{"title":"人体观察者对 X 射线透视和血管造影图像质量的印象与图像质量技术变化的比较。","authors":"Jelena M Mihailovic, Yoshihisa Kanaji, Daniel Miller, Malcolm R Bell, Kenneth A Fetterly","doi":"10.1117/1.JMI.11.4.045502","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Spatio-temporal variability in clinical fluoroscopy and cine angiography images combined with nonlinear image processing prevents the application of traditional image quality measurements in the cardiac catheterization laboratory. We aimed to develop and validate methods to measure human observer impressions of the image quality.</p><p><strong>Approach: </strong>Multi-frame images of the thorax of a euthanized pig were acquired to provide an anatomical background. The detector dose was varied from 6 to 200 nGy (increments 2×), and 0.6 and 1.0 mm focal spots were used. Two coronary stents with/without 0.5 mm separation and a synthetic right coronary artery (RCA) with hemispherical defects were embedded into the background images as test objects. The quantitative observer ( <math><mrow><mi>n</mi> <mo>=</mo> <mn>17</mn></mrow> </math> ) performance was measured using a two-alternating forced-choice test of whether stents were separated and by a count of visible right coronary artery defects. Qualitative impressions of noise, spatial resolution, and overall image quality were measured using a visual analog scale (VAS). A paired <math><mrow><mi>t</mi></mrow> </math> -test and multinomial logistic regression model were used to identify statistically significant factors affecting the observer's impression image quality.</p><p><strong>Results: </strong>The proportion of correct detection of stent separation and the number of reported right coronary artery defects changed significantly with detector dose increment in the 6 to 100 nGy ( <math><mrow><mi>p</mi> <mo><</mo> <mn>0.05</mn></mrow> </math> ). Although a trend favored the 0.6 versus 1.0 mm focal spot for these quantitative assessments, this was insignificant. Visual analog scale measurements changed significantly with detector dose increments in the range of 24 to 100 nGy and focal spot size ( <math><mrow><mi>p</mi> <mo><</mo> <mn>0.05</mn></mrow> </math> ). The application of multinomial logistic regression analysis to observer VAS scores demonstrated sensitivity matching of the paired <math><mrow><mi>t</mi></mrow> </math> -test applied to quantitative observer performance measurements.</p><p><strong>Conclusions: </strong>Both quantitative and qualitative measurements of observer impression of the image quality were sensitive to image quality changes associated with changing the detector dose and focal spot size. These findings encourage future work that uses qualitative image quality measurements to assess clinical fluoroscopy and angiography image quality.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11316400/pdf/","citationCount":"0","resultStr":"{\"title\":\"Comparison of human observer impression of X-ray fluoroscopy and angiography image quality with technical changes to image quality.\",\"authors\":\"Jelena M Mihailovic, Yoshihisa Kanaji, Daniel Miller, Malcolm R Bell, Kenneth A Fetterly\",\"doi\":\"10.1117/1.JMI.11.4.045502\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Spatio-temporal variability in clinical fluoroscopy and cine angiography images combined with nonlinear image processing prevents the application of traditional image quality measurements in the cardiac catheterization laboratory. We aimed to develop and validate methods to measure human observer impressions of the image quality.</p><p><strong>Approach: </strong>Multi-frame images of the thorax of a euthanized pig were acquired to provide an anatomical background. The detector dose was varied from 6 to 200 nGy (increments 2×), and 0.6 and 1.0 mm focal spots were used. Two coronary stents with/without 0.5 mm separation and a synthetic right coronary artery (RCA) with hemispherical defects were embedded into the background images as test objects. The quantitative observer ( <math><mrow><mi>n</mi> <mo>=</mo> <mn>17</mn></mrow> </math> ) performance was measured using a two-alternating forced-choice test of whether stents were separated and by a count of visible right coronary artery defects. Qualitative impressions of noise, spatial resolution, and overall image quality were measured using a visual analog scale (VAS). A paired <math><mrow><mi>t</mi></mrow> </math> -test and multinomial logistic regression model were used to identify statistically significant factors affecting the observer's impression image quality.</p><p><strong>Results: </strong>The proportion of correct detection of stent separation and the number of reported right coronary artery defects changed significantly with detector dose increment in the 6 to 100 nGy ( <math><mrow><mi>p</mi> <mo><</mo> <mn>0.05</mn></mrow> </math> ). Although a trend favored the 0.6 versus 1.0 mm focal spot for these quantitative assessments, this was insignificant. Visual analog scale measurements changed significantly with detector dose increments in the range of 24 to 100 nGy and focal spot size ( <math><mrow><mi>p</mi> <mo><</mo> <mn>0.05</mn></mrow> </math> ). The application of multinomial logistic regression analysis to observer VAS scores demonstrated sensitivity matching of the paired <math><mrow><mi>t</mi></mrow> </math> -test applied to quantitative observer performance measurements.</p><p><strong>Conclusions: </strong>Both quantitative and qualitative measurements of observer impression of the image quality were sensitive to image quality changes associated with changing the detector dose and focal spot size. These findings encourage future work that uses qualitative image quality measurements to assess clinical fluoroscopy and angiography image quality.</p>\",\"PeriodicalId\":1,\"journal\":{\"name\":\"Accounts of Chemical Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":16.4000,\"publicationDate\":\"2024-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11316400/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Accounts of Chemical Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1117/1.JMI.11.4.045502\",\"RegionNum\":1,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/8/10 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1117/1.JMI.11.4.045502","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/10 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
Comparison of human observer impression of X-ray fluoroscopy and angiography image quality with technical changes to image quality.
Purpose: Spatio-temporal variability in clinical fluoroscopy and cine angiography images combined with nonlinear image processing prevents the application of traditional image quality measurements in the cardiac catheterization laboratory. We aimed to develop and validate methods to measure human observer impressions of the image quality.
Approach: Multi-frame images of the thorax of a euthanized pig were acquired to provide an anatomical background. The detector dose was varied from 6 to 200 nGy (increments 2×), and 0.6 and 1.0 mm focal spots were used. Two coronary stents with/without 0.5 mm separation and a synthetic right coronary artery (RCA) with hemispherical defects were embedded into the background images as test objects. The quantitative observer ( ) performance was measured using a two-alternating forced-choice test of whether stents were separated and by a count of visible right coronary artery defects. Qualitative impressions of noise, spatial resolution, and overall image quality were measured using a visual analog scale (VAS). A paired -test and multinomial logistic regression model were used to identify statistically significant factors affecting the observer's impression image quality.
Results: The proportion of correct detection of stent separation and the number of reported right coronary artery defects changed significantly with detector dose increment in the 6 to 100 nGy ( ). Although a trend favored the 0.6 versus 1.0 mm focal spot for these quantitative assessments, this was insignificant. Visual analog scale measurements changed significantly with detector dose increments in the range of 24 to 100 nGy and focal spot size ( ). The application of multinomial logistic regression analysis to observer VAS scores demonstrated sensitivity matching of the paired -test applied to quantitative observer performance measurements.
Conclusions: Both quantitative and qualitative measurements of observer impression of the image quality were sensitive to image quality changes associated with changing the detector dose and focal spot size. These findings encourage future work that uses qualitative image quality measurements to assess clinical fluoroscopy and angiography image quality.
期刊介绍:
Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance.
Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.