利用幂律溶血模型中使用的不同系数集和标度剪切应力模型,对 FDA 基准血泵进行计算流体动力学模拟。

IF 1.1 4区 医学 Q4 ENGINEERING, BIOMEDICAL Journal of Artificial Organs Pub Date : 2024-08-23 DOI:10.1007/s10047-024-01468-6
Ahmet Onder, Omer Incebay, Rafet Yapici
{"title":"利用幂律溶血模型中使用的不同系数集和标度剪切应力模型,对 FDA 基准血泵进行计算流体动力学模拟。","authors":"Ahmet Onder, Omer Incebay, Rafet Yapici","doi":"10.1007/s10047-024-01468-6","DOIUrl":null,"url":null,"abstract":"<p><p>Hemolysis is the most important issue to consider in the design and optimization of blood-contacting devices. Although the use of Computational Fluid Dynamics (CFD) in hemolysis prediction studies provides convenience and has promising potential, it is an extremely challenging process. Hemolysis predictions with CFD depend on the mesh, implementation method, coefficient set, and scalar-shear-stress model. To this end, an attempt was made to find the combination that would provide the most accurate result in hemolysis prediction with the commonly cited power-law based hemolysis model. In the hemolysis predictions conducted using CFD on the Food and Drug Administration (FDA) benchmark blood pump, 3 different scalar-shear-stress models, and 5 different coefficient sets with the power-law based hemolysis model were used. Also, a mesh independence test based on hemolysis and pressure head was performed. The pressure head results of CFD simulations were compared with published pressure head of the FDA benchmark blood pump and a good agreement was observed. In addition, results of CFD-hemolysis predictions which are conducted with scalar-shear-stress model and coefficient set combinations were compared with experimental hemolysis data at three operating conditions such as 6-7 L/min flow rates at 3500 rpm rotational speeds and 6 L/min at 2500 rpm. One of the combinations of the scalar-shear-stress model and the coefficient set was found to be within the error limits of the experimental measurements, while all other combinations overestimated hemolysis.</p>","PeriodicalId":15177,"journal":{"name":"Journal of Artificial Organs","volume":null,"pages":null},"PeriodicalIF":1.1000,"publicationDate":"2024-08-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Computational fluid dynamics simulating of the FDA benchmark blood pump with different coefficient sets and scaler shear stress models used in the power-law hemolysis model.\",\"authors\":\"Ahmet Onder, Omer Incebay, Rafet Yapici\",\"doi\":\"10.1007/s10047-024-01468-6\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Hemolysis is the most important issue to consider in the design and optimization of blood-contacting devices. Although the use of Computational Fluid Dynamics (CFD) in hemolysis prediction studies provides convenience and has promising potential, it is an extremely challenging process. Hemolysis predictions with CFD depend on the mesh, implementation method, coefficient set, and scalar-shear-stress model. To this end, an attempt was made to find the combination that would provide the most accurate result in hemolysis prediction with the commonly cited power-law based hemolysis model. In the hemolysis predictions conducted using CFD on the Food and Drug Administration (FDA) benchmark blood pump, 3 different scalar-shear-stress models, and 5 different coefficient sets with the power-law based hemolysis model were used. Also, a mesh independence test based on hemolysis and pressure head was performed. The pressure head results of CFD simulations were compared with published pressure head of the FDA benchmark blood pump and a good agreement was observed. In addition, results of CFD-hemolysis predictions which are conducted with scalar-shear-stress model and coefficient set combinations were compared with experimental hemolysis data at three operating conditions such as 6-7 L/min flow rates at 3500 rpm rotational speeds and 6 L/min at 2500 rpm. One of the combinations of the scalar-shear-stress model and the coefficient set was found to be within the error limits of the experimental measurements, while all other combinations overestimated hemolysis.</p>\",\"PeriodicalId\":15177,\"journal\":{\"name\":\"Journal of Artificial Organs\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2024-08-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Artificial Organs\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1007/s10047-024-01468-6\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Artificial Organs","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1007/s10047-024-01468-6","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

溶血是设计和优化血液接触设备时需要考虑的最重要问题。尽管在溶血预测研究中使用计算流体动力学(CFD)提供了便利并具有广阔的前景,但这是一个极具挑战性的过程。利用 CFD 进行溶血预测取决于网格、实施方法、系数集和标量剪切应力模型。为此,我们尝试找到一种组合,利用通常引用的基于幂律的溶血模型进行溶血预测,从而获得最准确的结果。在使用 CFD 对食品与药物管理局 (FDA) 基准血泵进行溶血预测时,使用了 3 种不同的标量剪切应力模型和 5 种不同的系数集,以及基于幂律的溶血模型。此外,还进行了基于溶血和压头的网格独立性测试。将 CFD 模拟的压头结果与已公布的 FDA 基准血泵的压头结果进行了比较,结果显示两者吻合良好。此外,采用标量-剪切应力模型和系数集组合进行的 CFD 溶血预测结果与三种工作条件下的实验溶血数据进行了比较,如 3500 转/分钟转速下 6-7 升/分钟的流量和 2500 转/分钟转速下 6 升/分钟的流量。结果发现,其中一个标量-剪切应力模型和系数组合在实验测量误差范围之内,而所有其他组合都高估了溶血量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Computational fluid dynamics simulating of the FDA benchmark blood pump with different coefficient sets and scaler shear stress models used in the power-law hemolysis model.

Hemolysis is the most important issue to consider in the design and optimization of blood-contacting devices. Although the use of Computational Fluid Dynamics (CFD) in hemolysis prediction studies provides convenience and has promising potential, it is an extremely challenging process. Hemolysis predictions with CFD depend on the mesh, implementation method, coefficient set, and scalar-shear-stress model. To this end, an attempt was made to find the combination that would provide the most accurate result in hemolysis prediction with the commonly cited power-law based hemolysis model. In the hemolysis predictions conducted using CFD on the Food and Drug Administration (FDA) benchmark blood pump, 3 different scalar-shear-stress models, and 5 different coefficient sets with the power-law based hemolysis model were used. Also, a mesh independence test based on hemolysis and pressure head was performed. The pressure head results of CFD simulations were compared with published pressure head of the FDA benchmark blood pump and a good agreement was observed. In addition, results of CFD-hemolysis predictions which are conducted with scalar-shear-stress model and coefficient set combinations were compared with experimental hemolysis data at three operating conditions such as 6-7 L/min flow rates at 3500 rpm rotational speeds and 6 L/min at 2500 rpm. One of the combinations of the scalar-shear-stress model and the coefficient set was found to be within the error limits of the experimental measurements, while all other combinations overestimated hemolysis.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Artificial Organs
Journal of Artificial Organs 医学-工程:生物医学
CiteScore
2.80
自引率
15.40%
发文量
68
审稿时长
6-12 weeks
期刊介绍: The aim of the Journal of Artificial Organs is to introduce to colleagues worldwide a broad spectrum of important new achievements in the field of artificial organs, ranging from fundamental research to clinical applications. The scope of the Journal of Artificial Organs encompasses but is not restricted to blood purification, cardiovascular intervention, biomaterials, and artificial metabolic organs. Additionally, the journal will cover technical and industrial innovations. Membership in the Japanese Society for Artificial Organs is not a prerequisite for submission.
期刊最新文献
Exploring the role of plasmapheresis prior to thyroidectomy in managing thyrotoxicosis: a comprehensive scoping review. A compact machine perfusion device for whole blood perfusion in isolated rat liver. Advanced roll porous scaffold 3D bioprinting technology. Impact of the spleen size on short-term prognosis in patients with cardiogenic shock receiving Impella-incorporated temporary mechanical circulatory support Axial-flow polymer bridge pump with hydrodynamic bearings
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1