Mi-Jeong Yoon , Hyungtaek Kim , Yeun Jie Yoo , Sun Im , Tae-Woo Kim , Yasin Y. Dhaher , Donghyeon Kim , Seong Hoon Lim
{"title":"中风患者颅骨毛刺孔经颅直流电刺激电场调制的硅学建模:对安全临床应用的影响。","authors":"Mi-Jeong Yoon , Hyungtaek Kim , Yeun Jie Yoo , Sun Im , Tae-Woo Kim , Yasin Y. Dhaher , Donghyeon Kim , Seong Hoon Lim","doi":"10.1016/j.compbiomed.2024.109366","DOIUrl":null,"url":null,"abstract":"<div><h3>Background</h3><div>Transcranial direct current stimulation (tDCS) has emerged as a promising tool for stroke rehabilitation, supported by evidence demonstrating its beneficial effects on post-stroke recovery. However, patients with skull defects, such as burr holes, have been excluded from tDCS due to limited knowledge regarding the effect of skull defects on the electric field.</div></div><div><h3>Objective</h3><div>We investigated the effect of burr holes on the electric field induced by tDCS and identified the electrode location that modulates the electric field.</div></div><div><h3>Methods</h3><div>We generated mesh models of the heads of five patients with burr holes and five age-matched control patients who had never undergone brain surgery, based on magnetic resonance imaging. Then we conducted tDCS simulations, with the cathode fixed in one position and the anode in various positions. Regression analysis was employed to investigate the relationship between the electric field at the burr hole and the distance from the burr hole to the anode.</div></div><div><h3>Results</h3><div>In patients with burr holes, the electric field intensity increased as the anode approached the burr hole, reaching a maximum electric field when the anode covered it, with this pattern remaining consistent across all patient models. Assuming the holes were filled with cerebrospinal fluid, the maximum electric field was 1.20 ± 0.20 V/m (mean ± standard deviation, SD). When the anode was positioned more than 60 mm away from the burr hole, the electric field at the burr hole remained low and constant, with an average value of 0.29 ± 0.04V/m (mean ± SD). In contrast, for all patients without burr holes, the electric field intensity stayed constant regardless of the anode's position, with a maximum amplitude of 0.36 ± 0.04 V/m (mean ± SD). Furthermore, when the burr hole was assumed to be filled with scar tissue, the mean peak electric field was 0.93 ± 0.16 V/m, indicating that the electric field strength varies depending on the conductivity of the tissue filling the burr hole.</div></div><div><h3>Conclusion</h3><div>Based on the simulations, the minimum recommended distance from the burr hole to the anode is 60 mm to prevent unintended stimulation of the brain cortex during tDCS. These findings will contribute to the development of safe and effective tDCS treatments for patients with burr holes.</div></div>","PeriodicalId":10578,"journal":{"name":"Computers in biology and medicine","volume":"184 ","pages":"Article 109366"},"PeriodicalIF":7.0000,"publicationDate":"2024-11-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"In silico modeling of electric field modulation by transcranial direct current stimulation in stroke patients with skull burr holes: Implications for safe clinical application\",\"authors\":\"Mi-Jeong Yoon , Hyungtaek Kim , Yeun Jie Yoo , Sun Im , Tae-Woo Kim , Yasin Y. Dhaher , Donghyeon Kim , Seong Hoon Lim\",\"doi\":\"10.1016/j.compbiomed.2024.109366\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background</h3><div>Transcranial direct current stimulation (tDCS) has emerged as a promising tool for stroke rehabilitation, supported by evidence demonstrating its beneficial effects on post-stroke recovery. However, patients with skull defects, such as burr holes, have been excluded from tDCS due to limited knowledge regarding the effect of skull defects on the electric field.</div></div><div><h3>Objective</h3><div>We investigated the effect of burr holes on the electric field induced by tDCS and identified the electrode location that modulates the electric field.</div></div><div><h3>Methods</h3><div>We generated mesh models of the heads of five patients with burr holes and five age-matched control patients who had never undergone brain surgery, based on magnetic resonance imaging. Then we conducted tDCS simulations, with the cathode fixed in one position and the anode in various positions. Regression analysis was employed to investigate the relationship between the electric field at the burr hole and the distance from the burr hole to the anode.</div></div><div><h3>Results</h3><div>In patients with burr holes, the electric field intensity increased as the anode approached the burr hole, reaching a maximum electric field when the anode covered it, with this pattern remaining consistent across all patient models. Assuming the holes were filled with cerebrospinal fluid, the maximum electric field was 1.20 ± 0.20 V/m (mean ± standard deviation, SD). When the anode was positioned more than 60 mm away from the burr hole, the electric field at the burr hole remained low and constant, with an average value of 0.29 ± 0.04V/m (mean ± SD). In contrast, for all patients without burr holes, the electric field intensity stayed constant regardless of the anode's position, with a maximum amplitude of 0.36 ± 0.04 V/m (mean ± SD). Furthermore, when the burr hole was assumed to be filled with scar tissue, the mean peak electric field was 0.93 ± 0.16 V/m, indicating that the electric field strength varies depending on the conductivity of the tissue filling the burr hole.</div></div><div><h3>Conclusion</h3><div>Based on the simulations, the minimum recommended distance from the burr hole to the anode is 60 mm to prevent unintended stimulation of the brain cortex during tDCS. These findings will contribute to the development of safe and effective tDCS treatments for patients with burr holes.</div></div>\",\"PeriodicalId\":10578,\"journal\":{\"name\":\"Computers in biology and medicine\",\"volume\":\"184 \",\"pages\":\"Article 109366\"},\"PeriodicalIF\":7.0000,\"publicationDate\":\"2024-11-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Computers in biology and medicine\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0010482524014513\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"BIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Computers in biology and medicine","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0010482524014513","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"BIOLOGY","Score":null,"Total":0}
In silico modeling of electric field modulation by transcranial direct current stimulation in stroke patients with skull burr holes: Implications for safe clinical application
Background
Transcranial direct current stimulation (tDCS) has emerged as a promising tool for stroke rehabilitation, supported by evidence demonstrating its beneficial effects on post-stroke recovery. However, patients with skull defects, such as burr holes, have been excluded from tDCS due to limited knowledge regarding the effect of skull defects on the electric field.
Objective
We investigated the effect of burr holes on the electric field induced by tDCS and identified the electrode location that modulates the electric field.
Methods
We generated mesh models of the heads of five patients with burr holes and five age-matched control patients who had never undergone brain surgery, based on magnetic resonance imaging. Then we conducted tDCS simulations, with the cathode fixed in one position and the anode in various positions. Regression analysis was employed to investigate the relationship between the electric field at the burr hole and the distance from the burr hole to the anode.
Results
In patients with burr holes, the electric field intensity increased as the anode approached the burr hole, reaching a maximum electric field when the anode covered it, with this pattern remaining consistent across all patient models. Assuming the holes were filled with cerebrospinal fluid, the maximum electric field was 1.20 ± 0.20 V/m (mean ± standard deviation, SD). When the anode was positioned more than 60 mm away from the burr hole, the electric field at the burr hole remained low and constant, with an average value of 0.29 ± 0.04V/m (mean ± SD). In contrast, for all patients without burr holes, the electric field intensity stayed constant regardless of the anode's position, with a maximum amplitude of 0.36 ± 0.04 V/m (mean ± SD). Furthermore, when the burr hole was assumed to be filled with scar tissue, the mean peak electric field was 0.93 ± 0.16 V/m, indicating that the electric field strength varies depending on the conductivity of the tissue filling the burr hole.
Conclusion
Based on the simulations, the minimum recommended distance from the burr hole to the anode is 60 mm to prevent unintended stimulation of the brain cortex during tDCS. These findings will contribute to the development of safe and effective tDCS treatments for patients with burr holes.
期刊介绍:
Computers in Biology and Medicine is an international forum for sharing groundbreaking advancements in the use of computers in bioscience and medicine. This journal serves as a medium for communicating essential research, instruction, ideas, and information regarding the rapidly evolving field of computer applications in these domains. By encouraging the exchange of knowledge, we aim to facilitate progress and innovation in the utilization of computers in biology and medicine.