{"title":"拉丁美洲第一台 MRlinac 投入使用。","authors":"Rojas-López José Alejandro, Cabrera-Santiago Alexis, González Souto Xesús","doi":"10.4103/jmp.jmp_6_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To show the workflow for the commissioning of a MRlinac, and some proposed tests; off-axis targets, output factors for small fields, dose in inhomogeneities, and multileaf collimator quality assurance (MLC QA).</p><p><strong>Methods: </strong>The tests were performed based on TG-142, TG-119, ICRU 97, TRS-398, and TRS-483 recommendations as well as national regulations for radiation protection and safety.</p><p><strong>Results: </strong>The imaging tests are in agreement with the protocols. The radiation isocenter was 0.34 mm, and for off-axis targets location was up to 0.88 mm. The dose profiles measured and calculated in treatment planning system (TPS) passed in all cases the gamma analysis of 2%/2 mm (global dose differences). The output factors of fields larger than 2 cm × 2 cm are in agreement with the model of the MRlinac in the TPS. However, for smaller fields, their differences are higher than 10%. Picket fence test for different gantry angles showed a maximum leaf deviation up to 0.2 mm. Displacements observed in treatment couch adding weight (50 kg) are lower than 1 mm. Cryostat, bridge, and couch attenuation was up to 1.2%, 10%, and 24%, respectively.</p><p><strong>Conclusion: </strong>The implemented tests confirm that the studied MRlinac agrees with the standards reported in the literature and that the strict tolerances established as a baseline should allow a smoother implementation of stereotactic treatments in this machine.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 2","pages":"213-224"},"PeriodicalIF":0.7000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11309142/pdf/","citationCount":"0","resultStr":"{\"title\":\"Commissioning of the First MRlinac in Latin America.\",\"authors\":\"Rojas-López José Alejandro, Cabrera-Santiago Alexis, González Souto Xesús\",\"doi\":\"10.4103/jmp.jmp_6_24\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To show the workflow for the commissioning of a MRlinac, and some proposed tests; off-axis targets, output factors for small fields, dose in inhomogeneities, and multileaf collimator quality assurance (MLC QA).</p><p><strong>Methods: </strong>The tests were performed based on TG-142, TG-119, ICRU 97, TRS-398, and TRS-483 recommendations as well as national regulations for radiation protection and safety.</p><p><strong>Results: </strong>The imaging tests are in agreement with the protocols. The radiation isocenter was 0.34 mm, and for off-axis targets location was up to 0.88 mm. The dose profiles measured and calculated in treatment planning system (TPS) passed in all cases the gamma analysis of 2%/2 mm (global dose differences). The output factors of fields larger than 2 cm × 2 cm are in agreement with the model of the MRlinac in the TPS. However, for smaller fields, their differences are higher than 10%. Picket fence test for different gantry angles showed a maximum leaf deviation up to 0.2 mm. Displacements observed in treatment couch adding weight (50 kg) are lower than 1 mm. Cryostat, bridge, and couch attenuation was up to 1.2%, 10%, and 24%, respectively.</p><p><strong>Conclusion: </strong>The implemented tests confirm that the studied MRlinac agrees with the standards reported in the literature and that the strict tolerances established as a baseline should allow a smoother implementation of stereotactic treatments in this machine.</p>\",\"PeriodicalId\":51719,\"journal\":{\"name\":\"Journal of Medical Physics\",\"volume\":\"49 2\",\"pages\":\"213-224\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11309142/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/jmp.jmp_6_24\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/6/25 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_6_24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/25 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Commissioning of the First MRlinac in Latin America.
Purpose: To show the workflow for the commissioning of a MRlinac, and some proposed tests; off-axis targets, output factors for small fields, dose in inhomogeneities, and multileaf collimator quality assurance (MLC QA).
Methods: The tests were performed based on TG-142, TG-119, ICRU 97, TRS-398, and TRS-483 recommendations as well as national regulations for radiation protection and safety.
Results: The imaging tests are in agreement with the protocols. The radiation isocenter was 0.34 mm, and for off-axis targets location was up to 0.88 mm. The dose profiles measured and calculated in treatment planning system (TPS) passed in all cases the gamma analysis of 2%/2 mm (global dose differences). The output factors of fields larger than 2 cm × 2 cm are in agreement with the model of the MRlinac in the TPS. However, for smaller fields, their differences are higher than 10%. Picket fence test for different gantry angles showed a maximum leaf deviation up to 0.2 mm. Displacements observed in treatment couch adding weight (50 kg) are lower than 1 mm. Cryostat, bridge, and couch attenuation was up to 1.2%, 10%, and 24%, respectively.
Conclusion: The implemented tests confirm that the studied MRlinac agrees with the standards reported in the literature and that the strict tolerances established as a baseline should allow a smoother implementation of stereotactic treatments in this machine.
期刊介绍:
JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.