{"title":"一种用于便携式设备的自动咳嗽计数方法和系统构造。","authors":"Yixuan Wang, Kehaoyu Yang, Shaofeng Xu, Shuwang Rui, Jiaxing Xie, Juncheng Wang, Xin Wang","doi":"10.3389/fbioe.2024.1477694","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Cough is a common symptom of respiratory diseases, and prolonged monitoring of cough can help assist doctors in making judgments about patients' conditions, among which cough frequency is an indicator that characterizes the state of the patient's lungs. Therefore, the aim of this paper is to design an automatic cough counting system to monitor the number of coughs per minute for a long period of time.</p><p><strong>Methods: </strong>In this paper, a complete cough counting process is proposed, including denoising, segment extraction, eigenvalue calculation, recognition, and counting process; and a wearable automatic cough counting device containing acquisition and reception software. The design and construction of the algorithm is based on realistically captured cough-containing audio from 50 patients, combined with short-time features, and Meier cepstrum coefficients as features characterizing the cough.</p><p><strong>Results: </strong>The accuracy, sensitivity, specificity, and F1 score of the method were 93.24%, 97.58%, 86.97%, and 94.47%, respectively, with a Kappa value of 0.9209, an average counting error of 0.46 counts for a 60-s speech segment, and an average runtime of 2.80 ± 2.27 s.</p><p><strong>Discussion: </strong>This method improves the double threshold method in terms of the threshold and eigenvalues of the cough segments' sensitivity and has better performance in terms of accuracy, real-time performance, and computing speed, which can be applied to real-time cough counting and monitoring in small portable devices with limited computing power. The developed wearable portable automatic cough counting device and the accompanying host computer software application can realize the long-term monitoring of patients' coughing condition.</p>","PeriodicalId":12444,"journal":{"name":"Frontiers in Bioengineering and Biotechnology","volume":null,"pages":null},"PeriodicalIF":4.3000,"publicationDate":"2024-09-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11466865/pdf/","citationCount":"0","resultStr":"{\"title\":\"An automatic cough counting method and system construction for portable devices.\",\"authors\":\"Yixuan Wang, Kehaoyu Yang, Shaofeng Xu, Shuwang Rui, Jiaxing Xie, Juncheng Wang, Xin Wang\",\"doi\":\"10.3389/fbioe.2024.1477694\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Cough is a common symptom of respiratory diseases, and prolonged monitoring of cough can help assist doctors in making judgments about patients' conditions, among which cough frequency is an indicator that characterizes the state of the patient's lungs. Therefore, the aim of this paper is to design an automatic cough counting system to monitor the number of coughs per minute for a long period of time.</p><p><strong>Methods: </strong>In this paper, a complete cough counting process is proposed, including denoising, segment extraction, eigenvalue calculation, recognition, and counting process; and a wearable automatic cough counting device containing acquisition and reception software. The design and construction of the algorithm is based on realistically captured cough-containing audio from 50 patients, combined with short-time features, and Meier cepstrum coefficients as features characterizing the cough.</p><p><strong>Results: </strong>The accuracy, sensitivity, specificity, and F1 score of the method were 93.24%, 97.58%, 86.97%, and 94.47%, respectively, with a Kappa value of 0.9209, an average counting error of 0.46 counts for a 60-s speech segment, and an average runtime of 2.80 ± 2.27 s.</p><p><strong>Discussion: </strong>This method improves the double threshold method in terms of the threshold and eigenvalues of the cough segments' sensitivity and has better performance in terms of accuracy, real-time performance, and computing speed, which can be applied to real-time cough counting and monitoring in small portable devices with limited computing power. The developed wearable portable automatic cough counting device and the accompanying host computer software application can realize the long-term monitoring of patients' coughing condition.</p>\",\"PeriodicalId\":12444,\"journal\":{\"name\":\"Frontiers in Bioengineering and Biotechnology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":4.3000,\"publicationDate\":\"2024-09-27\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11466865/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Frontiers in Bioengineering and Biotechnology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.3389/fbioe.2024.1477694\",\"RegionNum\":3,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q1\",\"JCRName\":\"BIOTECHNOLOGY & APPLIED MICROBIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Frontiers in Bioengineering and Biotechnology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.3389/fbioe.2024.1477694","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q1","JCRName":"BIOTECHNOLOGY & APPLIED MICROBIOLOGY","Score":null,"Total":0}
An automatic cough counting method and system construction for portable devices.
Introduction: Cough is a common symptom of respiratory diseases, and prolonged monitoring of cough can help assist doctors in making judgments about patients' conditions, among which cough frequency is an indicator that characterizes the state of the patient's lungs. Therefore, the aim of this paper is to design an automatic cough counting system to monitor the number of coughs per minute for a long period of time.
Methods: In this paper, a complete cough counting process is proposed, including denoising, segment extraction, eigenvalue calculation, recognition, and counting process; and a wearable automatic cough counting device containing acquisition and reception software. The design and construction of the algorithm is based on realistically captured cough-containing audio from 50 patients, combined with short-time features, and Meier cepstrum coefficients as features characterizing the cough.
Results: The accuracy, sensitivity, specificity, and F1 score of the method were 93.24%, 97.58%, 86.97%, and 94.47%, respectively, with a Kappa value of 0.9209, an average counting error of 0.46 counts for a 60-s speech segment, and an average runtime of 2.80 ± 2.27 s.
Discussion: This method improves the double threshold method in terms of the threshold and eigenvalues of the cough segments' sensitivity and has better performance in terms of accuracy, real-time performance, and computing speed, which can be applied to real-time cough counting and monitoring in small portable devices with limited computing power. The developed wearable portable automatic cough counting device and the accompanying host computer software application can realize the long-term monitoring of patients' coughing condition.
期刊介绍:
The translation of new discoveries in medicine to clinical routine has never been easy. During the second half of the last century, thanks to the progress in chemistry, biochemistry and pharmacology, we have seen the development and the application of a large number of drugs and devices aimed at the treatment of symptoms, blocking unwanted pathways and, in the case of infectious diseases, fighting the micro-organisms responsible. However, we are facing, today, a dramatic change in the therapeutic approach to pathologies and diseases. Indeed, the challenge of the present and the next decade is to fully restore the physiological status of the diseased organism and to completely regenerate tissue and organs when they are so seriously affected that treatments cannot be limited to the repression of symptoms or to the repair of damage. This is being made possible thanks to the major developments made in basic cell and molecular biology, including stem cell science, growth factor delivery, gene isolation and transfection, the advances in bioengineering and nanotechnology, including development of new biomaterials, biofabrication technologies and use of bioreactors, and the big improvements in diagnostic tools and imaging of cells, tissues and organs.
In today`s world, an enhancement of communication between multidisciplinary experts, together with the promotion of joint projects and close collaborations among scientists, engineers, industry people, regulatory agencies and physicians are absolute requirements for the success of any attempt to develop and clinically apply a new biological therapy or an innovative device involving the collective use of biomaterials, cells and/or bioactive molecules. “Frontiers in Bioengineering and Biotechnology” aspires to be a forum for all people involved in the process by bridging the gap too often existing between a discovery in the basic sciences and its clinical application.