全面调试锥形束 CT 成像环,用于治疗 HDR 妇科病人。

IF 1.7 4区 医学 Q4 ONCOLOGY Brachytherapy Pub Date : 2024-11-01 DOI:10.1016/j.brachy.2024.07.008
Cullen Boyle, Firas Mourtada, Rani Anne, Shuying Wan, Yingxuan Chen, Yevgeniy Vinogradskiy, Reza Taleei
{"title":"全面调试锥形束 CT 成像环,用于治疗 HDR 妇科病人。","authors":"Cullen Boyle,&nbsp;Firas Mourtada,&nbsp;Rani Anne,&nbsp;Shuying Wan,&nbsp;Yingxuan Chen,&nbsp;Yevgeniy Vinogradskiy,&nbsp;Reza Taleei","doi":"10.1016/j.brachy.2024.07.008","DOIUrl":null,"url":null,"abstract":"<div><h3>PURPOSE</h3><div>A new mobile cone beam computed tomography (CBCT) imaging ring (IRm, Elekta, v2.10.6, Veenendaal, Netherlands) has recently been proposed for brachytherapy to improve procedure efficiency. We describe the commissioning process and end-to-end tests for GYN HDR brachytherapy employing IRm CBCT imaging.</div></div><div><h3>MATERIALS AND METHODS</h3><div>Commissioning included imaging isocenter test, 3D image quality, 2D imaging quality, image dose, and tube characteristics. CIRS HDR GYN phantom and Venezia CT/MR gynecological applicator were used to perform end-to-end (E2E) tests and optimize workflow. Venezia applicator and four interstitial needles were inserted into the phantom and IRm CBCT images were acquired. Phantom and applicator were scanned with CT scanner (Siemens SOMATOM go.Open Pro) using department's pelvis imaging protocol. MR imaging was performed using 0.35T MR Linac TRUFI pulse sequence. CBCT images were registered to CT and MR using rigid registration to assess image quality and applicator geometry fidelity.</div></div><div><h3>RESULTS</h3><div>All physics tests passed within acceptance tolerances. Registration of CBCT images to MR and CT scans was acceptable for applicator placement. Applicator registration of CBCT images to CT demonstrated excellent agreement of most distal source dwell position (&lt;1 mm). Slice thickness was also measured to be 1.25 mm, within 0.5 mm of its nominal value.</div></div><div><h3>CONCLUSION</h3><div>Based on E2E and commissioning results, IRm is an appropriate tool for brachytherapy treatment planning. This study demonstrated good image quality in GYN phantom and Venezia applicator using the IRm. Distal source dwell position agreement between CBCT and CT was acceptable for clinical use.</div></div>","PeriodicalId":55334,"journal":{"name":"Brachytherapy","volume":"23 6","pages":"Pages 687-697"},"PeriodicalIF":1.7000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comprehensive commissioning of a cone beam CT imaging ring for treatment of HDR GYN patients\",\"authors\":\"Cullen Boyle,&nbsp;Firas Mourtada,&nbsp;Rani Anne,&nbsp;Shuying Wan,&nbsp;Yingxuan Chen,&nbsp;Yevgeniy Vinogradskiy,&nbsp;Reza Taleei\",\"doi\":\"10.1016/j.brachy.2024.07.008\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>PURPOSE</h3><div>A new mobile cone beam computed tomography (CBCT) imaging ring (IRm, Elekta, v2.10.6, Veenendaal, Netherlands) has recently been proposed for brachytherapy to improve procedure efficiency. We describe the commissioning process and end-to-end tests for GYN HDR brachytherapy employing IRm CBCT imaging.</div></div><div><h3>MATERIALS AND METHODS</h3><div>Commissioning included imaging isocenter test, 3D image quality, 2D imaging quality, image dose, and tube characteristics. CIRS HDR GYN phantom and Venezia CT/MR gynecological applicator were used to perform end-to-end (E2E) tests and optimize workflow. Venezia applicator and four interstitial needles were inserted into the phantom and IRm CBCT images were acquired. Phantom and applicator were scanned with CT scanner (Siemens SOMATOM go.Open Pro) using department's pelvis imaging protocol. MR imaging was performed using 0.35T MR Linac TRUFI pulse sequence. CBCT images were registered to CT and MR using rigid registration to assess image quality and applicator geometry fidelity.</div></div><div><h3>RESULTS</h3><div>All physics tests passed within acceptance tolerances. Registration of CBCT images to MR and CT scans was acceptable for applicator placement. Applicator registration of CBCT images to CT demonstrated excellent agreement of most distal source dwell position (&lt;1 mm). Slice thickness was also measured to be 1.25 mm, within 0.5 mm of its nominal value.</div></div><div><h3>CONCLUSION</h3><div>Based on E2E and commissioning results, IRm is an appropriate tool for brachytherapy treatment planning. This study demonstrated good image quality in GYN phantom and Venezia applicator using the IRm. Distal source dwell position agreement between CBCT and CT was acceptable for clinical use.</div></div>\",\"PeriodicalId\":55334,\"journal\":{\"name\":\"Brachytherapy\",\"volume\":\"23 6\",\"pages\":\"Pages 687-697\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Brachytherapy\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S153847212400117X\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Brachytherapy","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S153847212400117X","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的:最近提出了一种用于近距离放射治疗的新型移动锥形束计算机断层扫描(CBCT)成像环(IRm,Elekta,v2.10.6,Veenendaal,荷兰),以提高手术效率。我们介绍了采用 IRm CBCT 成像技术的妇科 HDR 近距离放射治疗的调试过程和端到端测试:调试包括成像等中心测试、3D 图像质量、2D 图像质量、图像剂量和管道特性。使用 CIRS HDR 妇科模型和 Venezia CT/MR 妇科涂抹器进行端到端(E2E)测试并优化工作流程。将 Venezia 施用器和四个间隙针插入模型,并获取 IRm CBCT 图像。使用 CT 扫描仪(西门子 SOMATOM go.Open Pro)对模型和敷贴器进行扫描,采用的是该部门的骨盆成像协议。使用 0.35T MR Linac TRUFI 脉冲序列进行磁共振成像。使用刚性配准将 CBCT 图像与 CT 和 MR 图像配准,以评估图像质量和涂抹器几何形状的保真度:结果:所有物理测试均在合格范围内。CBCT 图像与 MR 和 CT 扫描的配准对于涂抹器的放置是可以接受的。CBCT 图像与 CT 图像的涂抹器登记显示,最远距离的源驻留位置非常一致(结论:根据 E2E 和调试结果,IRm 是近距离放射治疗规划的合适工具。这项研究表明,使用 IRm 在 GYN 模型和 Venezia 治疗器上获得的图像质量很好。CBCT 和 CT 之间的远端放射源停留位置一致,可用于临床。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Comprehensive commissioning of a cone beam CT imaging ring for treatment of HDR GYN patients

PURPOSE

A new mobile cone beam computed tomography (CBCT) imaging ring (IRm, Elekta, v2.10.6, Veenendaal, Netherlands) has recently been proposed for brachytherapy to improve procedure efficiency. We describe the commissioning process and end-to-end tests for GYN HDR brachytherapy employing IRm CBCT imaging.

MATERIALS AND METHODS

Commissioning included imaging isocenter test, 3D image quality, 2D imaging quality, image dose, and tube characteristics. CIRS HDR GYN phantom and Venezia CT/MR gynecological applicator were used to perform end-to-end (E2E) tests and optimize workflow. Venezia applicator and four interstitial needles were inserted into the phantom and IRm CBCT images were acquired. Phantom and applicator were scanned with CT scanner (Siemens SOMATOM go.Open Pro) using department's pelvis imaging protocol. MR imaging was performed using 0.35T MR Linac TRUFI pulse sequence. CBCT images were registered to CT and MR using rigid registration to assess image quality and applicator geometry fidelity.

RESULTS

All physics tests passed within acceptance tolerances. Registration of CBCT images to MR and CT scans was acceptable for applicator placement. Applicator registration of CBCT images to CT demonstrated excellent agreement of most distal source dwell position (<1 mm). Slice thickness was also measured to be 1.25 mm, within 0.5 mm of its nominal value.

CONCLUSION

Based on E2E and commissioning results, IRm is an appropriate tool for brachytherapy treatment planning. This study demonstrated good image quality in GYN phantom and Venezia applicator using the IRm. Distal source dwell position agreement between CBCT and CT was acceptable for clinical use.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Brachytherapy
Brachytherapy 医学-核医学
CiteScore
3.40
自引率
21.10%
发文量
119
审稿时长
9.1 weeks
期刊介绍: Brachytherapy is an international and multidisciplinary journal that publishes original peer-reviewed articles and selected reviews on the techniques and clinical applications of interstitial and intracavitary radiation in the management of cancers. Laboratory and experimental research relevant to clinical practice is also included. Related disciplines include medical physics, medical oncology, and radiation oncology and radiology. Brachytherapy publishes technical advances, original articles, reviews, and point/counterpoint on controversial issues. Original articles that address any aspect of brachytherapy are invited. Letters to the Editor-in-Chief are encouraged.
期刊最新文献
Table of Contents Editorial Board Masthead Surgically targeted radiation therapy versus stereotactic radiation therapy: A dosimetric comparison for brain metastasis resection cavities Commissioning considerations for the Bravos high-dose-rate afterloader: Towards improving treatment delivery accuracy
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1