设计一种基于伏安法测量人绒毛膜促性腺激素的电化学生物传感器

IF 1.3 Q4 ENGINEERING, BIOMEDICAL Journal of Medical Signals & Sensors Pub Date : 2024-07-25 eCollection Date: 2024-01-01 DOI:10.4103/jmss.jmss_64_23
Mohammad Samare-Najaf, Amirreza Dehghanian, Gholamreza Asadikaram, Maryam Mohamadi, Morteza Jafarinia, Amir Savardashtaki, Afrooz Afshari, Sina Vakili
{"title":"设计一种基于伏安法测量人绒毛膜促性腺激素的电化学生物传感器","authors":"Mohammad Samare-Najaf, Amirreza Dehghanian, Gholamreza Asadikaram, Maryam Mohamadi, Morteza Jafarinia, Amir Savardashtaki, Afrooz Afshari, Sina Vakili","doi":"10.4103/jmss.jmss_64_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Human chorionic gonadotropin (hCG) is a polypeptide hormone synthesized during pregnancy and is also upregulated in some pathologic conditions such as certain tumors. Its measurement is essential for diagnosing pregnancy and malignancies. Despite numerous attempts to introduce an accurate method capable of detecting hCG levels, several limitations are found in previous techniques. This study aimed to address the limitations of current hCG assay methods by designing an electrochemical biosensor based on voltammetry for the rapid, selective, inexpensive, and sensitive measurement of hCG levels.</p><p><strong>Methods: </strong>A carbon paste electrode was prepared and functionalized by para-aminobenzoic acid. The primary anti-β-hCG monoclonal antibody was immobilized on the electrode surface by activating the carboxyl groups with 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide and N-hydroxysuccinimide solutions. The study also involved optimizing parameters such as the time for primary antibody fixation, the time for hCG attachment, and the pH of the hydrogen peroxide solution to maximize the biosensor response. Different concentrations of hCG hormone were prepared and loaded on the electrode surface, the secondary antibody labeled with HRP enzyme was applied, thionine in phosphate-buffered saline solution was placed on the electrode surface, and the differential pulse electrical signal was recorded.</p><p><strong>Results: </strong>The linear range ranged from 5 to 100 mIU/ml, and the limit of detection was calculated as 0.11 mIU. The relative standard deviation was 3% and 2% for five repeated measurements of commercial standard samples with concentrations of 2 and 20 mIU/mL, respectively. The percent recovery was obtained from 98.3% to 101.5%.</p><p><strong>Conclusion: </strong>The sensor represents a promising advancement in hCG level measurement, offering a potential solution to overcome the existing limitations in current diagnostic strategies. Simple and inexpensive design, detecting hCG in its important clinical range during early pregnancy, and successful measurement of hCG in real serum samples are the advantages of this sensor.</p>","PeriodicalId":37680,"journal":{"name":"Journal of Medical Signals & Sensors","volume":"14 ","pages":"21"},"PeriodicalIF":1.3000,"publicationDate":"2024-07-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11373787/pdf/","citationCount":"0","resultStr":"{\"title\":\"Designing an Electrochemical Biosensor Based on Voltammetry for Measurement of Human Chorionic Gonadotropin.\",\"authors\":\"Mohammad Samare-Najaf, Amirreza Dehghanian, Gholamreza Asadikaram, Maryam Mohamadi, Morteza Jafarinia, Amir Savardashtaki, Afrooz Afshari, Sina Vakili\",\"doi\":\"10.4103/jmss.jmss_64_23\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Human chorionic gonadotropin (hCG) is a polypeptide hormone synthesized during pregnancy and is also upregulated in some pathologic conditions such as certain tumors. Its measurement is essential for diagnosing pregnancy and malignancies. Despite numerous attempts to introduce an accurate method capable of detecting hCG levels, several limitations are found in previous techniques. This study aimed to address the limitations of current hCG assay methods by designing an electrochemical biosensor based on voltammetry for the rapid, selective, inexpensive, and sensitive measurement of hCG levels.</p><p><strong>Methods: </strong>A carbon paste electrode was prepared and functionalized by para-aminobenzoic acid. The primary anti-β-hCG monoclonal antibody was immobilized on the electrode surface by activating the carboxyl groups with 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide and N-hydroxysuccinimide solutions. The study also involved optimizing parameters such as the time for primary antibody fixation, the time for hCG attachment, and the pH of the hydrogen peroxide solution to maximize the biosensor response. Different concentrations of hCG hormone were prepared and loaded on the electrode surface, the secondary antibody labeled with HRP enzyme was applied, thionine in phosphate-buffered saline solution was placed on the electrode surface, and the differential pulse electrical signal was recorded.</p><p><strong>Results: </strong>The linear range ranged from 5 to 100 mIU/ml, and the limit of detection was calculated as 0.11 mIU. The relative standard deviation was 3% and 2% for five repeated measurements of commercial standard samples with concentrations of 2 and 20 mIU/mL, respectively. The percent recovery was obtained from 98.3% to 101.5%.</p><p><strong>Conclusion: </strong>The sensor represents a promising advancement in hCG level measurement, offering a potential solution to overcome the existing limitations in current diagnostic strategies. Simple and inexpensive design, detecting hCG in its important clinical range during early pregnancy, and successful measurement of hCG in real serum samples are the advantages of this sensor.</p>\",\"PeriodicalId\":37680,\"journal\":{\"name\":\"Journal of Medical Signals & Sensors\",\"volume\":\"14 \",\"pages\":\"21\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2024-07-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11373787/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Signals & Sensors\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/jmss.jmss_64_23\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q4\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Signals & Sensors","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmss.jmss_64_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0

摘要

背景:人绒毛膜促性腺激素(hCG人绒毛膜促性腺激素(hCG)是一种在妊娠期间合成的多肽激素,在某些病理情况下(如某些肿瘤)也会上调。测量 hCG 对诊断妊娠和恶性肿瘤至关重要。尽管人们曾多次尝试引入一种能够检测 hCG 水平的精确方法,但发现以往的技术存在一些局限性。本研究旨在通过设计一种基于伏安法的电化学生物传感器来快速、选择性、廉价、灵敏地测量 hCG 水平,从而解决目前 hCG 检测方法的局限性:方法:制备碳糊电极并用对氨基苯甲酸进行功能化。方法:制备了碳糊电极,并用对氨基苯甲酸对其进行了功能化处理。通过用 1-乙基-3-(3-二甲氨基丙基)碳二亚胺和 N-羟基琥珀酰亚胺溶液激活羧基,将抗β-hCG 单克隆抗体固定在电极表面。研究还涉及优化参数,如一抗固定时间、hCG 附着时间和过氧化氢溶液的 pH 值,以最大限度地提高生物传感器的响应。制备不同浓度的 hCG 激素并将其装载在电极表面,涂上用 HRP 酶标记的第二抗体,将磷酸盐缓冲盐溶液中的亚硫酸置于电极表面,记录差分脉冲电信号:线性范围为 5 至 100 mIU/ml,检测限为 0.11 mIU。对浓度分别为 2 mIU/mL 和 20 mIU/mL 的商业标准样品重复测量 5 次,相对标准偏差分别为 3% 和 2%。回收率为 98.3% 至 101.5%:该传感器是测量 hCG 水平的一大进步,为克服现有诊断策略的局限性提供了潜在的解决方案。该传感器设计简单、成本低廉,能在临床上重要的早孕期范围内检测 hCG,并能成功测量真实血清样本中的 hCG,这些都是它的优点。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Designing an Electrochemical Biosensor Based on Voltammetry for Measurement of Human Chorionic Gonadotropin.

Background: Human chorionic gonadotropin (hCG) is a polypeptide hormone synthesized during pregnancy and is also upregulated in some pathologic conditions such as certain tumors. Its measurement is essential for diagnosing pregnancy and malignancies. Despite numerous attempts to introduce an accurate method capable of detecting hCG levels, several limitations are found in previous techniques. This study aimed to address the limitations of current hCG assay methods by designing an electrochemical biosensor based on voltammetry for the rapid, selective, inexpensive, and sensitive measurement of hCG levels.

Methods: A carbon paste electrode was prepared and functionalized by para-aminobenzoic acid. The primary anti-β-hCG monoclonal antibody was immobilized on the electrode surface by activating the carboxyl groups with 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide and N-hydroxysuccinimide solutions. The study also involved optimizing parameters such as the time for primary antibody fixation, the time for hCG attachment, and the pH of the hydrogen peroxide solution to maximize the biosensor response. Different concentrations of hCG hormone were prepared and loaded on the electrode surface, the secondary antibody labeled with HRP enzyme was applied, thionine in phosphate-buffered saline solution was placed on the electrode surface, and the differential pulse electrical signal was recorded.

Results: The linear range ranged from 5 to 100 mIU/ml, and the limit of detection was calculated as 0.11 mIU. The relative standard deviation was 3% and 2% for five repeated measurements of commercial standard samples with concentrations of 2 and 20 mIU/mL, respectively. The percent recovery was obtained from 98.3% to 101.5%.

Conclusion: The sensor represents a promising advancement in hCG level measurement, offering a potential solution to overcome the existing limitations in current diagnostic strategies. Simple and inexpensive design, detecting hCG in its important clinical range during early pregnancy, and successful measurement of hCG in real serum samples are the advantages of this sensor.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Signals & Sensors
Journal of Medical Signals & Sensors ENGINEERING, BIOMEDICAL-
CiteScore
2.30
自引率
0.00%
发文量
53
审稿时长
33 weeks
期刊介绍: JMSS is an interdisciplinary journal that incorporates all aspects of the biomedical engineering including bioelectrics, bioinformatics, medical physics, health technology assessment, etc. Subject areas covered by the journal include: - Bioelectric: Bioinstruments Biosensors Modeling Biomedical signal processing Medical image analysis and processing Medical imaging devices Control of biological systems Neuromuscular systems Cognitive sciences Telemedicine Robotic Medical ultrasonography Bioelectromagnetics Electrophysiology Cell tracking - Bioinformatics and medical informatics: Analysis of biological data Data mining Stochastic modeling Computational genomics Artificial intelligence & fuzzy Applications Medical softwares Bioalgorithms Electronic health - Biophysics and medical physics: Computed tomography Radiation therapy Laser therapy - Education in biomedical engineering - Health technology assessment - Standard in biomedical engineering.
期刊最新文献
Computed Tomography Scan and Clinical-based Complete Response Prediction in Locally Advanced Rectal Cancer after Neoadjuvant Chemoradiotherapy: A Machine Learning Approach. Radiomics based Machine Learning Models for Classification of Prostate Cancer Grade Groups from Multi Parametric MRI Images. Advancing Proton Therapy: A Review of Geant4 Simulation for Enhanced Planning and Optimization in Hadron Therapy. Evaluation of Dose Calculation Algorithms Accuracy for ISOgray Treatment Planning System in Motorized Wedged Treatment Fields. Diagnosis of Autism in Children Based on their Gait Pattern and Movement Signs Using the Kinect Sensor.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1