使用具有高空间分辨率和双能量选择功能的碲化镉平板探测器的光子计数 X 射线计算机断层扫描。

IF 0.7 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Journal of Medical Physics Pub Date : 2024-07-01 Epub Date: 2024-09-21 DOI:10.4103/jmp.jmp_33_24
Jiro Sato, Eiichi Sato, Kazuki Ito, Hodaka Moriyama, Osahiko Hagiwara, Toshiyuki Enomoto, Manabu Watanabe, Sohei Yoshida, Kunihiro Yoshioka, Hiroyuki Nitta
{"title":"使用具有高空间分辨率和双能量选择功能的碲化镉平板探测器的光子计数 X 射线计算机断层扫描。","authors":"Jiro Sato, Eiichi Sato, Kazuki Ito, Hodaka Moriyama, Osahiko Hagiwara, Toshiyuki Enomoto, Manabu Watanabe, Sohei Yoshida, Kunihiro Yoshioka, Hiroyuki Nitta","doi":"10.4103/jmp.jmp_33_24","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>To perform energy-dispersive X-ray imaging, we constructed a photon-counting X-ray computed tomography (CT) scanner to perform enhanced K-edge CT.</p><p><strong>Methods: </strong>X-ray photons penetrating through an object were detected using a cadmium telluride flat panel detector (FPD) with pixel dimensions of 100 × 100 mm<sup>2</sup>, and 720 radiograms from the FPD were sent to the personal computer to reconstruct tomograms. Gadolinium (Gd) K-edge energy is 50.2 keV, and Gd-Kedge CT was carried out using photons with an energy range of 50-100 keV.</p><p><strong>Results: </strong>Compared with low-energy CT of 15-50 keV, the gray density of muscle and bone substantially decreased, and the image contrast of Gd media was improved utilizing Gd-K-edge CT.</p><p><strong>Conclusion: </strong>Using the cone beam, the effective pixel dimensions were 80 × 80 μm<sup>2</sup>, and blood vessels were observed at a high contrast using Gd-Kedge CT.</p>","PeriodicalId":51719,"journal":{"name":"Journal of Medical Physics","volume":"49 3","pages":"441-447"},"PeriodicalIF":0.7000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11548080/pdf/","citationCount":"0","resultStr":"{\"title\":\"Photon-counting X-ray Computed Tomography Using a Cadmium Telluride Flat Panel Detector with High Spatial Resolutions and Dual-energy Selection.\",\"authors\":\"Jiro Sato, Eiichi Sato, Kazuki Ito, Hodaka Moriyama, Osahiko Hagiwara, Toshiyuki Enomoto, Manabu Watanabe, Sohei Yoshida, Kunihiro Yoshioka, Hiroyuki Nitta\",\"doi\":\"10.4103/jmp.jmp_33_24\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>To perform energy-dispersive X-ray imaging, we constructed a photon-counting X-ray computed tomography (CT) scanner to perform enhanced K-edge CT.</p><p><strong>Methods: </strong>X-ray photons penetrating through an object were detected using a cadmium telluride flat panel detector (FPD) with pixel dimensions of 100 × 100 mm<sup>2</sup>, and 720 radiograms from the FPD were sent to the personal computer to reconstruct tomograms. Gadolinium (Gd) K-edge energy is 50.2 keV, and Gd-Kedge CT was carried out using photons with an energy range of 50-100 keV.</p><p><strong>Results: </strong>Compared with low-energy CT of 15-50 keV, the gray density of muscle and bone substantially decreased, and the image contrast of Gd media was improved utilizing Gd-K-edge CT.</p><p><strong>Conclusion: </strong>Using the cone beam, the effective pixel dimensions were 80 × 80 μm<sup>2</sup>, and blood vessels were observed at a high contrast using Gd-Kedge CT.</p>\",\"PeriodicalId\":51719,\"journal\":{\"name\":\"Journal of Medical Physics\",\"volume\":\"49 3\",\"pages\":\"441-447\"},\"PeriodicalIF\":0.7000,\"publicationDate\":\"2024-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11548080/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/jmp.jmp_33_24\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/9/21 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/jmp.jmp_33_24","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/9/21 0:00:00","PubModel":"Epub","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

简介:为了进行能量色散 X 射线成像,我们构建了一个光子计数 X 射线计算机断层扫描(CT)扫描仪:为了进行能量色散 X 射线成像,我们建造了一台光子计数 X 射线计算机断层扫描(CT)扫描仪,以进行增强 K 边 CT:方法:使用像素尺寸为 100 × 100 mm2 的碲化镉平板探测器(FPD)检测穿透物体的 X 射线光子,并将 FPD 的 720 幅放射图发送到个人计算机,以重建断层图。钆(Gd)K 边能量为 50.2 keV,使用能量范围为 50-100 keV 的光子进行钆边 CT:结果:与 15-50 keV 的低能量 CT 相比,肌肉和骨骼的灰密度大幅降低,利用 Gd-Kedge CT 提高了钆介质的图像对比度:结论:利用锥形束,有效像素尺寸为 80 × 80 μm2,利用钆边缘 CT 可以观察到高对比度的血管。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Photon-counting X-ray Computed Tomography Using a Cadmium Telluride Flat Panel Detector with High Spatial Resolutions and Dual-energy Selection.

Introduction: To perform energy-dispersive X-ray imaging, we constructed a photon-counting X-ray computed tomography (CT) scanner to perform enhanced K-edge CT.

Methods: X-ray photons penetrating through an object were detected using a cadmium telluride flat panel detector (FPD) with pixel dimensions of 100 × 100 mm2, and 720 radiograms from the FPD were sent to the personal computer to reconstruct tomograms. Gadolinium (Gd) K-edge energy is 50.2 keV, and Gd-Kedge CT was carried out using photons with an energy range of 50-100 keV.

Results: Compared with low-energy CT of 15-50 keV, the gray density of muscle and bone substantially decreased, and the image contrast of Gd media was improved utilizing Gd-K-edge CT.

Conclusion: Using the cone beam, the effective pixel dimensions were 80 × 80 μm2, and blood vessels were observed at a high contrast using Gd-Kedge CT.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Medical Physics
Journal of Medical Physics RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.10
自引率
11.10%
发文量
55
审稿时长
30 weeks
期刊介绍: JOURNAL OF MEDICAL PHYSICS is the official journal of Association of Medical Physicists of India (AMPI). The association has been bringing out a quarterly publication since 1976. Till the end of 1993, it was known as Medical Physics Bulletin, which then became Journal of Medical Physics. The main objective of the Journal is to serve as a vehicle of communication to highlight all aspects of the practice of medical radiation physics. The areas covered include all aspects of the application of radiation physics to biological sciences, radiotherapy, radiodiagnosis, nuclear medicine, dosimetry and radiation protection. Papers / manuscripts dealing with the aspects of physics related to cancer therapy / radiobiology also fall within the scope of the journal.
期刊最新文献
A Segmentation-based Automated Calculation of Patient Size and Size-specific Dose Estimates in Pediatric Computed Tomography Scans. A Study on Radiation Level at the Treatment Plane Due to Induced Activity in Linear Accelerator Head. Advancements and Applications of Three-dimensional Printing Technology in Surgery. Agar-based Phantom for Evaluating Targeting of High-intensity Focused Ultrasound Systems for Breast Ablation. An Analysis of Radiotherapy Machine Requirements in India: Impact of the Pandemic and Regional Disparities.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1