种植体螺丝刀在多次拧紧/松开周期中的有效性评估。体外研究。

IF 1.1 Q3 DENTISTRY, ORAL SURGERY & MEDICINE European Journal of Prosthodontics and Restorative Dentistry Pub Date : 2024-12-11 DOI:10.1922/EJPRD_2743Batista09
R Batista, A Moreira, N Alves, M H Figueiral, J C Sampaio-Fernandes
{"title":"种植体螺丝刀在多次拧紧/松开周期中的有效性评估。体外研究。","authors":"R Batista, A Moreira, N Alves, M H Figueiral, J C Sampaio-Fernandes","doi":"10.1922/EJPRD_2743Batista09","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>It is unclear the number of times a screwdriver may be reused without loosening its efficacy or inducing damage to implant screws while keeping appropriate torque values.</p><p><strong>Objective: </strong>Evaluate the effect of the screwdriver and internal threads on the tightening efficacy by assessing the reverse torque value (RTV).</p><p><strong>Methods: </strong>The efficacy was tested with four groups of screws: two titanium groups and two gold-platted (n=5). Each screw was tightened 10 times at 20 Ncm using the screwdriver attached to ISD900 (NSK®). In the first protocol one implant was used for the five screws. In the second protocol for each screw a new implant was used. At each untightening, the RTV was measured and recorded.</p><p><strong>Results: </strong>RTVs for Gold were statistically lower compared to the Titanium. The RTVs differed among the tested screws. In the first protocol, the first serie tightening/untightening differed statistically from the last three series.</p><p><strong>Conclusion: </strong>Titanium screws from Zimmer Biomet achieve better RTV than gold-platted. By increasing the number cycles it is expected to have a decrease in the RTV. The extensive use of screwdrivers and the internal threads of the dental implants might be factors influencing the loss of RTV.</p>","PeriodicalId":45686,"journal":{"name":"European Journal of Prosthodontics and Restorative Dentistry","volume":" ","pages":""},"PeriodicalIF":1.1000,"publicationDate":"2024-12-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Assessment of Efficacy of Implant Screwdrivers' During Multiple Tightening/ Untightening Cycles. An in vitro Study.\",\"authors\":\"R Batista, A Moreira, N Alves, M H Figueiral, J C Sampaio-Fernandes\",\"doi\":\"10.1922/EJPRD_2743Batista09\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>It is unclear the number of times a screwdriver may be reused without loosening its efficacy or inducing damage to implant screws while keeping appropriate torque values.</p><p><strong>Objective: </strong>Evaluate the effect of the screwdriver and internal threads on the tightening efficacy by assessing the reverse torque value (RTV).</p><p><strong>Methods: </strong>The efficacy was tested with four groups of screws: two titanium groups and two gold-platted (n=5). Each screw was tightened 10 times at 20 Ncm using the screwdriver attached to ISD900 (NSK®). In the first protocol one implant was used for the five screws. In the second protocol for each screw a new implant was used. At each untightening, the RTV was measured and recorded.</p><p><strong>Results: </strong>RTVs for Gold were statistically lower compared to the Titanium. The RTVs differed among the tested screws. In the first protocol, the first serie tightening/untightening differed statistically from the last three series.</p><p><strong>Conclusion: </strong>Titanium screws from Zimmer Biomet achieve better RTV than gold-platted. By increasing the number cycles it is expected to have a decrease in the RTV. The extensive use of screwdrivers and the internal threads of the dental implants might be factors influencing the loss of RTV.</p>\",\"PeriodicalId\":45686,\"journal\":{\"name\":\"European Journal of Prosthodontics and Restorative Dentistry\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2024-12-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"European Journal of Prosthodontics and Restorative Dentistry\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1922/EJPRD_2743Batista09\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Prosthodontics and Restorative Dentistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1922/EJPRD_2743Batista09","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 0

摘要

简介:在保持适当的扭矩值的情况下,螺丝刀可以重复使用多少次,而不会松动其功效或导致植入螺钉损坏,目前尚不清楚。目的:通过评估反向扭矩值(RTV),评价螺丝刀和内螺纹对螺钉拧紧效果的影响。方法:采用4组螺钉:2组钛螺钉和2组镀金螺钉(n=5)进行疗效观察。使用ISD900 (NSK®)上的螺丝刀在20ncm处拧紧每颗螺钉10次。在第一个方案中,五个螺钉使用一个植入物。在第二个方案中,每个螺钉使用一个新的植入物。每次放松时,测量并记录RTV。结果:金的rtv低于钛的rtv。被测螺钉的rtv不同。在第一个方案中,第一个系列的拧紧/松开与后三个系列有统计学差异。结论:Zimmer Biomet公司钛合金螺钉的RTV优于镀金螺钉。通过增加循环次数,预计RTV会下降。螺丝刀的广泛使用和种植体的内螺纹可能是影响RTV丧失的因素。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Assessment of Efficacy of Implant Screwdrivers' During Multiple Tightening/ Untightening Cycles. An in vitro Study.

Introduction: It is unclear the number of times a screwdriver may be reused without loosening its efficacy or inducing damage to implant screws while keeping appropriate torque values.

Objective: Evaluate the effect of the screwdriver and internal threads on the tightening efficacy by assessing the reverse torque value (RTV).

Methods: The efficacy was tested with four groups of screws: two titanium groups and two gold-platted (n=5). Each screw was tightened 10 times at 20 Ncm using the screwdriver attached to ISD900 (NSK®). In the first protocol one implant was used for the five screws. In the second protocol for each screw a new implant was used. At each untightening, the RTV was measured and recorded.

Results: RTVs for Gold were statistically lower compared to the Titanium. The RTVs differed among the tested screws. In the first protocol, the first serie tightening/untightening differed statistically from the last three series.

Conclusion: Titanium screws from Zimmer Biomet achieve better RTV than gold-platted. By increasing the number cycles it is expected to have a decrease in the RTV. The extensive use of screwdrivers and the internal threads of the dental implants might be factors influencing the loss of RTV.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
2.30
自引率
7.70%
发文量
0
期刊介绍: The European Journal of Prosthodontics and Restorative Dentistry is published quarterly and includes clinical and research articles in subjects such as prosthodontics, operative dentistry, implantology, endodontics, periodontics and dental materials.
期刊最新文献
Accuracy of Intra-Oral Scanners for Full Crown Tooth Preparations with Subgingival Margins: A Systematic Review. Description and Durability of the Various Direct Restoration Techniques in Molar-Incisor Hypomineralization: A Systematic Review. Does the Learning Curve for Intraoral Scanning Vary Depending on the Device? Influence of Surface Roughness on the Adhesion of Hydroxyapatite Coatings to Titanium and Titanium Alloy Surfaces: A Systematic Review of in vitro Experimental Studies. Performance of Fiberglass Posts Versus Fiber- Reinforced Resin Composites in Endodontically Treated Anterior Teeth Without Ferrule: A Systematic Review.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1