PL02 演讲时间:下午 1:45

IF 1.7 4区 医学 Q4 ONCOLOGY Brachytherapy Pub Date : 2024-10-25 DOI:10.1016/j.brachy.2024.08.060
Christopher Deufel Ph.D., Eric Brost Ph.D., Justine Dupere Ph.D., Ivy A. Petersen M.D., Michael G. Haddock M.D., Allison E. Garda M.D.
{"title":"PL02 演讲时间:下午 1:45","authors":"Christopher Deufel Ph.D.,&nbsp;Eric Brost Ph.D.,&nbsp;Justine Dupere Ph.D.,&nbsp;Ivy A. Petersen M.D.,&nbsp;Michael G. Haddock M.D.,&nbsp;Allison E. Garda M.D.","doi":"10.1016/j.brachy.2024.08.060","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>To design, construct, and evaluate a system for assisted placement of brachytherapy applicators using electromagnetic tracking (EMT) technology that has been registered to CT or MRI images. The system provides real-time localization of needles during the insertion process, a three-dimensional display of planned needle sites, visibility of the anatomy and needle position during placement, and reference tracking to account for generator or target anatomy shifts. Such a system might be used to reduce brachytherapy procedure times, improve correspondence between intended and actual needle positions, or decrease the trainee learning curve. The system is notable for the following features: 1) Real-time visual and quantitative feedback of needle placement with respect to the underlying anatomy, as visualized by MRI or CT image, without continuous or repeated imaging 2) Pre-planning capability with a graphical overlay of target needle trajectories 3) Reference tracking to account for electromagnetic field generator or target anatomy shifts 4) DICOM-coordinate digital reconstruction of applicator locations for treatment planning and/or pre-treatment quality assurance 5) Compatibility with standard brachytherapy workflows including fixed table CT and MRI systems, procedures within or outside of a brachy suite, and insertion of needles in dorsal lithotomy position 6) Consists of commercially available EMT technology components</div></div><div><h3>Methods</h3><div>The system was constructed using an Aurora (Northern Digital Instruments, Waterloo, ON) planar 20 × 20 cm<sup>2</sup> field generator (EFG), System Control Unit, Sensor Interface Unit, and 6DOF and 5DOF Flextube sensor tools. The graphical user interface was written as a Matlab application with native functions and toolboxes. EMT-to-DICOM registration was based upon intracavitary applicators (e.g., tandem and ovoids), placed prior to imaging and digitized automatically using thresholding methods. The EFG was positioned above the pelvis (Figure 1A), EMT sensors were translated through the tandem and ovoid channels, and the EMT system was registered to the DICOM image set using an iterative closest-point algorithm. Next, a 5DOF EMT sensor was loaded into the distal inner lumen of a brachytherapy needle for placement. The system display provides axial, coronal, sagittal, and 3D-volumetric CT/MRI views. Proof-of concept and system accuracy were evaluated in phantom and human cadaver by comparing EM-tracked needle positions with ground-truth, post-implant CTs.</div></div><div><h3>Results</h3><div>Proof of concept was demonstrated for EMT-assisted placement of brachytherapy needles in a realistic clinical environment and on a brachy suite CT table. Figure 1B provides an example of how a pre-planned needle location (blue) can be visualized alongside real-time needle placement (red) to provide feedback to the user. The left-hand panel in Figure 1B shows an initial attempt where a guided needle is not in the intended location. The right-hand panel shows placement after readjustment. Accuracy in phantom (mean ± standard deviation) was 0.76 ± 0.13mm for needle tips placed up to 75mm from the tandem/ovoids and 0.52 ± 0.27mm for needle shafts at distances up to 100mm from the tandem/ovoids. Figure 1C illustrates the human cadaver EM-tracked needle positions and manually digitized positions from a post-implant CT scan. Tip and shaft accuracies were 0.77 ± 0.14mm and 0.40 ± 0.21mm, respectively.</div></div><div><h3>Conclusion</h3><div>An EMT-based guidance system provided sub-millimeter accuracy for the placement of brachytherapy needles without repeated or continuous imaging. The technology can be used to reduce brachytherapy procedure times, improve the correspondence between intended and actual needle positions, or decrease the trainee learning curve.</div></div>","PeriodicalId":55334,"journal":{"name":"Brachytherapy","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-10-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"PL02 Presentation Time: 1:45 PM\",\"authors\":\"Christopher Deufel Ph.D.,&nbsp;Eric Brost Ph.D.,&nbsp;Justine Dupere Ph.D.,&nbsp;Ivy A. Petersen M.D.,&nbsp;Michael G. Haddock M.D.,&nbsp;Allison E. Garda M.D.\",\"doi\":\"10.1016/j.brachy.2024.08.060\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><div>To design, construct, and evaluate a system for assisted placement of brachytherapy applicators using electromagnetic tracking (EMT) technology that has been registered to CT or MRI images. The system provides real-time localization of needles during the insertion process, a three-dimensional display of planned needle sites, visibility of the anatomy and needle position during placement, and reference tracking to account for generator or target anatomy shifts. Such a system might be used to reduce brachytherapy procedure times, improve correspondence between intended and actual needle positions, or decrease the trainee learning curve. The system is notable for the following features: 1) Real-time visual and quantitative feedback of needle placement with respect to the underlying anatomy, as visualized by MRI or CT image, without continuous or repeated imaging 2) Pre-planning capability with a graphical overlay of target needle trajectories 3) Reference tracking to account for electromagnetic field generator or target anatomy shifts 4) DICOM-coordinate digital reconstruction of applicator locations for treatment planning and/or pre-treatment quality assurance 5) Compatibility with standard brachytherapy workflows including fixed table CT and MRI systems, procedures within or outside of a brachy suite, and insertion of needles in dorsal lithotomy position 6) Consists of commercially available EMT technology components</div></div><div><h3>Methods</h3><div>The system was constructed using an Aurora (Northern Digital Instruments, Waterloo, ON) planar 20 × 20 cm<sup>2</sup> field generator (EFG), System Control Unit, Sensor Interface Unit, and 6DOF and 5DOF Flextube sensor tools. The graphical user interface was written as a Matlab application with native functions and toolboxes. EMT-to-DICOM registration was based upon intracavitary applicators (e.g., tandem and ovoids), placed prior to imaging and digitized automatically using thresholding methods. The EFG was positioned above the pelvis (Figure 1A), EMT sensors were translated through the tandem and ovoid channels, and the EMT system was registered to the DICOM image set using an iterative closest-point algorithm. Next, a 5DOF EMT sensor was loaded into the distal inner lumen of a brachytherapy needle for placement. The system display provides axial, coronal, sagittal, and 3D-volumetric CT/MRI views. Proof-of concept and system accuracy were evaluated in phantom and human cadaver by comparing EM-tracked needle positions with ground-truth, post-implant CTs.</div></div><div><h3>Results</h3><div>Proof of concept was demonstrated for EMT-assisted placement of brachytherapy needles in a realistic clinical environment and on a brachy suite CT table. Figure 1B provides an example of how a pre-planned needle location (blue) can be visualized alongside real-time needle placement (red) to provide feedback to the user. The left-hand panel in Figure 1B shows an initial attempt where a guided needle is not in the intended location. The right-hand panel shows placement after readjustment. Accuracy in phantom (mean ± standard deviation) was 0.76 ± 0.13mm for needle tips placed up to 75mm from the tandem/ovoids and 0.52 ± 0.27mm for needle shafts at distances up to 100mm from the tandem/ovoids. Figure 1C illustrates the human cadaver EM-tracked needle positions and manually digitized positions from a post-implant CT scan. Tip and shaft accuracies were 0.77 ± 0.14mm and 0.40 ± 0.21mm, respectively.</div></div><div><h3>Conclusion</h3><div>An EMT-based guidance system provided sub-millimeter accuracy for the placement of brachytherapy needles without repeated or continuous imaging. The technology can be used to reduce brachytherapy procedure times, improve the correspondence between intended and actual needle positions, or decrease the trainee learning curve.</div></div>\",\"PeriodicalId\":55334,\"journal\":{\"name\":\"Brachytherapy\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-10-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Brachytherapy\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S153847212400196X\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Brachytherapy","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S153847212400196X","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

目的设计、构建和评估一套系统,用于利用已在 CT 或 MRI 图像上注册的电磁跟踪 (EMT) 技术辅助放置近距离放射治疗应用器。该系统可在插入过程中实时定位穿刺针,提供计划穿刺针位置的三维显示,在穿刺过程中显示解剖结构和穿刺针位置,并提供参考跟踪以考虑发生器或目标解剖结构的移动。这种系统可用于缩短近距离放射治疗过程的时间,提高计划针位与实际针位之间的对应性,或降低受训者的学习曲线。该系统具有以下显著特点:1) 根据核磁共振成像或 CT 图像显示的下层解剖结构,对穿刺针位置进行实时可视化定量反馈,无需连续或重复成像 2) 通过目标穿刺针轨迹的图形叠加实现预规划功能 3) 参考跟踪,以考虑电磁场发生器或目标解剖结构的偏移 4) DICOM 坐标数字重建涂药器位置,用于治疗规划和/或治疗前质量保证 5) 与标准近距离放射治疗系统兼容。5) 与标准近距离放射治疗工作流程兼容,包括固定台 CT 和 MRI 系统、近距离放射治疗套间内外的手术以及在背侧碎石位置插入针头 6) 由市场上可用的 EMT 技术组件组成方法该系统由 Aurora(北方数字仪器公司,滑铁卢,安大略省)20 × 20 平方厘米平面场发生器 (EFG)、系统控制单元、传感器接口单元以及 6DOF 和 5DOF Flextube 传感器工具构成。图形用户界面是作为 Matlab 应用程序编写的,带有本地函数和工具箱。EMT到DICOM配准基于腔内涂抹器(如串联和卵圆形),在成像前放置,并使用阈值法自动数字化。EFG 位于骨盆上方(图 1A),EMT 传感器通过串联通道和卵形通道平移,EMT 系统使用迭代闭合点算法与 DICOM 图像集进行配准。接着,将 5DOF EMT 传感器装入近距离治疗针的远端内腔进行置放。系统显示屏提供轴向、冠状、矢状和三维容积 CT/MRI 视图。通过比较 EM 跟踪的针位置与地面实况、植入后 CT,在模型和人体尸体中对概念验证和系统准确性进行了评估。结果在真实的临床环境和近距离治疗室 CT 台上展示了 EMT 辅助近距离治疗针置放的概念验证。图 1B 举例说明了如何将预先计划的针头位置(蓝色)与实时针头放置(红色)同时显示出来,以便向用户提供反馈。图 1B 中的左侧面板显示的是最初的尝试,引导针没有到达预定位置。右侧面板显示的是重新调整后的位置。在模型中,针尖距离串联体/假体最远 75 毫米的精确度为 0.76 ± 0.13 毫米(平均值 ± 标准偏差),针轴距离串联体/假体最远 100 毫米的精确度为 0.52 ± 0.27 毫米(平均值 ± 标准偏差)。图 1C 展示了人体尸体 EM 跟踪针的位置和植入后 CT 扫描中手动数字化的位置。针尖和针杆的精确度分别为 0.77 ± 0.14 毫米和 0.40 ± 0.21 毫米。该技术可用于缩短近距离治疗手术时间,提高预期针头位置与实际针头位置之间的对应性,或降低受训者的学习曲线。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
PL02 Presentation Time: 1:45 PM

Purpose

To design, construct, and evaluate a system for assisted placement of brachytherapy applicators using electromagnetic tracking (EMT) technology that has been registered to CT or MRI images. The system provides real-time localization of needles during the insertion process, a three-dimensional display of planned needle sites, visibility of the anatomy and needle position during placement, and reference tracking to account for generator or target anatomy shifts. Such a system might be used to reduce brachytherapy procedure times, improve correspondence between intended and actual needle positions, or decrease the trainee learning curve. The system is notable for the following features: 1) Real-time visual and quantitative feedback of needle placement with respect to the underlying anatomy, as visualized by MRI or CT image, without continuous or repeated imaging 2) Pre-planning capability with a graphical overlay of target needle trajectories 3) Reference tracking to account for electromagnetic field generator or target anatomy shifts 4) DICOM-coordinate digital reconstruction of applicator locations for treatment planning and/or pre-treatment quality assurance 5) Compatibility with standard brachytherapy workflows including fixed table CT and MRI systems, procedures within or outside of a brachy suite, and insertion of needles in dorsal lithotomy position 6) Consists of commercially available EMT technology components

Methods

The system was constructed using an Aurora (Northern Digital Instruments, Waterloo, ON) planar 20 × 20 cm2 field generator (EFG), System Control Unit, Sensor Interface Unit, and 6DOF and 5DOF Flextube sensor tools. The graphical user interface was written as a Matlab application with native functions and toolboxes. EMT-to-DICOM registration was based upon intracavitary applicators (e.g., tandem and ovoids), placed prior to imaging and digitized automatically using thresholding methods. The EFG was positioned above the pelvis (Figure 1A), EMT sensors were translated through the tandem and ovoid channels, and the EMT system was registered to the DICOM image set using an iterative closest-point algorithm. Next, a 5DOF EMT sensor was loaded into the distal inner lumen of a brachytherapy needle for placement. The system display provides axial, coronal, sagittal, and 3D-volumetric CT/MRI views. Proof-of concept and system accuracy were evaluated in phantom and human cadaver by comparing EM-tracked needle positions with ground-truth, post-implant CTs.

Results

Proof of concept was demonstrated for EMT-assisted placement of brachytherapy needles in a realistic clinical environment and on a brachy suite CT table. Figure 1B provides an example of how a pre-planned needle location (blue) can be visualized alongside real-time needle placement (red) to provide feedback to the user. The left-hand panel in Figure 1B shows an initial attempt where a guided needle is not in the intended location. The right-hand panel shows placement after readjustment. Accuracy in phantom (mean ± standard deviation) was 0.76 ± 0.13mm for needle tips placed up to 75mm from the tandem/ovoids and 0.52 ± 0.27mm for needle shafts at distances up to 100mm from the tandem/ovoids. Figure 1C illustrates the human cadaver EM-tracked needle positions and manually digitized positions from a post-implant CT scan. Tip and shaft accuracies were 0.77 ± 0.14mm and 0.40 ± 0.21mm, respectively.

Conclusion

An EMT-based guidance system provided sub-millimeter accuracy for the placement of brachytherapy needles without repeated or continuous imaging. The technology can be used to reduce brachytherapy procedure times, improve the correspondence between intended and actual needle positions, or decrease the trainee learning curve.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Brachytherapy
Brachytherapy 医学-核医学
CiteScore
3.40
自引率
21.10%
发文量
119
审稿时长
9.1 weeks
期刊介绍: Brachytherapy is an international and multidisciplinary journal that publishes original peer-reviewed articles and selected reviews on the techniques and clinical applications of interstitial and intracavitary radiation in the management of cancers. Laboratory and experimental research relevant to clinical practice is also included. Related disciplines include medical physics, medical oncology, and radiation oncology and radiology. Brachytherapy publishes technical advances, original articles, reviews, and point/counterpoint on controversial issues. Original articles that address any aspect of brachytherapy are invited. Letters to the Editor-in-Chief are encouraged.
期刊最新文献
Editorial Board Masthead Table of Contents Thursday, July 11, 20244:00 PM - 5:00 PM PP01 Presentation Time: 4:00 PM MSOR12 Presentation Time: 5:55 PM
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1