{"title":"拉丁美洲第一台 MRlinac 投入使用。","authors":"Rojas-López José Alejandro, Cabrera-Santiago Alexis, González Souto Xesús","doi":"10.4103/jmp.jmp_6_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To show the workflow for the commissioning of a MRlinac, and some proposed tests; off-axis targets, output factors for small fields, dose in inhomogeneities, and multileaf collimator quality assurance (MLC QA).</p><p><strong>Methods: </strong>The tests were performed based on TG-142, TG-119, ICRU 97, TRS-398, and TRS-483 recommendations as well as national regulations for radiation protection and safety.</p><p><strong>Results: </strong>The imaging tests are in agreement with the protocols. The radiation isocenter was 0.34 mm, and for off-axis targets location was up to 0.88 mm. The dose profiles measured and calculated in treatment planning system (TPS) passed in all cases the gamma analysis of 2%/2 mm (global dose differences). The output factors of fields larger than 2 cm × 2 cm are in agreement with the model of the MRlinac in the TPS. However, for smaller fields, their differences are higher than 10%. Picket fence test for different gantry angles showed a maximum leaf deviation up to 0.2 mm. Displacements observed in treatment couch adding weight (50 kg) are lower than 1 mm. Cryostat, bridge, and couch attenuation was up to 1.2%, 10%, and 24%, respectively.</p><p><strong>Conclusion: </strong>The implemented tests confirm that the studied MRlinac agrees with the standards reported in the literature and that the strict tolerances established as a baseline should allow a smoother implementation of stereotactic treatments in this machine.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 2","pages":"213-224"},"PeriodicalIF":0.7000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11309142/pdf/","citationCount":"0","resultStr":"{\"title\":\"Commissioning of the First MRlinac in Latin America.\",\"authors\":\"Rojas-López José Alejandro, Cabrera-Santiago Alexis, González Souto Xesús\",\"doi\":\"10.4103/jmp.jmp_6_24\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To show the workflow for the commissioning of a MRlinac, and some proposed tests; off-axis targets, output factors for small fields, dose in inhomogeneities, and multileaf collimator quality assurance (MLC QA).</p><p><strong>Methods: </strong>The tests were performed based on TG-142, TG-119, ICRU 97, TRS-398, and TRS-483 recommendations as well as national regulations for radiation protection and safety.</p><p><strong>Results: </strong>The imaging tests are in agreement with the protocols. The radiation isocenter was 0.34 mm, and for off-axis targets location was up to 0.88 mm. The dose profiles measured and calculated in treatment planning system (TPS) passed in all cases the gamma analysis of 2%/2 mm (global dose differences). The output factors of fields larger than 2 cm × 2 cm are in agreement with the model of the MRlinac in the TPS. However, for smaller fields, their differences are higher than 10%. Picket fence test for different gantry angles showed a maximum leaf deviation up to 0.2 mm. Displacements observed in treatment couch adding weight (50 kg) are lower than 1 mm. Cryostat, bridge, and couch attenuation was up to 1.2%, 10%, and 24%, respectively.</p><p><strong>Conclusion: </strong>The implemented tests confirm that the studied MRlinac agrees with the standards reported in the literature and that the strict tolerances established as a baseline should allow a smoother implementation of stereotactic treatments in this machine.</p>\",\"PeriodicalId\":51719,\"journal\":{\"name\":\"Journal of Medical Physics\",\"volume\":\"49 2\",\"pages\":\"213-224\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2024-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11309142/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/jmp.jmp_6_24\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/6/25 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_6_24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/25 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:展示磁共振成像仪调试的工作流程和一些建议的测试;离轴目标、小场输出系数、不均匀剂量和多叶准直器质量保证(MLC QA):方法:根据 TG-142、TG-119、ICRU 97、TRS-398 和 TRS-483 建议以及国家辐射防护和安全法规进行测试:结果:成像测试与规程一致。辐射等中心为 0.34 毫米,离轴目标位置可达 0.88 毫米。在治疗计划系统(TPS)中测量和计算的剂量曲线在所有情况下都通过了 2%/2 mm 的伽马分析(全局剂量差异)。大于 2 厘米×2 厘米的磁场的输出因子与 TPS 中的 MRlinac 模型一致。然而,对于较小的磁场,它们之间的差异高于 10%。不同龙门角度的栅栏测试显示,最大叶片偏差达 0.2 毫米。在增加重量(50 千克)的处理沙发上观察到的位移低于 1 毫米。低温恒温器、桥梁和沙发的衰减分别高达 1.2%、10% 和 24%:实施的测试证实,所研究的磁共振成像仪符合文献报道的标准,而且作为基线建立的严格公差应能使该设备更顺利地实施立体定向治疗。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Commissioning of the First MRlinac in Latin America.

Purpose: To show the workflow for the commissioning of a MRlinac, and some proposed tests; off-axis targets, output factors for small fields, dose in inhomogeneities, and multileaf collimator quality assurance (MLC QA).

Methods: The tests were performed based on TG-142, TG-119, ICRU 97, TRS-398, and TRS-483 recommendations as well as national regulations for radiation protection and safety.

Results: The imaging tests are in agreement with the protocols. The radiation isocenter was 0.34 mm, and for off-axis targets location was up to 0.88 mm. The dose profiles measured and calculated in treatment planning system (TPS) passed in all cases the gamma analysis of 2%/2 mm (global dose differences). The output factors of fields larger than 2 cm × 2 cm are in agreement with the model of the MRlinac in the TPS. However, for smaller fields, their differences are higher than 10%. Picket fence test for different gantry angles showed a maximum leaf deviation up to 0.2 mm. Displacements observed in treatment couch adding weight (50 kg) are lower than 1 mm. Cryostat, bridge, and couch attenuation was up to 1.2%, 10%, and 24%, respectively.

Conclusion: The implemented tests confirm that the studied MRlinac agrees with the standards reported in the literature and that the strict tolerances established as a baseline should allow a smoother implementation of stereotactic treatments in this machine.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Physics
Journal of Medical Physics RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.10
自引率
11.10%
发文量
55
审稿时长
30 weeks
期刊介绍: JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.
期刊最新文献
A Segmentation-based Automated Calculation of Patient Size and Size-specific Dose Estimates in Pediatric Computed Tomography Scans. A Study on Radiation Level at the Treatment Plane Due to Induced Activity in Linear Accelerator Head. Advancements and Applications of Three-dimensional Printing Technology in Surgery. Agar-based Phantom for Evaluating Targeting of High-intensity Focused Ultrasound Systems for Breast Ablation. An Analysis of Radiotherapy Machine Requirements in India: Impact of the Pandemic and Regional Disparities.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1