[基于配重校正法的超轻植入体磁感应位移力测试方法]。

Yong Yin, Jin Deng, Xiaomei Wu
{"title":"[基于配重校正法的超轻植入体磁感应位移力测试方法]。","authors":"Yong Yin, Jin Deng, Xiaomei Wu","doi":"10.12455/j.issn.1671-7104.230711","DOIUrl":null,"url":null,"abstract":"<p><p>ASTM (American Society for Testing and Materials) F2052-21 (Chinese corresponding standard titled as YY/T 0987.2) introduces a test method to measure the magnetically induced displacement force on medical devices in the magnetic resonance environment. This method has been widely accepted and utilized globally, which is intended for implantable medical devices that can be suspended from a string. The angular deflection of the string from the vertical is measured and thereby measuring out the magnitude of the magnetically induced deflection force relative to gravity. When applying this method on ultralight implants, it is practically almost impossible to use a suspension string that weighs less than 1% of the mass of the implant. That study proposes a method for measuring the magnetically induced displacement force on ultralight implants based on counterweight correction by using a string and clamp that exceed 1% of the weight of the implant. And then to measure the deflection angle according to the standard requirement, following that to correct the deflection angle by using the proportion of the mass of the implant under test to the total mass of the tested object, to obtain the measurement result. That study also provides theoretical calculations and experimental verification to the counterweight correction method, the results prove that the proposed method can effectively measure the magnetically induced displacement force on ultralight implants.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-09-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Magnetically Induced Displacement Force Test Method for Ultralight Implants Based on Counterweight Correction Method].\",\"authors\":\"Yong Yin, Jin Deng, Xiaomei Wu\",\"doi\":\"10.12455/j.issn.1671-7104.230711\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>ASTM (American Society for Testing and Materials) F2052-21 (Chinese corresponding standard titled as YY/T 0987.2) introduces a test method to measure the magnetically induced displacement force on medical devices in the magnetic resonance environment. This method has been widely accepted and utilized globally, which is intended for implantable medical devices that can be suspended from a string. The angular deflection of the string from the vertical is measured and thereby measuring out the magnitude of the magnetically induced deflection force relative to gravity. When applying this method on ultralight implants, it is practically almost impossible to use a suspension string that weighs less than 1% of the mass of the implant. That study proposes a method for measuring the magnetically induced displacement force on ultralight implants based on counterweight correction by using a string and clamp that exceed 1% of the weight of the implant. And then to measure the deflection angle according to the standard requirement, following that to correct the deflection angle by using the proportion of the mass of the implant under test to the total mass of the tested object, to obtain the measurement result. That study also provides theoretical calculations and experimental verification to the counterweight correction method, the results prove that the proposed method can effectively measure the magnetically induced displacement force on ultralight implants.</p>\",\"PeriodicalId\":52535,\"journal\":{\"name\":\"中国医疗器械杂志\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-09-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中国医疗器械杂志\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://doi.org/10.12455/j.issn.1671-7104.230711\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.230711","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

ASTM(美国材料与试验协会)F2052-21(中国相应标准的名称为 YY/T 0987.2)介绍了一种在磁共振环境中测量医疗器械磁感应位移力的测试方法。该方法已在全球范围内被广泛接受和使用,适用于可悬挂在绳索上的植入式医疗器械。测量绳索偏离垂直方向的角度,从而测出相对于重力的磁感应偏转力大小。将这种方法应用于超轻型植入物时,实际上几乎不可能使用重量小于植入物质量 1%的悬挂绳。该研究提出了一种基于配重校正的超轻植入体磁感应位移力测量方法,即使用超过植入体重量 1%的绳索和夹具。然后根据标准要求测量偏转角,再利用被测植入体质量占被测物体总质量的比例对偏转角进行修正,从而得出测量结果。该研究还对配重校正方法进行了理论计算和实验验证,结果证明所提出的方法可以有效测量超轻植入物的磁感应位移力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[Magnetically Induced Displacement Force Test Method for Ultralight Implants Based on Counterweight Correction Method].

ASTM (American Society for Testing and Materials) F2052-21 (Chinese corresponding standard titled as YY/T 0987.2) introduces a test method to measure the magnetically induced displacement force on medical devices in the magnetic resonance environment. This method has been widely accepted and utilized globally, which is intended for implantable medical devices that can be suspended from a string. The angular deflection of the string from the vertical is measured and thereby measuring out the magnitude of the magnetically induced deflection force relative to gravity. When applying this method on ultralight implants, it is practically almost impossible to use a suspension string that weighs less than 1% of the mass of the implant. That study proposes a method for measuring the magnetically induced displacement force on ultralight implants based on counterweight correction by using a string and clamp that exceed 1% of the weight of the implant. And then to measure the deflection angle according to the standard requirement, following that to correct the deflection angle by using the proportion of the mass of the implant under test to the total mass of the tested object, to obtain the measurement result. That study also provides theoretical calculations and experimental verification to the counterweight correction method, the results prove that the proposed method can effectively measure the magnetically induced displacement force on ultralight implants.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
中国医疗器械杂志
中国医疗器械杂志 Medicine-Medicine (all)
CiteScore
0.40
自引率
0.00%
发文量
8086
期刊介绍:
期刊最新文献
[12-Lead Holter Integrated with Sleep Monitoring Module]. [Bowel Sounds Detection Method Based on ResNet-BiLSTM and Attention Mechanism]. [Clinical Application of Equivalent Uniform Dose in Intensity-Modulated Rotational Radiotherapy Based on Eclipse TPS]. [Clinical Validation of a Prototype Smart Non-Invasive Pregnancy Glucose Monitor]. [Clinical Validation Study of Deep Learning-Generated Magnetic Resonance Images].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1