{"title":"单平面透视检查早期多胫骨错位致聚乙烯断裂1例。","authors":"Atsushi Sugimoto, Teruya Ishibashi, Takaharu Yamazaki, Masashi Tamaki, Seiji Okada, Tetsuya Tomita","doi":"10.2106/JBJS.CC.24.00466","DOIUrl":null,"url":null,"abstract":"<p><strong>Case: </strong>A 75-year-old woman underwent cruciate retaining total knee arthroplasty (TKA) for osteonecrosis of the knee, converted from unicompartmental arthroplasty due to an intraoperative medial tibial plateau fracture. Four years later, she experienced persistent knee pain. Computed tomography revealed tibial malalignments in 2 planes. In vivo kinematic analysis using single-plane fluoroscopy, which can visualize the femorotibial contact, indicated varus thrust with lateral subluxation of the tibial component and the femoral component recessing into the polyethylene insert. Revision TKA affirmed the polyethylene insert breakage.</p><p><strong>Conclusion: </strong>Malalignment might have caused mechanical failure, and in vivo kinematic analysis was useful to visualize abnormal kinematics in TKA.</p>","PeriodicalId":14748,"journal":{"name":"JBJS case connector","volume":"15 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2025-01-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Detection of Multiple Tibial Malalignment-Induced Early Polyethylene Breakage Using Single Plane Fluoroscopy: A Case Report.\",\"authors\":\"Atsushi Sugimoto, Teruya Ishibashi, Takaharu Yamazaki, Masashi Tamaki, Seiji Okada, Tetsuya Tomita\",\"doi\":\"10.2106/JBJS.CC.24.00466\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Case: </strong>A 75-year-old woman underwent cruciate retaining total knee arthroplasty (TKA) for osteonecrosis of the knee, converted from unicompartmental arthroplasty due to an intraoperative medial tibial plateau fracture. Four years later, she experienced persistent knee pain. Computed tomography revealed tibial malalignments in 2 planes. In vivo kinematic analysis using single-plane fluoroscopy, which can visualize the femorotibial contact, indicated varus thrust with lateral subluxation of the tibial component and the femoral component recessing into the polyethylene insert. Revision TKA affirmed the polyethylene insert breakage.</p><p><strong>Conclusion: </strong>Malalignment might have caused mechanical failure, and in vivo kinematic analysis was useful to visualize abnormal kinematics in TKA.</p>\",\"PeriodicalId\":14748,\"journal\":{\"name\":\"JBJS case connector\",\"volume\":\"15 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2025-01-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"JBJS case connector\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.2106/JBJS.CC.24.00466\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"JBJS case connector","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2106/JBJS.CC.24.00466","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
Detection of Multiple Tibial Malalignment-Induced Early Polyethylene Breakage Using Single Plane Fluoroscopy: A Case Report.
Case: A 75-year-old woman underwent cruciate retaining total knee arthroplasty (TKA) for osteonecrosis of the knee, converted from unicompartmental arthroplasty due to an intraoperative medial tibial plateau fracture. Four years later, she experienced persistent knee pain. Computed tomography revealed tibial malalignments in 2 planes. In vivo kinematic analysis using single-plane fluoroscopy, which can visualize the femorotibial contact, indicated varus thrust with lateral subluxation of the tibial component and the femoral component recessing into the polyethylene insert. Revision TKA affirmed the polyethylene insert breakage.
Conclusion: Malalignment might have caused mechanical failure, and in vivo kinematic analysis was useful to visualize abnormal kinematics in TKA.
期刊介绍:
JBJS Case Connector helps improve patient care by providing the medical community with a journal that harnesses technology to provide information tools for discovery and reporting of unusual musculoskeletal problems, findings, treatment, and outcomes. Co-edited by Thomas W. Bauer, MD, PhD, and Ronald W. Lindsey, MD, JBJS Case Connector assists orthopaedic surgeons in the search for precedents, connections, and trends in their efforts to improve patient care. Using this unique journal, surgeons can find the commonalities between cases, benefit from the experience of their peers, and filter case information by many important variables in order to provide the best possible care for orthopaedic patients. This cross-referenced online journal includes thousands of orthopaedic case reports. It compiles symptoms, conditions, and demographic details to empower surgeons to find cases similar to theirs. Surgeons can mine the database to reveal emerging trends and identify patterns, distinguishing between truly rare cases and repeated, related single instances of a larger problem. The JBJS Case Connector Image Quiz feature provides interactive quizzes based on images from content published by JBJS and includes a discussion area for further exploration of ideas and concepts. The JBJS Image Quiz app for iPad and iPhone is available in the App Store. Contributions to JBJS Case Connector are welcomed from anywhere in the world and are considered on their merits. Articles must be written in English and should be submitted as outlined in the Instructions to Authors. All authors must abide by the JBJS ethics policies and all submissions to JBJS Case Connector are covered by the JBJS embargo policy.