离子镁和离子钙在体内和体外的关系。

IF 2.5 3区 医学 Q2 MEDICAL LABORATORY TECHNOLOGY Clinical biochemistry Pub Date : 2024-09-04 DOI:10.1016/j.clinbiochem.2024.110815
{"title":"离子镁和离子钙在体内和体外的关系。","authors":"","doi":"10.1016/j.clinbiochem.2024.110815","DOIUrl":null,"url":null,"abstract":"<div><h3>Objectives</h3><p>The objective of this study was to determine the in vivo correlation of ionized magnesium (iMg) with ionized calcium (iCa), total calcium, albumin and pH. In addition, the analytical interference of iCa on iMg measurement on the Stat Profile Prime Plus (Nova Biomedical) and vice versa was defined.</p></div><div><h3>Methods</h3><p>In vivo correlation of iCa, iMg and pH was studied in 238 paired blood gas samples of 109 different patients admitted to the intensive care unit. Albumin and total magnesium (tMg) were measured in heparinized plasma samples. Measurement of iMg was performed with the ion selective magnesium electrode (ISE) of the Stat Profile Prime Plus (Nova Biomedical) and iCa and pH were measured with a Rapid Point 500 blood gas analyzer (Siemens). Albumin, total calcium and total magnesium were analyzed with a Siemens Atellica CH. Analytical interference of iCa with iMg and vice versa was investigated using unbuffered saline solutions.</p></div><div><h3>Results</h3><p>In the studied patient population, no significant correlations were observed between iMg and iCa, albumin, and pH. An inverse relationship was observed between iCa and Mg-ISE. For every 0.1 mmol/L change in iCa concentration, the iMg concentration deviated by 0.01 mmol/L at an iMg concentration of 0.5 mmol/L and by 0.013 mmol/L at an iMg concentration of 1.0 mmol/L. The measurement of iCa was not affected by iMg.</p></div><div><h3>Conclusions</h3><p>In vivo, no correlation was observed between iMg with iCa, albumin and pH. Interference of iCa on iMg measurement was noted, with a maximum deviation of ±0.02 mmol/L iMg across the reference range of iCa (1.15–1.32 mmol/L). Additionally, the iCa measurement was not affected by the iMg concentration.</p></div>","PeriodicalId":10172,"journal":{"name":"Clinical biochemistry","volume":null,"pages":null},"PeriodicalIF":2.5000,"publicationDate":"2024-09-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S0009912024001097/pdfft?md5=cfeee0d962ed4277d52c9c1c630cb302&pid=1-s2.0-S0009912024001097-main.pdf","citationCount":"0","resultStr":"{\"title\":\"In vivo and in vitro relationship between ionized magnesium and ionized calcium\",\"authors\":\"\",\"doi\":\"10.1016/j.clinbiochem.2024.110815\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Objectives</h3><p>The objective of this study was to determine the in vivo correlation of ionized magnesium (iMg) with ionized calcium (iCa), total calcium, albumin and pH. In addition, the analytical interference of iCa on iMg measurement on the Stat Profile Prime Plus (Nova Biomedical) and vice versa was defined.</p></div><div><h3>Methods</h3><p>In vivo correlation of iCa, iMg and pH was studied in 238 paired blood gas samples of 109 different patients admitted to the intensive care unit. Albumin and total magnesium (tMg) were measured in heparinized plasma samples. Measurement of iMg was performed with the ion selective magnesium electrode (ISE) of the Stat Profile Prime Plus (Nova Biomedical) and iCa and pH were measured with a Rapid Point 500 blood gas analyzer (Siemens). Albumin, total calcium and total magnesium were analyzed with a Siemens Atellica CH. Analytical interference of iCa with iMg and vice versa was investigated using unbuffered saline solutions.</p></div><div><h3>Results</h3><p>In the studied patient population, no significant correlations were observed between iMg and iCa, albumin, and pH. An inverse relationship was observed between iCa and Mg-ISE. For every 0.1 mmol/L change in iCa concentration, the iMg concentration deviated by 0.01 mmol/L at an iMg concentration of 0.5 mmol/L and by 0.013 mmol/L at an iMg concentration of 1.0 mmol/L. The measurement of iCa was not affected by iMg.</p></div><div><h3>Conclusions</h3><p>In vivo, no correlation was observed between iMg with iCa, albumin and pH. Interference of iCa on iMg measurement was noted, with a maximum deviation of ±0.02 mmol/L iMg across the reference range of iCa (1.15–1.32 mmol/L). Additionally, the iCa measurement was not affected by the iMg concentration.</p></div>\",\"PeriodicalId\":10172,\"journal\":{\"name\":\"Clinical biochemistry\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.5000,\"publicationDate\":\"2024-09-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S0009912024001097/pdfft?md5=cfeee0d962ed4277d52c9c1c630cb302&pid=1-s2.0-S0009912024001097-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Clinical biochemistry\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0009912024001097\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical biochemistry","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0009912024001097","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

研究目的本研究旨在确定体内离子化镁(iMg)与离子化钙(iCa)、总钙、白蛋白和 pH 值的相关性。此外,还确定了 iCa 对 Stat Profile Prime Plus(Nova Biomedical)上 iMg 测量的分析干扰,反之亦然:方法:研究了重症监护室 109 名不同患者的 238 份配对血气样本中 iCa、iMg 和 pH 的体内相关性。在肝素化血浆样本中测量了白蛋白和总镁(tMg)。iMg 用 Stat Profile Prime Plus(Nova Biomedical)的离子选择性镁电极(ISE)测量,iCa 和 pH 用 Rapid Point 500 血气分析仪(西门子)测量。使用西门子 Atellica CH 分析白蛋白、总钙和总镁。使用无缓冲生理盐水溶液研究了 iCa 对 iMg 的分析干扰,以及 iMg 对 iCa 的分析干扰:结果:在所研究的患者群体中,未观察到 iMg 与 iCa、白蛋白和 pH 值之间存在明显的相关性。iCa 和 Mg-ISE 之间呈反比关系。iCa 浓度每变化 0.1 毫摩尔/升,iMg 浓度就会偏离 0.01 毫摩尔/升(iMg 浓度为 0.5 毫摩尔/升);iMg 浓度为 1.0 毫摩尔/升,iMg 浓度就会偏离 0.013 毫摩尔/升。iCa 的测量不受 iMg 的影响:在体内,iMg 与 iCa、白蛋白和 pH 之间没有相关性。注意到 iCa 对 iMg 测量的干扰,在 iCa 的参考范围(1.15-1.32 mmol/L)内,iMg 的最大偏差为 ±0.02 mmol/L。此外,iCa 测量不受 iMg 浓度的影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
In vivo and in vitro relationship between ionized magnesium and ionized calcium

Objectives

The objective of this study was to determine the in vivo correlation of ionized magnesium (iMg) with ionized calcium (iCa), total calcium, albumin and pH. In addition, the analytical interference of iCa on iMg measurement on the Stat Profile Prime Plus (Nova Biomedical) and vice versa was defined.

Methods

In vivo correlation of iCa, iMg and pH was studied in 238 paired blood gas samples of 109 different patients admitted to the intensive care unit. Albumin and total magnesium (tMg) were measured in heparinized plasma samples. Measurement of iMg was performed with the ion selective magnesium electrode (ISE) of the Stat Profile Prime Plus (Nova Biomedical) and iCa and pH were measured with a Rapid Point 500 blood gas analyzer (Siemens). Albumin, total calcium and total magnesium were analyzed with a Siemens Atellica CH. Analytical interference of iCa with iMg and vice versa was investigated using unbuffered saline solutions.

Results

In the studied patient population, no significant correlations were observed between iMg and iCa, albumin, and pH. An inverse relationship was observed between iCa and Mg-ISE. For every 0.1 mmol/L change in iCa concentration, the iMg concentration deviated by 0.01 mmol/L at an iMg concentration of 0.5 mmol/L and by 0.013 mmol/L at an iMg concentration of 1.0 mmol/L. The measurement of iCa was not affected by iMg.

Conclusions

In vivo, no correlation was observed between iMg with iCa, albumin and pH. Interference of iCa on iMg measurement was noted, with a maximum deviation of ±0.02 mmol/L iMg across the reference range of iCa (1.15–1.32 mmol/L). Additionally, the iCa measurement was not affected by the iMg concentration.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Clinical biochemistry
Clinical biochemistry 医学-医学实验技术
CiteScore
5.10
自引率
0.00%
发文量
151
审稿时长
25 days
期刊介绍: Clinical Biochemistry publishes articles relating to clinical chemistry, molecular biology and genetics, therapeutic drug monitoring and toxicology, laboratory immunology and laboratory medicine in general, with the focus on analytical and clinical investigation of laboratory tests in humans used for diagnosis, prognosis, treatment and therapy, and monitoring of disease.
期刊最新文献
Deteriorated thiol-disulphide and oxidized-reduced glutathione status in blood in Alzheimer’s disease Editorial Board In vivo and in vitro relationship between ionized magnesium and ionized calcium Heart failure biomarkers and prediction of early left ventricle remodeling after acute coronary syndromes Serum glucose mediated association of serum lactate with acute kidney injury among AIS patients
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1