Ying Zhang, Ziyun Shi, Haixia Liang, Meixia Chen, Yan Sun, Fujun Shang, Yi Wan
{"title":"以 AAMI/ESH/ISO 为参考标准,G.LAB MD41A0 上臂式血压计在妊娠期和子痫前期的临床准确性。","authors":"Ying Zhang, Ziyun Shi, Haixia Liang, Meixia Chen, Yan Sun, Fujun Shang, Yi Wan","doi":"10.2147/VHRM.S479380","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Validation of blood pressure (BP) monitors in pregnant women are needed to ensure the accurate measurement of BP in pregnancy. Therefore the study aimed to evaluate the measurement accuracy of the G.LAB MD41A0 oscillometric automatic upper-arm blood pressure monitor in pregnancy and pre-eclampsia women according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018).</p><p><strong>Methods: </strong>A total of 45 pregnant women were included in the cross-sectional validation study. The same left-arm sequential method was used for blood pressure measurement, and the blood pressure differences between the test device and the mercury standard reference were assessed according to the AAMI/ESH/ISO Universal Standard.</p><p><strong>Results: </strong>The participants included 15 normotensives, 15 gestational hypertension and 15 pre-eclampsia. The average (mean ± SD) differences between the device and mercury standard was -0.84±3.88 mmHg and -0.58±3.35 mmHg for systolic and diastolic blood pressure, respectively, for the validation Criterion 1. The SD of the averaged differences between the device and reference readings per participant was 2.92 mmHg and 2.28 mmHg for systolic and diastolic blood pressure, respectively, for the Criterion 2 of the universal standard.</p><p><strong>Conclusion: </strong>The G.LAB MD41A0 automatic upper-arm blood pressure monitor fulfills the criteria of the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement in pregnancy and pre-eclampsia.</p>","PeriodicalId":23597,"journal":{"name":"Vascular Health and Risk Management","volume":"20 ","pages":"487-492"},"PeriodicalIF":2.6000,"publicationDate":"2024-11-03","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11549012/pdf/","citationCount":"0","resultStr":"{\"title\":\"Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard.\",\"authors\":\"Ying Zhang, Ziyun Shi, Haixia Liang, Meixia Chen, Yan Sun, Fujun Shang, Yi Wan\",\"doi\":\"10.2147/VHRM.S479380\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Objective: </strong>Validation of blood pressure (BP) monitors in pregnant women are needed to ensure the accurate measurement of BP in pregnancy. Therefore the study aimed to evaluate the measurement accuracy of the G.LAB MD41A0 oscillometric automatic upper-arm blood pressure monitor in pregnancy and pre-eclampsia women according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018).</p><p><strong>Methods: </strong>A total of 45 pregnant women were included in the cross-sectional validation study. The same left-arm sequential method was used for blood pressure measurement, and the blood pressure differences between the test device and the mercury standard reference were assessed according to the AAMI/ESH/ISO Universal Standard.</p><p><strong>Results: </strong>The participants included 15 normotensives, 15 gestational hypertension and 15 pre-eclampsia. The average (mean ± SD) differences between the device and mercury standard was -0.84±3.88 mmHg and -0.58±3.35 mmHg for systolic and diastolic blood pressure, respectively, for the validation Criterion 1. The SD of the averaged differences between the device and reference readings per participant was 2.92 mmHg and 2.28 mmHg for systolic and diastolic blood pressure, respectively, for the Criterion 2 of the universal standard.</p><p><strong>Conclusion: </strong>The G.LAB MD41A0 automatic upper-arm blood pressure monitor fulfills the criteria of the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement in pregnancy and pre-eclampsia.</p>\",\"PeriodicalId\":23597,\"journal\":{\"name\":\"Vascular Health and Risk Management\",\"volume\":\"20 \",\"pages\":\"487-492\"},\"PeriodicalIF\":2.6000,\"publicationDate\":\"2024-11-03\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11549012/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Vascular Health and Risk Management\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2147/VHRM.S479380\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q2\",\"JCRName\":\"PERIPHERAL VASCULAR DISEASE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Vascular Health and Risk Management","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2147/VHRM.S479380","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q2","JCRName":"PERIPHERAL VASCULAR DISEASE","Score":null,"Total":0}
Clinical Accuracy of the G.LAB MD41A0 Upper Arm Blood Pressure Monitor in Pregnancy and Pre-Eclampsia Using the AAMI/ESH/ISO as Reference Standard.
Objective: Validation of blood pressure (BP) monitors in pregnant women are needed to ensure the accurate measurement of BP in pregnancy. Therefore the study aimed to evaluate the measurement accuracy of the G.LAB MD41A0 oscillometric automatic upper-arm blood pressure monitor in pregnancy and pre-eclampsia women according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018).
Methods: A total of 45 pregnant women were included in the cross-sectional validation study. The same left-arm sequential method was used for blood pressure measurement, and the blood pressure differences between the test device and the mercury standard reference were assessed according to the AAMI/ESH/ISO Universal Standard.
Results: The participants included 15 normotensives, 15 gestational hypertension and 15 pre-eclampsia. The average (mean ± SD) differences between the device and mercury standard was -0.84±3.88 mmHg and -0.58±3.35 mmHg for systolic and diastolic blood pressure, respectively, for the validation Criterion 1. The SD of the averaged differences between the device and reference readings per participant was 2.92 mmHg and 2.28 mmHg for systolic and diastolic blood pressure, respectively, for the Criterion 2 of the universal standard.
Conclusion: The G.LAB MD41A0 automatic upper-arm blood pressure monitor fulfills the criteria of the AAMI/ESH/ISO Universal Standard and can be recommended for clinical use and self-measurement in pregnancy and pre-eclampsia.
期刊介绍:
An international, peer-reviewed journal of therapeutics and risk management, focusing on concise rapid reporting of clinical studies on the processes involved in the maintenance of vascular health; the monitoring, prevention, and treatment of vascular disease and its sequelae; and the involvement of metabolic disorders, particularly diabetes. In addition, the journal will also seek to define drug usage in terms of ultimate uptake and acceptance by the patient and healthcare professional.