{"title":"一种利用阻力变化帮助针尖插入猴子硬膜外阻滞的仪器。","authors":"S T Young, K H Chan, C F Chen","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A new method is proposed to detect the entrance of a needle into the epidural space. Measuring the changes in electric resistance during the insertion of the needle through different tissues in the interspinal space can help anesthesiologists judge whether and when the needle has entered the epidural space. This preliminary, animal study showed that the proposed method performed quite satisfactorily. The electric resistance of the tissues in the interspinal space varied from 1.57 +/- 1.22 k omega to 2.93 +/- 1.14 k omega, and the resistance of the epidural space varied from 9.00 +/- 1.12 k omega to 14.64 +/- 1.25 k omega. The difference was statistically significant (p less than 0.001). This difference in the resistance can be used to verify placement of the tip of the needle in the epidural space and help anesthesiologists administer epidural anesthesia.</p>","PeriodicalId":76133,"journal":{"name":"Medical instrumentation","volume":"21 5","pages":"266-8"},"PeriodicalIF":0.0000,"publicationDate":"1987-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"An instrument using variation of resistance to aid in needle tip insertion in epidural block in monkeys.\",\"authors\":\"S T Young, K H Chan, C F Chen\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>A new method is proposed to detect the entrance of a needle into the epidural space. Measuring the changes in electric resistance during the insertion of the needle through different tissues in the interspinal space can help anesthesiologists judge whether and when the needle has entered the epidural space. This preliminary, animal study showed that the proposed method performed quite satisfactorily. The electric resistance of the tissues in the interspinal space varied from 1.57 +/- 1.22 k omega to 2.93 +/- 1.14 k omega, and the resistance of the epidural space varied from 9.00 +/- 1.12 k omega to 14.64 +/- 1.25 k omega. The difference was statistically significant (p less than 0.001). This difference in the resistance can be used to verify placement of the tip of the needle in the epidural space and help anesthesiologists administer epidural anesthesia.</p>\",\"PeriodicalId\":76133,\"journal\":{\"name\":\"Medical instrumentation\",\"volume\":\"21 5\",\"pages\":\"266-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1987-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical instrumentation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical instrumentation","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
提出了一种检测针头进入硬膜外腔的新方法。测量针头穿过脊髓间隙不同组织时的电阻变化,可以帮助麻醉师判断针头是否以及何时进入硬膜外间隙。这项初步的动物研究表明,所提出的方法效果相当令人满意。脊髓间隙组织的电阻范围为1.57 +/- 1.22 k ω ~ 2.93 +/- 1.14 k ω,硬膜外间隙组织的电阻范围为9.00 +/- 1.12 k ω ~ 14.64 +/- 1.25 k ω。差异有统计学意义(p < 0.001)。这种阻力的差异可以用来验证针尖在硬膜外间隙的位置,并帮助麻醉师进行硬膜外麻醉。
An instrument using variation of resistance to aid in needle tip insertion in epidural block in monkeys.
A new method is proposed to detect the entrance of a needle into the epidural space. Measuring the changes in electric resistance during the insertion of the needle through different tissues in the interspinal space can help anesthesiologists judge whether and when the needle has entered the epidural space. This preliminary, animal study showed that the proposed method performed quite satisfactorily. The electric resistance of the tissues in the interspinal space varied from 1.57 +/- 1.22 k omega to 2.93 +/- 1.14 k omega, and the resistance of the epidural space varied from 9.00 +/- 1.12 k omega to 14.64 +/- 1.25 k omega. The difference was statistically significant (p less than 0.001). This difference in the resistance can be used to verify placement of the tip of the needle in the epidural space and help anesthesiologists administer epidural anesthesia.