{"title":"在弹道模型中比较弯头和直头射频套管的偏转情况","authors":"Allen Chen ,&nbsp;Udai Nanda ,&nbsp;Joseph Solberg ,&nbsp;Ethan Rand ,&nbsp;George Christolias ,&nbsp;Jaspal Ricky Singh","doi":"10.1016/j.inpm.2024.100429","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><p>Percutaneous pain and spine procedures play an important diagnostic and therapeutic role in the treatment of various pain diagnoses. Accurate placement of needles or cannulae during these procedures is paramount to the success of these procedures.</p></div><div><h3>Objective</h3><p>The purpose of this study is to examine and quantify the amount of deflection of radiofrequency cannulae based on curved tip versus no curved tip, using a ballistic gel tissue simulant.</p></div><div><h3>Materials and methods</h3><p>Six different types of cannulae commonly used for spinal and peripheral nerve ablations were selected, including 18, 20, and 22 gauge curved and straight radiofrequency cannulae. Ballistic gel samples were made in molds of 40 mm and 80 mm. Each cannula was mounted in a drill press to ensure accurate trajectory.</p></div><div><h3>Results</h3><p>Curved RFA cannula had increased deflection when compared to straight cannula for 18-, 20-, and 22-gauge cannulae at a depth of 40 mm. Curved RFA cannula had increased deflection when compared to straight cannula for 20- and 22-gauge cannulae at a depth of 80 mm. Overall, the mean deflection for a curved cannula increased 1.9x for 20-gauge cannulae and 2.5x for 22-gauge cannulae when compared to a straight cannula.</p></div><div><h3>Conclusions</h3><p>For interventionalists, understanding the effects of needle or cannula shape is crucial for accurate placement. When a procedure requires additional steerability, additional deflection up to 2.5x obtained by placing a bend in the needle or cannula tip should be considered.</p></div>","PeriodicalId":100727,"journal":{"name":"Interventional Pain Medicine","volume":"3 3","pages":"Article 100429"},"PeriodicalIF":0.0000,"publicationDate":"2024-07-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2772594424000505/pdfft?md5=b88f037b6b9a7dc41a9333b499ff6967&pid=1-s2.0-S2772594424000505-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Comparison of curved and straight tip radiofrequency cannula deflection in a ballistic model\",\"authors\":\"Allen Chen ,&nbsp;Udai Nanda ,&nbsp;Joseph Solberg ,&nbsp;Ethan Rand ,&nbsp;George Christolias ,&nbsp;Jaspal Ricky Singh\",\"doi\":\"10.1016/j.inpm.2024.100429\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><p>Percutaneous pain and spine procedures play an important diagnostic and therapeutic role in the treatment of various pain diagnoses. Accurate placement of needles or cannulae during these procedures is paramount to the success of these procedures.</p></div><div><h3>Objective</h3><p>The purpose of this study is to examine and quantify the amount of deflection of radiofrequency cannulae based on curved tip versus no curved tip, using a ballistic gel tissue simulant.</p></div><div><h3>Materials and methods</h3><p>Six different types of cannulae commonly used for spinal and peripheral nerve ablations were selected, including 18, 20, and 22 gauge curved and straight radiofrequency cannulae. Ballistic gel samples were made in molds of 40 mm and 80 mm. Each cannula was mounted in a drill press to ensure accurate trajectory.</p></div><div><h3>Results</h3><p>Curved RFA cannula had increased deflection when compared to straight cannula for 18-, 20-, and 22-gauge cannulae at a depth of 40 mm. Curved RFA cannula had increased deflection when compared to straight cannula for 20- and 22-gauge cannulae at a depth of 80 mm. Overall, the mean deflection for a curved cannula increased 1.9x for 20-gauge cannulae and 2.5x for 22-gauge cannulae when compared to a straight cannula.</p></div><div><h3>Conclusions</h3><p>For interventionalists, understanding the effects of needle or cannula shape is crucial for accurate placement. When a procedure requires additional steerability, additional deflection up to 2.5x obtained by placing a bend in the needle or cannula tip should be considered.</p></div>\",\"PeriodicalId\":100727,\"journal\":{\"name\":\"Interventional Pain Medicine\",\"volume\":\"3 3\",\"pages\":\"Article 100429\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-07-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2772594424000505/pdfft?md5=b88f037b6b9a7dc41a9333b499ff6967&pid=1-s2.0-S2772594424000505-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Interventional Pain Medicine\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2772594424000505\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Interventional Pain Medicine","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2772594424000505","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

背景经皮疼痛和脊柱手术在治疗各种疼痛诊断中发挥着重要的诊断和治疗作用。本研究的目的是使用弹道凝胶组织模拟物检查和量化射频套管的偏转量,根据弯曲尖端与无弯曲尖端进行比较。材料和方法选择了脊柱和周围神经消融常用的六种不同类型的套管,包括 18 号、20 号和 22 号弯曲和笔直射频套管。弹道凝胶样品分别用 40 毫米和 80 毫米的模具制成。结果与直套管相比,在深度为 40 毫米的 18、20 和 22 号套管中,弧形射频消融套管的偏转增加。对于深度为 80 毫米的 20 号和 22 号插管,弧形 RFA 插管与直插管相比偏转更大。总体而言,与直套管相比,20 号套管的弧形套管的平均偏转增加了 1.9 倍,22 号套管的平均偏转增加了 2.5 倍。当手术需要额外的可转向性时,应考虑通过在针头或套管顶端设置弯曲来获得额外的偏转,最高可达 2.5 倍。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Comparison of curved and straight tip radiofrequency cannula deflection in a ballistic model

Background

Percutaneous pain and spine procedures play an important diagnostic and therapeutic role in the treatment of various pain diagnoses. Accurate placement of needles or cannulae during these procedures is paramount to the success of these procedures.

Objective

The purpose of this study is to examine and quantify the amount of deflection of radiofrequency cannulae based on curved tip versus no curved tip, using a ballistic gel tissue simulant.

Materials and methods

Six different types of cannulae commonly used for spinal and peripheral nerve ablations were selected, including 18, 20, and 22 gauge curved and straight radiofrequency cannulae. Ballistic gel samples were made in molds of 40 mm and 80 mm. Each cannula was mounted in a drill press to ensure accurate trajectory.

Results

Curved RFA cannula had increased deflection when compared to straight cannula for 18-, 20-, and 22-gauge cannulae at a depth of 40 mm. Curved RFA cannula had increased deflection when compared to straight cannula for 20- and 22-gauge cannulae at a depth of 80 mm. Overall, the mean deflection for a curved cannula increased 1.9x for 20-gauge cannulae and 2.5x for 22-gauge cannulae when compared to a straight cannula.

Conclusions

For interventionalists, understanding the effects of needle or cannula shape is crucial for accurate placement. When a procedure requires additional steerability, additional deflection up to 2.5x obtained by placing a bend in the needle or cannula tip should be considered.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Intraosseous basivertebral nerve ablation: A 5-year pooled analysis from three prospective clinical trials. Evaluating the effectiveness of interlaminar epidural steroid injections for cervical radiculopathy using PROMIS as an outcome measure. Spinal cord stimulation for the treatment of complex regional pain syndrome: A systematic review of randomized controlled trials. Postherpetic neuralgia mimicking lumbar radiculopathy. Comparing the efficacy of intra-articular injection of Platelet Rich Plasma (PRP) with corticosteroids (CS) in patients with chronic zygapophyseal joint low back pain confirmed by double intra-articular diagnostic blocks: A triple-blinded randomized multicentric controlled trial with a 6-month follow-up.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1