{"title":"使用磁共振光谱法测量聚乙烯吡咯烷酮(PVP)溶液中的温度。","authors":"Neville D. Gai, Ruifeng Dong, Jan Willem van der Veen, Ronald Ouwerkerk, Carlo Pierpaoli","doi":"10.1002/mp.17683","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>Polyvinylpyrrolidone (PVP) water solutions could be used for cross-site and cross-vendor validation of diffusion-related measurements. However, since water diffusivity varies as a function of temperature, knowing the temperature of the PVP solution at the time of the measurement is fundamental in accomplishing this task.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>MR spectroscopy (MRS) could provide absolute temperature measurements since the water peak moves relative to any stable peak as temperature changes. In this work, the PVP proton spectrum was investigated to see if any stable peaks would allow for temperature determination. Reproducibility and repeatability for three scanners from three vendors were also assessed.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>A spherical 17 cm container filled with 40% PVP w/w in distilled water was used for the experiments. A Point REsolved Spectroscopy Sequence (PRESS) with water suppression was employed on three 3T scanners from different vendors—GE, Siemens, and Philips. Frequency separation (in ppm) between peaks was measured in a voxel at the location of a fiber optic temperature probe and mapped to probe measured temperature. The center peak of the first methylene proton triplet closest to water peak was selected for analysis in jMRUI due to its ease of identification and echo time shift invariance. Shift in ppm of the central methylene peak proton was mapped against measured temperatures.</p>\n \n <p>Repeatability and reproducibility across the three scanners were determined at room temperature using 10 repeated PRESS scans. MRS established ppm shift versus temperature relationship was used to predict temperature in different PVP phantoms which were then compared against fiber optic probe measured temperature values.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Several <sup>1</sup>H peaks were identified on all scans of the PVP phantom. The water peak moved by ∼−0.01 ppm/°C on the three scanners relative to a central methylene peak. The maximum mean absolute temperature difference over a temperature range of 18–35°C between the three scanners was 0.16°C while the minimum was 0.057°C. Repeatability on each scanner was excellent (std range: 0.00–0.14°C) over 10 repeated PRESS scans. Reproducibility across the three scanners was also excellent with mean temperature difference between scanners ranging between 0.1 and 0.4°C. Temperature values from MRS were within prediction bounds on the three scanners for another in-house prepared 40% PVP phantom (maximum difference<0.3°C), while they were consistently overestimated for another 30% PVP phantom (<1°C) and underestimated for a CaliberMRI 40% PVP phantom (<2.8°C).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>PVP solutions exhibit stable proton peaks, one of which was used for assessing the temperature of the solution using MR proton spectroscopy. These measurements are fast and feasible with standard sequences and postprocessing MRS software and provide fundamental information for calibration of diffusion MRI using PVP solutions.</p>\n </section>\n </div>","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 5","pages":"3151-3160"},"PeriodicalIF":3.2000,"publicationDate":"2025-02-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Measuring temperature in polyvinylpyrrolidone (PVP) solutions using MR spectroscopy\",\"authors\":\"Neville D. Gai, Ruifeng Dong, Jan Willem van der Veen, Ronald Ouwerkerk, Carlo Pierpaoli\",\"doi\":\"10.1002/mp.17683\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>Polyvinylpyrrolidone (PVP) water solutions could be used for cross-site and cross-vendor validation of diffusion-related measurements. However, since water diffusivity varies as a function of temperature, knowing the temperature of the PVP solution at the time of the measurement is fundamental in accomplishing this task.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Purpose</h3>\\n \\n <p>MR spectroscopy (MRS) could provide absolute temperature measurements since the water peak moves relative to any stable peak as temperature changes. In this work, the PVP proton spectrum was investigated to see if any stable peaks would allow for temperature determination. Reproducibility and repeatability for three scanners from three vendors were also assessed.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>A spherical 17 cm container filled with 40% PVP w/w in distilled water was used for the experiments. A Point REsolved Spectroscopy Sequence (PRESS) with water suppression was employed on three 3T scanners from different vendors—GE, Siemens, and Philips. Frequency separation (in ppm) between peaks was measured in a voxel at the location of a fiber optic temperature probe and mapped to probe measured temperature. The center peak of the first methylene proton triplet closest to water peak was selected for analysis in jMRUI due to its ease of identification and echo time shift invariance. Shift in ppm of the central methylene peak proton was mapped against measured temperatures.</p>\\n \\n <p>Repeatability and reproducibility across the three scanners were determined at room temperature using 10 repeated PRESS scans. MRS established ppm shift versus temperature relationship was used to predict temperature in different PVP phantoms which were then compared against fiber optic probe measured temperature values.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>Several <sup>1</sup>H peaks were identified on all scans of the PVP phantom. The water peak moved by ∼−0.01 ppm/°C on the three scanners relative to a central methylene peak. The maximum mean absolute temperature difference over a temperature range of 18–35°C between the three scanners was 0.16°C while the minimum was 0.057°C. Repeatability on each scanner was excellent (std range: 0.00–0.14°C) over 10 repeated PRESS scans. Reproducibility across the three scanners was also excellent with mean temperature difference between scanners ranging between 0.1 and 0.4°C. Temperature values from MRS were within prediction bounds on the three scanners for another in-house prepared 40% PVP phantom (maximum difference<0.3°C), while they were consistently overestimated for another 30% PVP phantom (<1°C) and underestimated for a CaliberMRI 40% PVP phantom (<2.8°C).</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>PVP solutions exhibit stable proton peaks, one of which was used for assessing the temperature of the solution using MR proton spectroscopy. These measurements are fast and feasible with standard sequences and postprocessing MRS software and provide fundamental information for calibration of diffusion MRI using PVP solutions.</p>\\n </section>\\n </div>\",\"PeriodicalId\":18384,\"journal\":{\"name\":\"Medical physics\",\"volume\":\"52 5\",\"pages\":\"3151-3160\"},\"PeriodicalIF\":3.2000,\"publicationDate\":\"2025-02-17\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://aapm.onlinelibrary.wiley.com/doi/10.1002/mp.17683\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"3","ListUrlMain":"https://aapm.onlinelibrary.wiley.com/doi/10.1002/mp.17683","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Measuring temperature in polyvinylpyrrolidone (PVP) solutions using MR spectroscopy
Background
Polyvinylpyrrolidone (PVP) water solutions could be used for cross-site and cross-vendor validation of diffusion-related measurements. However, since water diffusivity varies as a function of temperature, knowing the temperature of the PVP solution at the time of the measurement is fundamental in accomplishing this task.
Purpose
MR spectroscopy (MRS) could provide absolute temperature measurements since the water peak moves relative to any stable peak as temperature changes. In this work, the PVP proton spectrum was investigated to see if any stable peaks would allow for temperature determination. Reproducibility and repeatability for three scanners from three vendors were also assessed.
Methods
A spherical 17 cm container filled with 40% PVP w/w in distilled water was used for the experiments. A Point REsolved Spectroscopy Sequence (PRESS) with water suppression was employed on three 3T scanners from different vendors—GE, Siemens, and Philips. Frequency separation (in ppm) between peaks was measured in a voxel at the location of a fiber optic temperature probe and mapped to probe measured temperature. The center peak of the first methylene proton triplet closest to water peak was selected for analysis in jMRUI due to its ease of identification and echo time shift invariance. Shift in ppm of the central methylene peak proton was mapped against measured temperatures.
Repeatability and reproducibility across the three scanners were determined at room temperature using 10 repeated PRESS scans. MRS established ppm shift versus temperature relationship was used to predict temperature in different PVP phantoms which were then compared against fiber optic probe measured temperature values.
Results
Several 1H peaks were identified on all scans of the PVP phantom. The water peak moved by ∼−0.01 ppm/°C on the three scanners relative to a central methylene peak. The maximum mean absolute temperature difference over a temperature range of 18–35°C between the three scanners was 0.16°C while the minimum was 0.057°C. Repeatability on each scanner was excellent (std range: 0.00–0.14°C) over 10 repeated PRESS scans. Reproducibility across the three scanners was also excellent with mean temperature difference between scanners ranging between 0.1 and 0.4°C. Temperature values from MRS were within prediction bounds on the three scanners for another in-house prepared 40% PVP phantom (maximum difference<0.3°C), while they were consistently overestimated for another 30% PVP phantom (<1°C) and underestimated for a CaliberMRI 40% PVP phantom (<2.8°C).
Conclusions
PVP solutions exhibit stable proton peaks, one of which was used for assessing the temperature of the solution using MR proton spectroscopy. These measurements are fast and feasible with standard sequences and postprocessing MRS software and provide fundamental information for calibration of diffusion MRI using PVP solutions.
期刊介绍:
Medical Physics publishes original, high impact physics, imaging science, and engineering research that advances patient diagnosis and therapy through contributions in 1) Basic science developments with high potential for clinical translation 2) Clinical applications of cutting edge engineering and physics innovations 3) Broadly applicable and innovative clinical physics developments
Medical Physics is a journal of global scope and reach. By publishing in Medical Physics your research will reach an international, multidisciplinary audience including practicing medical physicists as well as physics- and engineering based translational scientists. We work closely with authors of promising articles to improve their quality.