[阻塞性肺通气功能障碍患者的脉冲振荡测量系统指数与常规肺功能测试的相关性]。

B Wei, K Zhang, Z Y Wang, B H Fu, X M Huang, Y T Chen, J P Zhao, J M Wang, M Xie, W Ni
{"title":"[阻塞性肺通气功能障碍患者的脉冲振荡测量系统指数与常规肺功能测试的相关性]。","authors":"B Wei, K Zhang, Z Y Wang, B H Fu, X M Huang, Y T Chen, J P Zhao, J M Wang, M Xie, W Ni","doi":"10.3760/cma.j.cn112138-20240410-00231","DOIUrl":null,"url":null,"abstract":"<p><p><b>Objective:</b> To investigate the correlation between impulse oscillometry system examination indicators and conventional pulmonary ventilation function. <b>Methods:</b> The pulmonary ventilation function data of 10 883 patients from January 1, 2020 to December 31, 2022 at Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology were included. The one-second rate [ratio of forced expiratory volume in the first second (FEV<sub>1</sub>) to forced vital capacity (FVC)] measured as a percentage of the predicted value was ≥92% for the control group (<i>n=</i>3 478) and <92% for the pulmonary obstruction group (<i>n=</i>7 405). The obstruction group was subdivided into five groups according to the degree of pulmonary dysfunction: mild group (<i>n=</i>3 938),moderate group (<i>n=</i>1 142),oderate-severe group (<i>n=</i>917),severe group (<i>n=</i>737),and extremely severe group (<i>n=</i>671). Conventional pulmonary ventilatory function FVC, FEV<sub>1</sub>, one-second rate, and forced expired flow at 50% of FVC (MEF50%), forced expired flow at 75% FVC (MEF25%), maximal mid-expiratory flow (MMEF), peak expiratory flow (PEF), and pulsed oscillation pulmonary function test were detected in both groups of patients. Impedance at 5 Hz (Z5) means total respiratory resistance, resistance at 5 Hz (R5) means total airway resistance, reactance at 5 Hz (X5) indicates the elastic recoil of the peripheral airways, and resistance at 20 Hz (R20) represents resistance of the central airways. R5-R20 reflects resistance in the small airways. Additionally, peripheral resistance (Rp), respiratory resonance frequency (Frex), and area under the reactance curve (Ax) were also measured. Correlation between the indicators of the two groups and the sensitivity and specificity of the impulse oscillometry system parameters for the diagnosis of obstructive pulmonary ventilation dysfunction were analyzed. <b>Results:</b> Pulmonary function force expiratory volume in the first second as a percentage of predicted value (FEV<sub>1</sub>%Pre) [80.10 (54.95,97.10)%],one-second rate [62.43(48.67, 67.02)%],MEF50% [1.33 (0.62,1.97)L/s],MEF25% [0.28 (0.17,0.41)L/s], MMEF [0.85 (0.43,1.29)L/s],and PEF [5.64 (3.73,7.50)]L/s in the obstruction group were significantly lower than those in the control group (<i>P<</i>0.05). The differences within the subgroups of the obstruction group were also significant (<i>P<</i>0.05). Pulsed oscillation Z5 [0.42 (0.33,0.55)kPa·L<sup>-1</sup>·s<sup>-1</sup>],Rp [0.25 (0.20,0.45)kPa·L<sup>-1</sup>·s<sup>-1</sup>], R5 [0.39 (0.31,0.49)kPa·L<sup>-1</sup>·s<sup>-1</sup>], R20 [0.28 (0.24,0.34)kPa·L<sup>-1</sup>·s<sup>-1</sup>], R5-R20 [0.09 (0.05,0.17)kPa·L<sup>-1</sup>·s<sup>-1</sup>],Frex [16.32 (13.07,20.84)Hz], and Ax [0.67 (0.28,1.64)] indices in the obstruction group were significantly higher than those in the control group. X5 [-0.14 (-0.23, -0.10)kPa·L<sup>-1</sup>·s<sup>-1</sup>] was significantly lower than that in the control group (<i>P<</i>0.05). Z5, Rp, X5, R5, R5-R20, Frex, and Ax were statistically significant between different degrees of obstruction in the obstruction group (<i>P<</i>0.05). The impulse oscillometry system parameters Z5, Rp, R5, R20, R5-20, Frex, and Ax were negatively correlated with the indices of conventional pulmonary ventilation (<i>r</i>=-0.21-0.68, <i>P<</i>0.05), and the parameter X5 was positively correlated with the indices of conventional pulmonary ventilation (<i>r=</i>0.41-0.68, <i>P<</i>0.05). The pulsed oscillation pulmonary function test parameters X5 (58.60%-95.68%) and Ax (57.08%-98.06%) presented the best sensitivity; X5 (86.29%-98.82%), Frex (86.69%-94.71%), and Ax (88.10%-98.53%) displayed the best specificity; and R20 presented the worst sensitivity and specificity. The sensitivity and specificity were slightly better in female patients than in male patients. <b>Conclusion:</b> The technical parameters of the impulse oscillometry system showed significant correlation with relevant indices of conventional pulmonary ventilation function detection. These well reflect the changes of different degrees of pulmonary ventilation function and have greater significance for reference in evaluating the degree of pulmonary function impairment.</p>","PeriodicalId":68309,"journal":{"name":"中华内科杂志","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Correlation of Impulse oscillometry system indices with conventional pulmonary function tests in patients with obstructive pulmonary ventilation dysfunction].\",\"authors\":\"B Wei, K Zhang, Z Y Wang, B H Fu, X M Huang, Y T Chen, J P Zhao, J M Wang, M Xie, W Ni\",\"doi\":\"10.3760/cma.j.cn112138-20240410-00231\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p><b>Objective:</b> To investigate the correlation between impulse oscillometry system examination indicators and conventional pulmonary ventilation function. <b>Methods:</b> The pulmonary ventilation function data of 10 883 patients from January 1, 2020 to December 31, 2022 at Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology were included. The one-second rate [ratio of forced expiratory volume in the first second (FEV<sub>1</sub>) to forced vital capacity (FVC)] measured as a percentage of the predicted value was ≥92% for the control group (<i>n=</i>3 478) and <92% for the pulmonary obstruction group (<i>n=</i>7 405). The obstruction group was subdivided into five groups according to the degree of pulmonary dysfunction: mild group (<i>n=</i>3 938),moderate group (<i>n=</i>1 142),oderate-severe group (<i>n=</i>917),severe group (<i>n=</i>737),and extremely severe group (<i>n=</i>671). Conventional pulmonary ventilatory function FVC, FEV<sub>1</sub>, one-second rate, and forced expired flow at 50% of FVC (MEF50%), forced expired flow at 75% FVC (MEF25%), maximal mid-expiratory flow (MMEF), peak expiratory flow (PEF), and pulsed oscillation pulmonary function test were detected in both groups of patients. Impedance at 5 Hz (Z5) means total respiratory resistance, resistance at 5 Hz (R5) means total airway resistance, reactance at 5 Hz (X5) indicates the elastic recoil of the peripheral airways, and resistance at 20 Hz (R20) represents resistance of the central airways. R5-R20 reflects resistance in the small airways. Additionally, peripheral resistance (Rp), respiratory resonance frequency (Frex), and area under the reactance curve (Ax) were also measured. Correlation between the indicators of the two groups and the sensitivity and specificity of the impulse oscillometry system parameters for the diagnosis of obstructive pulmonary ventilation dysfunction were analyzed. <b>Results:</b> Pulmonary function force expiratory volume in the first second as a percentage of predicted value (FEV<sub>1</sub>%Pre) [80.10 (54.95,97.10)%],one-second rate [62.43(48.67, 67.02)%],MEF50% [1.33 (0.62,1.97)L/s],MEF25% [0.28 (0.17,0.41)L/s], MMEF [0.85 (0.43,1.29)L/s],and PEF [5.64 (3.73,7.50)]L/s in the obstruction group were significantly lower than those in the control group (<i>P<</i>0.05). The differences within the subgroups of the obstruction group were also significant (<i>P<</i>0.05). Pulsed oscillation Z5 [0.42 (0.33,0.55)kPa·L<sup>-1</sup>·s<sup>-1</sup>],Rp [0.25 (0.20,0.45)kPa·L<sup>-1</sup>·s<sup>-1</sup>], R5 [0.39 (0.31,0.49)kPa·L<sup>-1</sup>·s<sup>-1</sup>], R20 [0.28 (0.24,0.34)kPa·L<sup>-1</sup>·s<sup>-1</sup>], R5-R20 [0.09 (0.05,0.17)kPa·L<sup>-1</sup>·s<sup>-1</sup>],Frex [16.32 (13.07,20.84)Hz], and Ax [0.67 (0.28,1.64)] indices in the obstruction group were significantly higher than those in the control group. X5 [-0.14 (-0.23, -0.10)kPa·L<sup>-1</sup>·s<sup>-1</sup>] was significantly lower than that in the control group (<i>P<</i>0.05). Z5, Rp, X5, R5, R5-R20, Frex, and Ax were statistically significant between different degrees of obstruction in the obstruction group (<i>P<</i>0.05). The impulse oscillometry system parameters Z5, Rp, R5, R20, R5-20, Frex, and Ax were negatively correlated with the indices of conventional pulmonary ventilation (<i>r</i>=-0.21-0.68, <i>P<</i>0.05), and the parameter X5 was positively correlated with the indices of conventional pulmonary ventilation (<i>r=</i>0.41-0.68, <i>P<</i>0.05). The pulsed oscillation pulmonary function test parameters X5 (58.60%-95.68%) and Ax (57.08%-98.06%) presented the best sensitivity; X5 (86.29%-98.82%), Frex (86.69%-94.71%), and Ax (88.10%-98.53%) displayed the best specificity; and R20 presented the worst sensitivity and specificity. The sensitivity and specificity were slightly better in female patients than in male patients. <b>Conclusion:</b> The technical parameters of the impulse oscillometry system showed significant correlation with relevant indices of conventional pulmonary ventilation function detection. These well reflect the changes of different degrees of pulmonary ventilation function and have greater significance for reference in evaluating the degree of pulmonary function impairment.</p>\",\"PeriodicalId\":68309,\"journal\":{\"name\":\"中华内科杂志\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-11-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中华内科杂志\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.3760/cma.j.cn112138-20240410-00231\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中华内科杂志","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3760/cma.j.cn112138-20240410-00231","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的研究脉冲振荡测量系统检查指标与常规肺通气功能之间的相关性。方法:纳入华中科技大学同济医学院附属同济医院 2020 年 1 月 1 日至 2022 年 12 月 31 日 10 883 例患者的肺通气功能数据。对照组(3 478 人)和阻塞组(7 405 人)的一秒率[第一秒用力呼气容积(FEV1)与用力肺活量(FVC)之比]占预测值的百分比≥92%。根据肺功能障碍程度将阻塞组细分为五组:轻度组(n=3 938)、中度组(n=1 142)、中重度组(n=917)、重度组(n=737)和极重度组(n=671)。两组患者均检测了常规肺通气功能 FVC、FEV1、一秒钟速率、50% FVC 时的强迫呼气流量(MEF50%)、75% FVC 时的强迫呼气流量(MEF25%)、最大呼气中流量(MMEF)、呼气峰值流量(PEF)和脉冲振荡肺功能测试。5 赫兹阻抗(Z5)表示总呼吸阻力,5 赫兹阻力(R5)表示总气道阻力,5 赫兹反应(X5)表示外周气道的弹性反冲,20 赫兹阻力(R20)表示中央气道的阻力。R5-R20 反映了小气道的阻力。此外,还测量了外周阻力(Rp)、呼吸共振频率(Frex)和反应曲线下面积(Ax)。分析了两组指标之间的相关性以及脉冲振荡测量系统参数对阻塞性肺通气功能障碍诊断的敏感性和特异性。结果肺功能第一秒用力呼气容积占预测值的百分比(FEV1%Pre)[80.10 (54.95,97.10)%], 一秒钟速率[62.43(48.67,67.02)%], MEF50% [1.33 (0. 62,1.97)L/s], MEF50% [1.33 (0. 62,1.97)L/s].62,1.97)L/s]、MEF25%[0.28(0.17,0.41)L/s]、MMEF[0.85(0.43,1.29)L/s]和PEF[5.64(3.73,7.50)]L/s均显著低于对照组(P0.05)。阻塞组各亚组间的差异也很明显(P0.05)。脉冲振荡 Z5 [0.42 (0.33,0.55)kPa-L-1-s-1],Rp [0.25 (0.20,0.45)kPa-L-1-s-1], R5 [0.39 (0.31,0.49)kPa-L-1-s-1], R20 [0.28 (0.24,0.34)kPa-L-1-s-1], R5-R20 [0.09(0.05,0.17)kPa-L-1-s-1]、Frex[16.32(13.07,20.84)Hz]和 Ax[0.67(0.28,1.64)]指数均显著高于对照组。X5 [-0.14 (-0.23, -0.10)kPa-L-1-s-1]明显低于对照组(P0.05)。阻塞组不同阻塞程度之间的 Z5、Rp、X5、R5、R5-R20、Frex 和 Ax 均有统计学意义(P0.05)。脉冲振荡系统参数 Z5、Rp、R5、R20、R5-20、Frex 和 Ax 与常规肺通气指数呈负相关(r=-0.21-0.68,P0.05),参数 X5 与常规肺通气指数呈正相关(r=0.41-0.68,P0.05)。脉冲振荡肺功能测试参数 X5(58.60%-95.68%)和 Ax(57.08%-98.06%)的敏感性最好;X5(86.29%-98.82%)、Frex(86.69%-94.71%)和 Ax(88.10%-98.53%)的特异性最好;R20 的敏感性和特异性最差。女性患者的敏感性和特异性略好于男性患者。结论脉冲振荡测量系统的技术参数与常规肺通气功能检测的相关指标有显著相关性。这些指标能很好地反映不同程度的肺通气功能变化,对评估肺功能损伤程度具有较大的参考意义。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
[Correlation of Impulse oscillometry system indices with conventional pulmonary function tests in patients with obstructive pulmonary ventilation dysfunction].

Objective: To investigate the correlation between impulse oscillometry system examination indicators and conventional pulmonary ventilation function. Methods: The pulmonary ventilation function data of 10 883 patients from January 1, 2020 to December 31, 2022 at Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology were included. The one-second rate [ratio of forced expiratory volume in the first second (FEV1) to forced vital capacity (FVC)] measured as a percentage of the predicted value was ≥92% for the control group (n=3 478) and <92% for the pulmonary obstruction group (n=7 405). The obstruction group was subdivided into five groups according to the degree of pulmonary dysfunction: mild group (n=3 938),moderate group (n=1 142),oderate-severe group (n=917),severe group (n=737),and extremely severe group (n=671). Conventional pulmonary ventilatory function FVC, FEV1, one-second rate, and forced expired flow at 50% of FVC (MEF50%), forced expired flow at 75% FVC (MEF25%), maximal mid-expiratory flow (MMEF), peak expiratory flow (PEF), and pulsed oscillation pulmonary function test were detected in both groups of patients. Impedance at 5 Hz (Z5) means total respiratory resistance, resistance at 5 Hz (R5) means total airway resistance, reactance at 5 Hz (X5) indicates the elastic recoil of the peripheral airways, and resistance at 20 Hz (R20) represents resistance of the central airways. R5-R20 reflects resistance in the small airways. Additionally, peripheral resistance (Rp), respiratory resonance frequency (Frex), and area under the reactance curve (Ax) were also measured. Correlation between the indicators of the two groups and the sensitivity and specificity of the impulse oscillometry system parameters for the diagnosis of obstructive pulmonary ventilation dysfunction were analyzed. Results: Pulmonary function force expiratory volume in the first second as a percentage of predicted value (FEV1%Pre) [80.10 (54.95,97.10)%],one-second rate [62.43(48.67, 67.02)%],MEF50% [1.33 (0.62,1.97)L/s],MEF25% [0.28 (0.17,0.41)L/s], MMEF [0.85 (0.43,1.29)L/s],and PEF [5.64 (3.73,7.50)]L/s in the obstruction group were significantly lower than those in the control group (P<0.05). The differences within the subgroups of the obstruction group were also significant (P<0.05). Pulsed oscillation Z5 [0.42 (0.33,0.55)kPa·L-1·s-1],Rp [0.25 (0.20,0.45)kPa·L-1·s-1], R5 [0.39 (0.31,0.49)kPa·L-1·s-1], R20 [0.28 (0.24,0.34)kPa·L-1·s-1], R5-R20 [0.09 (0.05,0.17)kPa·L-1·s-1],Frex [16.32 (13.07,20.84)Hz], and Ax [0.67 (0.28,1.64)] indices in the obstruction group were significantly higher than those in the control group. X5 [-0.14 (-0.23, -0.10)kPa·L-1·s-1] was significantly lower than that in the control group (P<0.05). Z5, Rp, X5, R5, R5-R20, Frex, and Ax were statistically significant between different degrees of obstruction in the obstruction group (P<0.05). The impulse oscillometry system parameters Z5, Rp, R5, R20, R5-20, Frex, and Ax were negatively correlated with the indices of conventional pulmonary ventilation (r=-0.21-0.68, P<0.05), and the parameter X5 was positively correlated with the indices of conventional pulmonary ventilation (r=0.41-0.68, P<0.05). The pulsed oscillation pulmonary function test parameters X5 (58.60%-95.68%) and Ax (57.08%-98.06%) presented the best sensitivity; X5 (86.29%-98.82%), Frex (86.69%-94.71%), and Ax (88.10%-98.53%) displayed the best specificity; and R20 presented the worst sensitivity and specificity. The sensitivity and specificity were slightly better in female patients than in male patients. Conclusion: The technical parameters of the impulse oscillometry system showed significant correlation with relevant indices of conventional pulmonary ventilation function detection. These well reflect the changes of different degrees of pulmonary ventilation function and have greater significance for reference in evaluating the degree of pulmonary function impairment.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
1.00
自引率
0.00%
发文量
19688
期刊介绍:
期刊最新文献
[2024 Chinese guidelines for the diagnosis and treatment of rheumatoid arthritis]. [A single-center prospective study of vitamin D levels and its supplementary effect in the first trimester]. [Advancements and outstanding questions in neuroimmune diseases over the past decade]. [Advances in targeting the interleukin-6 signalling pathway in cancer therapy]. [Correlation of Impulse oscillometry system indices with conventional pulmonary function tests in patients with obstructive pulmonary ventilation dysfunction].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1