Ana Marija Kožuljević, Tomislav Bokulić, Darko Grošev, Zdenka Kuncic, Siddharth Parashari, Luka Pavelić, Mihael Makek
{"title":"研究 PET 成像系统测量偏振相关康普顿事件的空间分辨率","authors":"Ana Marija Kožuljević, Tomislav Bokulić, Darko Grošev, Zdenka Kuncic, Siddharth Parashari, Luka Pavelić, Mihael Makek","doi":"arxiv-2409.01238","DOIUrl":null,"url":null,"abstract":"Recent studies of positron emission tomography (PET) devices have shown that\nthe detection of polarization-correlated annihilation quanta can potentially\nreduce the background and creation of false lines of response (LORs) leading to\nimproved image quality. We developed a novel PET demonstrator system, capable\nof measuring correlated gamma photons with single-layer Compton polarimeters to\nexplore the potential of the method. We tested the system using sources with\nclinically relevant activities at the University Hospital Centre Zagreb. Here\nwe present, for the first time, the images of two Ge-68 line sources,\nreconstructed solely from the correlated annihilation events. The spatial\nresolution at two different diameters is determined and compared to the one\nobtained from events with photoelectric interaction.","PeriodicalId":501378,"journal":{"name":"arXiv - PHYS - Medical Physics","volume":"87 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-09-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Investigation of the spatial resolution of PET imaging system measuring polarization-correlated Compton events\",\"authors\":\"Ana Marija Kožuljević, Tomislav Bokulić, Darko Grošev, Zdenka Kuncic, Siddharth Parashari, Luka Pavelić, Mihael Makek\",\"doi\":\"arxiv-2409.01238\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Recent studies of positron emission tomography (PET) devices have shown that\\nthe detection of polarization-correlated annihilation quanta can potentially\\nreduce the background and creation of false lines of response (LORs) leading to\\nimproved image quality. We developed a novel PET demonstrator system, capable\\nof measuring correlated gamma photons with single-layer Compton polarimeters to\\nexplore the potential of the method. We tested the system using sources with\\nclinically relevant activities at the University Hospital Centre Zagreb. Here\\nwe present, for the first time, the images of two Ge-68 line sources,\\nreconstructed solely from the correlated annihilation events. The spatial\\nresolution at two different diameters is determined and compared to the one\\nobtained from events with photoelectric interaction.\",\"PeriodicalId\":501378,\"journal\":{\"name\":\"arXiv - PHYS - Medical Physics\",\"volume\":\"87 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-09-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"arXiv - PHYS - Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/arxiv-2409.01238\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"arXiv - PHYS - Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/arxiv-2409.01238","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
最近对正电子发射断层扫描(PET)设备的研究表明,偏振相关湮没量子的检测有可能减少背景和虚假响应线(LOR)的产生,从而提高图像质量。我们开发了一种新型 PET 演示系统,能够利用单层康普顿偏振计测量相关伽马光子,以探索该方法的潜力。我们使用萨格勒布大学医院中心的临床相关活动源对该系统进行了测试。在这里,我们首次展示了两个 Ge-68 线源的图像,它们完全是由相关湮灭事件重建的。我们确定了两种不同直径下的空间分辨率,并将其与光电相互作用事件获得的空间分辨率进行了比较。
Investigation of the spatial resolution of PET imaging system measuring polarization-correlated Compton events
Recent studies of positron emission tomography (PET) devices have shown that
the detection of polarization-correlated annihilation quanta can potentially
reduce the background and creation of false lines of response (LORs) leading to
improved image quality. We developed a novel PET demonstrator system, capable
of measuring correlated gamma photons with single-layer Compton polarimeters to
explore the potential of the method. We tested the system using sources with
clinically relevant activities at the University Hospital Centre Zagreb. Here
we present, for the first time, the images of two Ge-68 line sources,
reconstructed solely from the correlated annihilation events. The spatial
resolution at two different diameters is determined and compared to the one
obtained from events with photoelectric interaction.