封闭系统传输装置中的死体积可能导致稳定性研究中的采样误差。

IF 1.1 Q4 PHARMACOLOGY & PHARMACY Journal of Pharmacy Technology Pub Date : 2024-12-01 Epub Date: 2024-10-07 DOI:10.1177/87551225241285319
Mélanie Closset, Maire-Lise Colsoul, Benoît Bihin, Jean-Daniel Hecq, Pascal Odou, Laurence Galanti
{"title":"封闭系统传输装置中的死体积可能导致稳定性研究中的采样误差。","authors":"Mélanie Closset, Maire-Lise Colsoul, Benoît Bihin, Jean-Daniel Hecq, Pascal Odou, Laurence Galanti","doi":"10.1177/87551225241285319","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Closed system transfer devices (CSTD) help to reduce the exposure of healthcare professionals to hazardous drugs. They may be used in stability studies conducted on anticancer drugs. During a stability study about polyolefin bags of gemcitabine, Tevadaptor® device was suspected of causing a bias in the evaluation of the concentrations of the first aliquots extracted from the bags.</p><p><strong>Objective: </strong>The objectives are to determine whether the use of a CSTD to prepare a drug solution and to withdraw it from a bag can interfere on the measured concentration compared to the expected one and to suggest hypothesis to explain the phenomenon.</p><p><strong>Method: </strong>In the first experiment, three polyolefin bags of gemcitabine (5.4 mg/mL) were prepared under aseptic conditions using the Tevadaptor Luer Lock Adaptor®. The day of preparation, five aliquots of 3.8 ml each were sequentially withdrawn from each polyolefin bag using the same device. After one day, a new aliquot was withdrawn from each bag. In the second experiment, three polyolefin bags of gemcitabine (5.4 mg/ml) were prepared under aseptic conditions using a needle. One aliquot was extracted using a needle after the preparation from each bag, and another aliquot was extracted after one day. The concentrations of all aliquots were measured by liquid chromatography coupled to a photodiode array detector during the same run.</p><p><strong>Results and discussion: </strong>The concentrations of the first aliquots extracted on day zero from the polyolefin bags using the Tevadaptor Luer Lock Adaptor® exhibit an overestimation of 26% ([95%CI: 23%-29%] P<0.001) compared to the others. Overestimation is not found for subsequent aliquots, or while using a needle to bypass the Tevadaptor® device.</p><p><strong>Conclusion: </strong>This case highlights the bias that may arise when using CSTDs in stability studies. They should be used with comprehensive understanding of their technical specifications.</p>","PeriodicalId":16796,"journal":{"name":"Journal of Pharmacy Technology","volume":null,"pages":null},"PeriodicalIF":1.1000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11536552/pdf/","citationCount":"0","resultStr":"{\"title\":\"Potential Sampling Errors in Stability Studies Due to Dead Volume in Closed System Transfer Devices.\",\"authors\":\"Mélanie Closset, Maire-Lise Colsoul, Benoît Bihin, Jean-Daniel Hecq, Pascal Odou, Laurence Galanti\",\"doi\":\"10.1177/87551225241285319\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Closed system transfer devices (CSTD) help to reduce the exposure of healthcare professionals to hazardous drugs. They may be used in stability studies conducted on anticancer drugs. During a stability study about polyolefin bags of gemcitabine, Tevadaptor® device was suspected of causing a bias in the evaluation of the concentrations of the first aliquots extracted from the bags.</p><p><strong>Objective: </strong>The objectives are to determine whether the use of a CSTD to prepare a drug solution and to withdraw it from a bag can interfere on the measured concentration compared to the expected one and to suggest hypothesis to explain the phenomenon.</p><p><strong>Method: </strong>In the first experiment, three polyolefin bags of gemcitabine (5.4 mg/mL) were prepared under aseptic conditions using the Tevadaptor Luer Lock Adaptor®. The day of preparation, five aliquots of 3.8 ml each were sequentially withdrawn from each polyolefin bag using the same device. After one day, a new aliquot was withdrawn from each bag. In the second experiment, three polyolefin bags of gemcitabine (5.4 mg/ml) were prepared under aseptic conditions using a needle. One aliquot was extracted using a needle after the preparation from each bag, and another aliquot was extracted after one day. The concentrations of all aliquots were measured by liquid chromatography coupled to a photodiode array detector during the same run.</p><p><strong>Results and discussion: </strong>The concentrations of the first aliquots extracted on day zero from the polyolefin bags using the Tevadaptor Luer Lock Adaptor® exhibit an overestimation of 26% ([95%CI: 23%-29%] P<0.001) compared to the others. Overestimation is not found for subsequent aliquots, or while using a needle to bypass the Tevadaptor® device.</p><p><strong>Conclusion: </strong>This case highlights the bias that may arise when using CSTDs in stability studies. They should be used with comprehensive understanding of their technical specifications.</p>\",\"PeriodicalId\":16796,\"journal\":{\"name\":\"Journal of Pharmacy Technology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2024-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11536552/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Pharmacy Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1177/87551225241285319\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/10/7 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"PHARMACOLOGY & PHARMACY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Pharmacy Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/87551225241285319","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/10/7 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"PHARMACOLOGY & PHARMACY","Score":null,"Total":0}
引用次数: 0

摘要

背景:封闭系统转移装置(CSTD)有助于减少医护人员接触危险药物的机会。它们可用于抗癌药物的稳定性研究。在一项关于吉西他滨聚烯烃袋的稳定性研究中,Tevadaptor® 装置被怀疑在评估从聚烯烃袋中提取的第一份等分样品的浓度时造成了偏差:目的:确定使用 CSTD 配制药液并从药袋中提取药液是否会干扰与预期浓度相比的测量浓度,并提出解释该现象的假设:在第一次实验中,使用 Tevadaptor Luer Lock Adaptor® 在无菌条件下配制了三袋吉西他滨(5.4 mg/mL)。制备当天,使用同一装置从每个聚烯烃袋中依次抽取 5 份等分样品,每份 3.8 毫升。一天后,从每个袋中抽取新的等分试样。在第二个实验中,使用针头在无菌条件下制备了三袋吉西他滨(5.4 毫克/毫升)。制备完成后,用针从每个袋子中抽取一份等分样品,一天后再抽取另一份等分样品。通过液相色谱法和光电二极管阵列检测器测量所有等分样品的浓度:结果:使用 Tevadaptor Luer Lock Adaptor® 在第 0 天从聚烯烃袋中提取的第一份等分样品的浓度被高估了 26%([95%CI:23%-29%] PC 结论:这一案例凸显了使用 Tevadaptor Luer Lock Adaptor® 时可能出现的偏差:本案例凸显了在稳定性研究中使用 CSTD 时可能出现的偏差。在使用时应全面了解其技术规格。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Potential Sampling Errors in Stability Studies Due to Dead Volume in Closed System Transfer Devices.

Background: Closed system transfer devices (CSTD) help to reduce the exposure of healthcare professionals to hazardous drugs. They may be used in stability studies conducted on anticancer drugs. During a stability study about polyolefin bags of gemcitabine, Tevadaptor® device was suspected of causing a bias in the evaluation of the concentrations of the first aliquots extracted from the bags.

Objective: The objectives are to determine whether the use of a CSTD to prepare a drug solution and to withdraw it from a bag can interfere on the measured concentration compared to the expected one and to suggest hypothesis to explain the phenomenon.

Method: In the first experiment, three polyolefin bags of gemcitabine (5.4 mg/mL) were prepared under aseptic conditions using the Tevadaptor Luer Lock Adaptor®. The day of preparation, five aliquots of 3.8 ml each were sequentially withdrawn from each polyolefin bag using the same device. After one day, a new aliquot was withdrawn from each bag. In the second experiment, three polyolefin bags of gemcitabine (5.4 mg/ml) were prepared under aseptic conditions using a needle. One aliquot was extracted using a needle after the preparation from each bag, and another aliquot was extracted after one day. The concentrations of all aliquots were measured by liquid chromatography coupled to a photodiode array detector during the same run.

Results and discussion: The concentrations of the first aliquots extracted on day zero from the polyolefin bags using the Tevadaptor Luer Lock Adaptor® exhibit an overestimation of 26% ([95%CI: 23%-29%] P<0.001) compared to the others. Overestimation is not found for subsequent aliquots, or while using a needle to bypass the Tevadaptor® device.

Conclusion: This case highlights the bias that may arise when using CSTDs in stability studies. They should be used with comprehensive understanding of their technical specifications.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Pharmacy Technology
Journal of Pharmacy Technology PHARMACOLOGY & PHARMACY-
CiteScore
1.50
自引率
0.00%
发文量
49
期刊介绍: For both pharmacists and technicians, jPT provides valuable information for those interested in the entire body of pharmacy practice. jPT covers new drugs, products, and equipment; therapeutic trends; organizational, legal, and educational activities; drug distribution and administration; and includes continuing education articles.
期刊最新文献
Implementation and Evaluation of APOTECAchemo in a Community Cancer Center: A Comparative Study of Robotic Versus Manual Antineoplastic Preparation. Implementation of Computerized Prescriber Order Entry Systems: A Review of Impacts, Barriers, and Facilitators. Adverse Effects Associated With Multimodal Analgesic Regimens in Critically Ill, Nonintubated Patients: A Systematic Review and Meta-Analysis. Outcomes of Hospitalized Patients With Sepsis Before and After Implementation of a Sepsis Care Improvement Initiative at a Community Hospital. Potential Sampling Errors in Stability Studies Due to Dead Volume in Closed System Transfer Devices.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1