{"title":"自适应光学视觉模拟器和哈特曼-沙克像差仪在接受过角膜屈光手术和未接受过角膜屈光手术的假性角膜病眼中的重复性。","authors":"Dylan McBee,Karim Kozhaya,Li Wang,Mitchell P Weikert,Douglas D Koch","doi":"10.3928/1081597x-20240718-02","DOIUrl":null,"url":null,"abstract":"PURPOSE\r\nTo evaluate the intrasession repeatability of wavefront aberrations obtained by a combined adaptive optics visual simulator and Hartman-Shack aberrometer in pseudophakic eyes with and without previous corneal refractive surgery.\r\n\r\nMETHODS\r\nThree consecutive measurements were performed in one eye of each individual. Total ocular aberrations were recorded up to the 5th Zernike order for a 4.5-mm pupil. Repeatability was assessed by calculating the within-subject standard deviation (Sw), the repeatability limit (R), and the intraclass correlation coefficient (ICC). Vector analysis was performed to assess astigmatism variability between scans.\r\n\r\nRESULTS\r\nThe study enrolled 32 normal individuals and 24 individuals with a history of refractive surgery. In normal and eyes that had previous refractive surgery, respectively, the Sw values were 0.155 and 0.176 diopters (D) for sphere and 0.184 and 0.265 D for cylinder. The Sw values for all 3rd order terms ranged from 0.037 to 0.047 µm in normal eyes and 0.044 to 0.063 µm in eyes that had previous refractive surgery. The Sw for primary spherical aberration was 0.020 µm in normal eyes and 0.026 µm in eyes that had previous refractive surgery. ICC values for measurements of astigmatism yielded larger variability (ICC = 0.751 and 0.879). However, both groups demonstrated excellent repeatability (ICC > 0.9) for root mean square higher order aberrations (RMS-HOA) and total RMS values.\r\n\r\nCONCLUSIONS\r\nIn pseudophakic eyes, the adaptive optics Hartmann-Shack device demonstrated acceptable repeatability for measurement of sphere and 3rd and 4th order HOAs with higher variability for astigmatism measurements, especially in eyes with a prior history of corneal refractive surgery. [J Refract Surg. 2024;40(9):e645-e653.].","PeriodicalId":16951,"journal":{"name":"Journal of refractive surgery","volume":"93 1","pages":"e645-e653"},"PeriodicalIF":2.9000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Repeatability of a Combined Adaptive Optics Visual Simulator and Hartman-Shack Aberrometer in Pseudophakic Eyes With and Without Previous Corneal Refractive Surgery.\",\"authors\":\"Dylan McBee,Karim Kozhaya,Li Wang,Mitchell P Weikert,Douglas D Koch\",\"doi\":\"10.3928/1081597x-20240718-02\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"PURPOSE\\r\\nTo evaluate the intrasession repeatability of wavefront aberrations obtained by a combined adaptive optics visual simulator and Hartman-Shack aberrometer in pseudophakic eyes with and without previous corneal refractive surgery.\\r\\n\\r\\nMETHODS\\r\\nThree consecutive measurements were performed in one eye of each individual. Total ocular aberrations were recorded up to the 5th Zernike order for a 4.5-mm pupil. Repeatability was assessed by calculating the within-subject standard deviation (Sw), the repeatability limit (R), and the intraclass correlation coefficient (ICC). Vector analysis was performed to assess astigmatism variability between scans.\\r\\n\\r\\nRESULTS\\r\\nThe study enrolled 32 normal individuals and 24 individuals with a history of refractive surgery. In normal and eyes that had previous refractive surgery, respectively, the Sw values were 0.155 and 0.176 diopters (D) for sphere and 0.184 and 0.265 D for cylinder. The Sw values for all 3rd order terms ranged from 0.037 to 0.047 µm in normal eyes and 0.044 to 0.063 µm in eyes that had previous refractive surgery. The Sw for primary spherical aberration was 0.020 µm in normal eyes and 0.026 µm in eyes that had previous refractive surgery. ICC values for measurements of astigmatism yielded larger variability (ICC = 0.751 and 0.879). However, both groups demonstrated excellent repeatability (ICC > 0.9) for root mean square higher order aberrations (RMS-HOA) and total RMS values.\\r\\n\\r\\nCONCLUSIONS\\r\\nIn pseudophakic eyes, the adaptive optics Hartmann-Shack device demonstrated acceptable repeatability for measurement of sphere and 3rd and 4th order HOAs with higher variability for astigmatism measurements, especially in eyes with a prior history of corneal refractive surgery. [J Refract Surg. 2024;40(9):e645-e653.].\",\"PeriodicalId\":16951,\"journal\":{\"name\":\"Journal of refractive surgery\",\"volume\":\"93 1\",\"pages\":\"e645-e653\"},\"PeriodicalIF\":2.9000,\"publicationDate\":\"2024-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of refractive surgery\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3928/1081597x-20240718-02\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of refractive surgery","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3928/1081597x-20240718-02","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
Repeatability of a Combined Adaptive Optics Visual Simulator and Hartman-Shack Aberrometer in Pseudophakic Eyes With and Without Previous Corneal Refractive Surgery.
PURPOSE
To evaluate the intrasession repeatability of wavefront aberrations obtained by a combined adaptive optics visual simulator and Hartman-Shack aberrometer in pseudophakic eyes with and without previous corneal refractive surgery.
METHODS
Three consecutive measurements were performed in one eye of each individual. Total ocular aberrations were recorded up to the 5th Zernike order for a 4.5-mm pupil. Repeatability was assessed by calculating the within-subject standard deviation (Sw), the repeatability limit (R), and the intraclass correlation coefficient (ICC). Vector analysis was performed to assess astigmatism variability between scans.
RESULTS
The study enrolled 32 normal individuals and 24 individuals with a history of refractive surgery. In normal and eyes that had previous refractive surgery, respectively, the Sw values were 0.155 and 0.176 diopters (D) for sphere and 0.184 and 0.265 D for cylinder. The Sw values for all 3rd order terms ranged from 0.037 to 0.047 µm in normal eyes and 0.044 to 0.063 µm in eyes that had previous refractive surgery. The Sw for primary spherical aberration was 0.020 µm in normal eyes and 0.026 µm in eyes that had previous refractive surgery. ICC values for measurements of astigmatism yielded larger variability (ICC = 0.751 and 0.879). However, both groups demonstrated excellent repeatability (ICC > 0.9) for root mean square higher order aberrations (RMS-HOA) and total RMS values.
CONCLUSIONS
In pseudophakic eyes, the adaptive optics Hartmann-Shack device demonstrated acceptable repeatability for measurement of sphere and 3rd and 4th order HOAs with higher variability for astigmatism measurements, especially in eyes with a prior history of corneal refractive surgery. [J Refract Surg. 2024;40(9):e645-e653.].
期刊介绍:
The Journal of Refractive Surgery, the official journal of the International Society of Refractive Surgery, a partner of the American Academy of Ophthalmology, has been a monthly peer-reviewed forum for original research, review, and evaluation of refractive and lens-based surgical procedures for more than 30 years. Practical, clinically valuable articles provide readers with the most up-to-date information regarding advances in the field of refractive surgery. Begin to explore the Journal and all of its great benefits such as:
• Columns including “Translational Science,” “Surgical Techniques,” and “Biomechanics”
• Supplemental videos and materials available for many articles
• Access to current articles, as well as several years of archived content
• Articles posted online just 2 months after acceptance.