{"title":"[300 万和 500 万次循环时聚乙烯衬垫磨损性能的体外分析]。","authors":"Shitong Yan, Jianjun Wang, Jian Su","doi":"10.12455/j.issn.1671-7104.230705","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>The purpose of this study is to compare the wear properties of UHMWPE acetabular liners after undergoing 3 million (3 Mc) and 5 million (5 Mc) cycles of <i>in-vitro</i> wear testing. The results will serve as a reference for the design of <i>in-vitro</i> testing for hip prostheses and as a control for clinical revision removals.</p><p><strong>Methods: </strong>Wear tests were conducted on three different sizes of acetabular liners (28 mm, 32 mm, and 36 mm internal diameters) using a hip simulator to determine the amount of wear after 3 and 5 million cycles. The analysis included the number, size, and shape of abrasive particles.</p><p><strong>Results: </strong>After 3 and 5 million cycles of wear, the amount of wear on the acetabular liner increased as the inner diameter increased. The abrasive particles had an average equivalent circular area diameter (ECD) of 0.27 μm and 0.29 μm, and 94.4% and 90.1% of the aspect ratio (AR) less than 4.</p><p><strong>Conclusion: </strong>The amount of wear on the acetabular liner after 3 Mc wear can indicate the wear performance of the product. The number of particles increased and the percentage of fibrous particles was higher after 5 Mc wear compared to 3 Mc wear.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-07-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[<i>In-Vitro</i> Analysis of Polyethylene Liner Wear Performance at 3 and 5 Million Cycles].\",\"authors\":\"Shitong Yan, Jianjun Wang, Jian Su\",\"doi\":\"10.12455/j.issn.1671-7104.230705\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>The purpose of this study is to compare the wear properties of UHMWPE acetabular liners after undergoing 3 million (3 Mc) and 5 million (5 Mc) cycles of <i>in-vitro</i> wear testing. The results will serve as a reference for the design of <i>in-vitro</i> testing for hip prostheses and as a control for clinical revision removals.</p><p><strong>Methods: </strong>Wear tests were conducted on three different sizes of acetabular liners (28 mm, 32 mm, and 36 mm internal diameters) using a hip simulator to determine the amount of wear after 3 and 5 million cycles. The analysis included the number, size, and shape of abrasive particles.</p><p><strong>Results: </strong>After 3 and 5 million cycles of wear, the amount of wear on the acetabular liner increased as the inner diameter increased. The abrasive particles had an average equivalent circular area diameter (ECD) of 0.27 μm and 0.29 μm, and 94.4% and 90.1% of the aspect ratio (AR) less than 4.</p><p><strong>Conclusion: </strong>The amount of wear on the acetabular liner after 3 Mc wear can indicate the wear performance of the product. The number of particles increased and the percentage of fibrous particles was higher after 5 Mc wear compared to 3 Mc wear.</p>\",\"PeriodicalId\":52535,\"journal\":{\"name\":\"中国医疗器械杂志\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-07-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中国医疗器械杂志\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://doi.org/10.12455/j.issn.1671-7104.230705\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.230705","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究旨在比较超高分子量聚乙烯髋臼衬垫在经过 300 万 (3 Mc) 次和 500 万 (5 Mc) 次体外磨损测试后的磨损特性。结果将作为髋关节假体体外测试设计的参考,以及临床翻修取出的对照:使用髋关节模拟器对三种不同尺寸的髋臼衬垫(内径分别为 28 毫米、32 毫米和 36 毫米)进行了磨损测试,以确定 300 万和 500 万次循环后的磨损量。分析包括磨粒的数量、大小和形状:结果:经过 300 万和 500 万次循环磨损后,髋臼衬垫的磨损量随着内径的增加而增加。磨料颗粒的平均等效圆面积直径(ECD)分别为 0.27 μm 和 0.29 μm,94.4% 和 90.1% 的长宽比(AR)小于 4.结论:经过 3 Mc 磨损后,髋臼衬垫的磨损量可以说明产品的磨损性能。与 3 Mc 磨损相比,5 Mc 磨损后颗粒数量增加,纤维颗粒的百分比更高。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[In-Vitro Analysis of Polyethylene Liner Wear Performance at 3 and 5 Million Cycles].

Objective: The purpose of this study is to compare the wear properties of UHMWPE acetabular liners after undergoing 3 million (3 Mc) and 5 million (5 Mc) cycles of in-vitro wear testing. The results will serve as a reference for the design of in-vitro testing for hip prostheses and as a control for clinical revision removals.

Methods: Wear tests were conducted on three different sizes of acetabular liners (28 mm, 32 mm, and 36 mm internal diameters) using a hip simulator to determine the amount of wear after 3 and 5 million cycles. The analysis included the number, size, and shape of abrasive particles.

Results: After 3 and 5 million cycles of wear, the amount of wear on the acetabular liner increased as the inner diameter increased. The abrasive particles had an average equivalent circular area diameter (ECD) of 0.27 μm and 0.29 μm, and 94.4% and 90.1% of the aspect ratio (AR) less than 4.

Conclusion: The amount of wear on the acetabular liner after 3 Mc wear can indicate the wear performance of the product. The number of particles increased and the percentage of fibrous particles was higher after 5 Mc wear compared to 3 Mc wear.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中国医疗器械杂志
中国医疗器械杂志 Medicine-Medicine (all)
CiteScore
0.40
自引率
0.00%
发文量
8086
期刊介绍:
期刊最新文献
[12-Lead Holter Integrated with Sleep Monitoring Module]. [Bowel Sounds Detection Method Based on ResNet-BiLSTM and Attention Mechanism]. [Clinical Application of Equivalent Uniform Dose in Intensity-Modulated Rotational Radiotherapy Based on Eclipse TPS]. [Clinical Validation of a Prototype Smart Non-Invasive Pregnancy Glucose Monitor]. [Clinical Validation Study of Deep Learning-Generated Magnetic Resonance Images].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1