{"title":"在近视屈光手术候选者中,基于皮拉米德波前的自动屈光度与干性屈光度、环形屈光度和主观屈光度的一致性。","authors":"Siamak Zarei-Ghanavati, Marzieh Najjaran, Samira Hassanzadeh, Mehdi Khabazkhoob, Akilesh Gokul, Mohammed Ziaei","doi":"10.4103/joco.joco_261_23","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the repeatability of a pyramidal wavefront-based refraction (WFR) measurement and its agreement with dry autorefraction (DR), cycloplegic autorefraction (CR), and subjective refraction (SR) in myopic refractive surgery candidates.</p><p><strong>Methods: </strong>One hundred eighty-nine eyes from 189 participants were evaluated. PERAMIS aberrometry (PERAMIS; SCHWIND eye-tech-solutions, Kleinostheim, Germany), DR, and CR, as well as SR, were completed for all candidates. The repeatability of PERAMIS measurements was assessed, and the Bland-Altman plots were used to test the agreement between different methods.</p><p><strong>Results: </strong>Repeatability of the PERAMIS aberrometer was very high in the measurement of all refractive elements (Sphere, cylinder, spherical equivalent [M], J0, and J45) (interclass correlation coefficient >0.980 for all). A significant myopic shift was found with WFR compared to CR (0.45 diopter [D]) and SR (0.28 D) (<i>P</i> < 0.05). For the M component, there was a significant difference between WFR and CR (<i>P</i> < 0.05). J0 component measured with WFR was significantly different from SR and CR (<i>P</i> < 0.05). For the J45 variable, all three refraction methods were comparable (all, <i>P</i> > 0.05). In M > 5.00 D, a difference of 0.79 D (limit of agreement: -3.50-1.90) was found between WFR and CR.</p><p><strong>Conclusions: </strong>In lower degrees of myopia, WFR was in good agreement with the manifest autorefraction. WFR, CR, and SR techniques were comparable in measuring astigmatism, especially in oblique astigmatism.</p>","PeriodicalId":15423,"journal":{"name":"Journal of Current Ophthalmology","volume":"36 1","pages":"54-60"},"PeriodicalIF":1.2000,"publicationDate":"2024-10-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11567603/pdf/","citationCount":"0","resultStr":"{\"title\":\"Agreement of a Pyramidal Wavefront-Based Autorefraction with Dry, Cycloplegic, and Subjective Refraction in Myopic Refractive Surgery Candidates.\",\"authors\":\"Siamak Zarei-Ghanavati, Marzieh Najjaran, Samira Hassanzadeh, Mehdi Khabazkhoob, Akilesh Gokul, Mohammed Ziaei\",\"doi\":\"10.4103/joco.joco_261_23\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To evaluate the repeatability of a pyramidal wavefront-based refraction (WFR) measurement and its agreement with dry autorefraction (DR), cycloplegic autorefraction (CR), and subjective refraction (SR) in myopic refractive surgery candidates.</p><p><strong>Methods: </strong>One hundred eighty-nine eyes from 189 participants were evaluated. PERAMIS aberrometry (PERAMIS; SCHWIND eye-tech-solutions, Kleinostheim, Germany), DR, and CR, as well as SR, were completed for all candidates. The repeatability of PERAMIS measurements was assessed, and the Bland-Altman plots were used to test the agreement between different methods.</p><p><strong>Results: </strong>Repeatability of the PERAMIS aberrometer was very high in the measurement of all refractive elements (Sphere, cylinder, spherical equivalent [M], J0, and J45) (interclass correlation coefficient >0.980 for all). A significant myopic shift was found with WFR compared to CR (0.45 diopter [D]) and SR (0.28 D) (<i>P</i> < 0.05). For the M component, there was a significant difference between WFR and CR (<i>P</i> < 0.05). J0 component measured with WFR was significantly different from SR and CR (<i>P</i> < 0.05). For the J45 variable, all three refraction methods were comparable (all, <i>P</i> > 0.05). In M > 5.00 D, a difference of 0.79 D (limit of agreement: -3.50-1.90) was found between WFR and CR.</p><p><strong>Conclusions: </strong>In lower degrees of myopia, WFR was in good agreement with the manifest autorefraction. WFR, CR, and SR techniques were comparable in measuring astigmatism, especially in oblique astigmatism.</p>\",\"PeriodicalId\":15423,\"journal\":{\"name\":\"Journal of Current Ophthalmology\",\"volume\":\"36 1\",\"pages\":\"54-60\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2024-10-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11567603/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Current Ophthalmology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.4103/joco.joco_261_23\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q3\",\"JCRName\":\"OPHTHALMOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Current Ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/joco.joco_261_23","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"OPHTHALMOLOGY","Score":null,"Total":0}
Agreement of a Pyramidal Wavefront-Based Autorefraction with Dry, Cycloplegic, and Subjective Refraction in Myopic Refractive Surgery Candidates.
Purpose: To evaluate the repeatability of a pyramidal wavefront-based refraction (WFR) measurement and its agreement with dry autorefraction (DR), cycloplegic autorefraction (CR), and subjective refraction (SR) in myopic refractive surgery candidates.
Methods: One hundred eighty-nine eyes from 189 participants were evaluated. PERAMIS aberrometry (PERAMIS; SCHWIND eye-tech-solutions, Kleinostheim, Germany), DR, and CR, as well as SR, were completed for all candidates. The repeatability of PERAMIS measurements was assessed, and the Bland-Altman plots were used to test the agreement between different methods.
Results: Repeatability of the PERAMIS aberrometer was very high in the measurement of all refractive elements (Sphere, cylinder, spherical equivalent [M], J0, and J45) (interclass correlation coefficient >0.980 for all). A significant myopic shift was found with WFR compared to CR (0.45 diopter [D]) and SR (0.28 D) (P < 0.05). For the M component, there was a significant difference between WFR and CR (P < 0.05). J0 component measured with WFR was significantly different from SR and CR (P < 0.05). For the J45 variable, all three refraction methods were comparable (all, P > 0.05). In M > 5.00 D, a difference of 0.79 D (limit of agreement: -3.50-1.90) was found between WFR and CR.
Conclusions: In lower degrees of myopia, WFR was in good agreement with the manifest autorefraction. WFR, CR, and SR techniques were comparable in measuring astigmatism, especially in oblique astigmatism.
期刊介绍:
Peer Review under the responsibility of Iranian Society of Ophthalmology Journal of Current Ophthalmology, the official publication of the Iranian Society of Ophthalmology, is a peer-reviewed, open-access, scientific journal that welcomes high quality original articles related to vision science and all fields of ophthalmology. Journal of Current Ophthalmology is the continuum of Iranian Journal of Ophthalmology published since 1969.