优化因素:方法:构建了一个硅内 DBT 管道,以评估 WA DBT 系统在各种成像条件下的μ Calc 可探测性:焦斑运动 (FSM)、角度剂量分布 (ADS)、探测器像素间距和探测器电子噪声 (EN)。使用插入 120 μ m μ Calc 簇的数字拟人乳房模型模拟图像。评估指标包括滤波通道观测器的信噪比(SNR)和多阅图器多案例分析的接收器运算曲线下面积(AUC):结果表明,FSM 降低了 μ Calcs 的清晰度,信噪比和 AUC 分别下降了 5.2% 和 1.8%。在典型的临床滤波器设置下,非均匀 ADS 使滤波后投影重建的信噪比提高了 62.8%,AUC 提高了 10.2%。当EN从2000电子减少到200电子时,信噪比和AUC分别增加了21.6%和5.0%。探测器像素间距从 85 μ m 减小到 50 μ m 后,信噪比和 AUC 分别提高了 55.6% 和 7.5%。50 μ m 像素间距与 EN200 相结合,信噪比提高了 89.2%,AUC 提高了 12.8%:根据影响程度,在 WA DBT 中提高 μ Calc 可探测性的优先顺序如下:(1) 使用小像素间距和低 EN 水平的探测器;(2) 为中心投影分配更高的剂量;(3) 减少 FSM。这项研究的结果有可能为未来的 DBT 系统优化提供指导。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
In-silico study of the impact of system design parameters on microcalcification detection in wide-angle digital breast tomosynthesis.

Purpose: Accurate detection of microcalcifications ( μ Calcs ) is crucial for the early detection of breast cancer. Some clinical studies have indicated that digital breast tomosynthesis (DBT) systems with a wide angular range have inferior μ Calc detectability compared with those with a narrow angular range. This study aims to (1) provide guidance for optimizing wide-angle (WA) DBT for improving μ Calcs detectability and (2) prioritize key optimization factors.

Approach: An in-silico DBT pipeline was constructed to evaluate μ Calc detectability of a WA DBT system under various imaging conditions: focal spot motion (FSM), angular dose distribution (ADS), detector pixel pitch, and detector electronic noise (EN). Images were simulated using a digital anthropomorphic breast phantom inserted with 120 μ m μ Calc clusters. Evaluation metrics included the signal-to-noise ratio (SNR) of the filtered channel observer and the area under the receiver operator curve (AUC) of multiple-reader multiple-case analysis.

Results: Results showed that FSM degraded μ Calcs sharpness and decreased the SNR and AUC by 5.2% and 1.8%, respectively. Non-uniform ADS increased the SNR by 62.8% and the AUC by 10.2% for filtered backprojection reconstruction with a typical clinical filter setting. When EN decreased from 2000 to 200 electrons, the SNR and AUC increased by 21.6% and 5.0%, respectively. Decreasing the detector pixel pitch from 85 to 50    μ m improved the SNR and AUC by 55.6% and 7.5%, respectively. The combined improvement of a 50 μ m pixel pitch and EN200 was 89.2% in the SNR and 12.8% in the AUC.

Conclusions: Based on the magnitude of impact, the priority for enhancing μ Calc detectability in WA DBT is as follows: (1) utilizing detectors with a small pixel pitch and low EN level, (2) allocating a higher dose to central projections, and (3) reducing FSM. The results from this study can potentially provide guidance for DBT system optimization in the future.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Accounts of Chemical Research
Accounts of Chemical Research 化学-化学综合
CiteScore
31.40
自引率
1.10%
发文量
312
审稿时长
2 months
期刊介绍: Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance. Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.
期刊最新文献
Management of Cholesteatoma: Hearing Rehabilitation. Congenital Cholesteatoma. Evaluation of Cholesteatoma. Management of Cholesteatoma: Extension Beyond Middle Ear/Mastoid. Recidivism and Recurrence.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1