{"title":"一种新的检测方法会错误地增加血浆中的乳酸脱氢酶,但不会增加血清中的乳酸脱氢酶","authors":"","doi":"10.1016/j.clinbiochem.2024.110804","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>Serum is the International Federation of Clinical Chemistry (IFCC)-recommended matrix for the measurement of lactate dehydrogenase (LD); however, many laboratories opt for lithium heparin plasma to achieve quicker turnaround times and minimize tube usage.</p><p>When introducing the new Sigma-Strong IFCC-recommended LDH2 assay from Abbott Laboratories on lithium-heparin collected samples, we observed a rise in the patient median LD activity as well as several samples exhibiting falsely elevated values.</p></div><div><h3>Materials and methods</h3><p>120 + serum and plasma samples from consenting patients were collected and evaluated for complete blood count and lactate dehydrogenase using two different assays. Aggregated patient results before and after introduction of the LDH2 assay were compared.</p></div><div><h3>Results</h3><p>Mean LD was 14% higher in plasma than in serum when using the LDH2 assay but only 5% higher when using the previous LDH legacy assay from Abbott Laboratories. Similarly, platelets and leukocytes were 10–30 times higher in plasma than in serum. Aggregated lactate dehydrogenase patient results demonstrated a dramatic increase in patient median following introduction of the LDH2 assay. Various experiments were tried to reduce cellular interference, but the only viable solution we found, apart from reverting to the LDH legacy assay, was to utilize serum tubes.</p></div><div><h3>Conclusion</h3><p>We conclude that lithium-heparin plasma leads to falsely elevated lactate dehydrogenase activity when using the LDH2 assay. These errors can be prevented by using serum collected in gel separator tubes.</p></div>","PeriodicalId":10172,"journal":{"name":"Clinical biochemistry","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2024-07-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A new assay falsely increases lactate dehydrogenase in plasma but not in serum\",\"authors\":\"\",\"doi\":\"10.1016/j.clinbiochem.2024.110804\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>Serum is the International Federation of Clinical Chemistry (IFCC)-recommended matrix for the measurement of lactate dehydrogenase (LD); however, many laboratories opt for lithium heparin plasma to achieve quicker turnaround times and minimize tube usage.</p><p>When introducing the new Sigma-Strong IFCC-recommended LDH2 assay from Abbott Laboratories on lithium-heparin collected samples, we observed a rise in the patient median LD activity as well as several samples exhibiting falsely elevated values.</p></div><div><h3>Materials and methods</h3><p>120 + serum and plasma samples from consenting patients were collected and evaluated for complete blood count and lactate dehydrogenase using two different assays. Aggregated patient results before and after introduction of the LDH2 assay were compared.</p></div><div><h3>Results</h3><p>Mean LD was 14% higher in plasma than in serum when using the LDH2 assay but only 5% higher when using the previous LDH legacy assay from Abbott Laboratories. Similarly, platelets and leukocytes were 10–30 times higher in plasma than in serum. Aggregated lactate dehydrogenase patient results demonstrated a dramatic increase in patient median following introduction of the LDH2 assay. Various experiments were tried to reduce cellular interference, but the only viable solution we found, apart from reverting to the LDH legacy assay, was to utilize serum tubes.</p></div><div><h3>Conclusion</h3><p>We conclude that lithium-heparin plasma leads to falsely elevated lactate dehydrogenase activity when using the LDH2 assay. These errors can be prevented by using serum collected in gel separator tubes.</p></div>\",\"PeriodicalId\":10172,\"journal\":{\"name\":\"Clinical biochemistry\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.5000,\"publicationDate\":\"2024-07-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical biochemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0009912024000985\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical biochemistry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0009912024000985","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

导言:血清是国际临床化学联合会(IFCC)推荐用于测定乳酸脱氢酶(LD)的基质;然而,许多实验室选择使用肝素锂血浆,以实现更快的周转时间并最大限度地减少试管用量。在将雅培实验室新推出的 Sigma-Strong IFCC 推荐的 LDH2 检测法引入锂肝素采集的样本时,我们观察到患者 LD 活性的中位数有所上升,而且有几个样本的值出现了假性升高。材料与方法 120 + 经患者同意采集的血清和血浆样本,并使用两种不同的检测法对其进行全血细胞计数和乳酸脱氢酶评估。结果 使用 LDH2 检测法时,血浆中的平均乳酸脱氢酶比血清中高 14%,而使用雅培实验室以前的 LDH 传统检测法时仅高 5%。同样,血浆中的血小板和白细胞也比血清中高 10-30 倍。患者的乳酸脱氢酶聚集结果表明,采用 LDH2 检测法后,患者的中位数急剧增加。我们尝试了各种实验来减少细胞干扰,但除了重新使用 LDH 传统测定法外,我们发现唯一可行的办法就是使用血清试管。使用凝胶分离管收集血清可以避免这些错误。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A new assay falsely increases lactate dehydrogenase in plasma but not in serum

Introduction

Serum is the International Federation of Clinical Chemistry (IFCC)-recommended matrix for the measurement of lactate dehydrogenase (LD); however, many laboratories opt for lithium heparin plasma to achieve quicker turnaround times and minimize tube usage.

When introducing the new Sigma-Strong IFCC-recommended LDH2 assay from Abbott Laboratories on lithium-heparin collected samples, we observed a rise in the patient median LD activity as well as several samples exhibiting falsely elevated values.

Materials and methods

120 + serum and plasma samples from consenting patients were collected and evaluated for complete blood count and lactate dehydrogenase using two different assays. Aggregated patient results before and after introduction of the LDH2 assay were compared.

Results

Mean LD was 14% higher in plasma than in serum when using the LDH2 assay but only 5% higher when using the previous LDH legacy assay from Abbott Laboratories. Similarly, platelets and leukocytes were 10–30 times higher in plasma than in serum. Aggregated lactate dehydrogenase patient results demonstrated a dramatic increase in patient median following introduction of the LDH2 assay. Various experiments were tried to reduce cellular interference, but the only viable solution we found, apart from reverting to the LDH legacy assay, was to utilize serum tubes.

Conclusion

We conclude that lithium-heparin plasma leads to falsely elevated lactate dehydrogenase activity when using the LDH2 assay. These errors can be prevented by using serum collected in gel separator tubes.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical biochemistry
Clinical biochemistry 医学-医学实验技术
CiteScore
5.10
自引率
0.00%
发文量
151
审稿时长
25 days
期刊介绍: Clinical Biochemistry publishes articles relating to clinical chemistry, molecular biology and genetics, therapeutic drug monitoring and toxicology, laboratory immunology and laboratory medicine in general, with the focus on analytical and clinical investigation of laboratory tests in humans used for diagnosis, prognosis, treatment and therapy, and monitoring of disease.
期刊最新文献
Challenges and Perspectives on the Adoption of Cystatin C testing in China: A laboratory technician’s perspective Quantitative evaluation of adalimumab and anti-adalimumab antibodies in sera using a surface plasmon resonance biosensor Inflammation and lipoperoxidation in mucopolysaccharidoses type II patients at diagnosis and post-hematopoietic stem cell transplantation. Multidisciplinary approach to redefining thyroid hormone reference intervals with big data analysis Relationship between analytical imprecision and coefficient of determination (R2) of the calibration curve
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1