三种无创血氧仪在健康志愿者中的同时比较。

Medical instrumentation Pub Date : 1987-06-01
Y Mendelson, J C Kent, A Shahnarian, G W Welch, R M Giasi
{"title":"三种无创血氧仪在健康志愿者中的同时比较。","authors":"Y Mendelson,&nbsp;J C Kent,&nbsp;A Shahnarian,&nbsp;G W Welch,&nbsp;R M Giasi","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Noninvasive monitoring of arterial hemoglobin oxygen saturation (SaO2) from the Nellcor N-100 and the Ohmeda-Biox 3700 pulse oximeters were compared with SaO2 measured simultaneously by the Hewlett-Packard 47201A ear oximeter. A total of 868 pairs of data points ranging in saturation from 55-100% were obtained from 31 healthy adult volunteers of different ages, sex, skin pigmentation, height, weight, hematocrit, and smoking habits. Steady state and rapidly changing hypoxic conditions were achieved by varying the inspired O2 concentration between 10 and 100%. Measurements were analyzed by means of linear regression. The equations for the best fitted linear regression line between the Nellcor and the Ohmeda-Biox pulse oximeters as compared to the reference Hewlett-Packard ear oximeter readings were Y = 0.92X + 7.45 and Y = 1.10X - 10.73, respectively. Our study showed that despite a very close correlation among these three oximeters (r = 0.98), when compared to the Hewlett-Packard ear oximeter, the Nellcor pulse oximeter overestimates SaO2 by 2-4%, whereas the Ohmeda-Biox pulse oximeter underestimates SaO2 by 3-6% for SaO2 levels between 70 and 55%.</p>","PeriodicalId":76133,"journal":{"name":"Medical instrumentation","volume":"21 3","pages":"183-8"},"PeriodicalIF":0.0000,"publicationDate":"1987-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Simultaneous comparison of three noninvasive oximeters in healthy volunteers.\",\"authors\":\"Y Mendelson,&nbsp;J C Kent,&nbsp;A Shahnarian,&nbsp;G W Welch,&nbsp;R M Giasi\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Noninvasive monitoring of arterial hemoglobin oxygen saturation (SaO2) from the Nellcor N-100 and the Ohmeda-Biox 3700 pulse oximeters were compared with SaO2 measured simultaneously by the Hewlett-Packard 47201A ear oximeter. A total of 868 pairs of data points ranging in saturation from 55-100% were obtained from 31 healthy adult volunteers of different ages, sex, skin pigmentation, height, weight, hematocrit, and smoking habits. Steady state and rapidly changing hypoxic conditions were achieved by varying the inspired O2 concentration between 10 and 100%. Measurements were analyzed by means of linear regression. The equations for the best fitted linear regression line between the Nellcor and the Ohmeda-Biox pulse oximeters as compared to the reference Hewlett-Packard ear oximeter readings were Y = 0.92X + 7.45 and Y = 1.10X - 10.73, respectively. Our study showed that despite a very close correlation among these three oximeters (r = 0.98), when compared to the Hewlett-Packard ear oximeter, the Nellcor pulse oximeter overestimates SaO2 by 2-4%, whereas the Ohmeda-Biox pulse oximeter underestimates SaO2 by 3-6% for SaO2 levels between 70 and 55%.</p>\",\"PeriodicalId\":76133,\"journal\":{\"name\":\"Medical instrumentation\",\"volume\":\"21 3\",\"pages\":\"183-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1987-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical instrumentation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical instrumentation","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

将Nellcor N-100和Ohmeda-Biox 3700脉搏血氧仪无创监测的动脉血红蛋白氧饱和度(SaO2)与Hewlett-Packard 47201A耳血氧仪同时测量的SaO2进行比较。从31名不同年龄、性别、皮肤色素、身高、体重、红细胞压积和吸烟习惯的健康成年志愿者中获得饱和度为55-100%的868对数据点。通过将吸入的O2浓度在10%到100%之间变化,可以实现稳态和快速变化的缺氧条件。测量结果采用线性回归分析。与参考hp耳血氧仪读数相比,Nellcor和Ohmeda-Biox脉搏血氧仪之间的最佳拟合线性回归方程分别为Y = 0.92X + 7.45和Y = 1.10X - 10.73。我们的研究表明,尽管这三种血氧仪之间的相关性非常密切(r = 0.98),但与惠普耳式血氧仪相比,Nellcor脉搏血氧仪对SaO2的高估幅度为2-4%,而Ohmeda-Biox脉搏血氧仪在SaO2水平在70 - 55%之间时低估了3-6%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Simultaneous comparison of three noninvasive oximeters in healthy volunteers.

Noninvasive monitoring of arterial hemoglobin oxygen saturation (SaO2) from the Nellcor N-100 and the Ohmeda-Biox 3700 pulse oximeters were compared with SaO2 measured simultaneously by the Hewlett-Packard 47201A ear oximeter. A total of 868 pairs of data points ranging in saturation from 55-100% were obtained from 31 healthy adult volunteers of different ages, sex, skin pigmentation, height, weight, hematocrit, and smoking habits. Steady state and rapidly changing hypoxic conditions were achieved by varying the inspired O2 concentration between 10 and 100%. Measurements were analyzed by means of linear regression. The equations for the best fitted linear regression line between the Nellcor and the Ohmeda-Biox pulse oximeters as compared to the reference Hewlett-Packard ear oximeter readings were Y = 0.92X + 7.45 and Y = 1.10X - 10.73, respectively. Our study showed that despite a very close correlation among these three oximeters (r = 0.98), when compared to the Hewlett-Packard ear oximeter, the Nellcor pulse oximeter overestimates SaO2 by 2-4%, whereas the Ohmeda-Biox pulse oximeter underestimates SaO2 by 3-6% for SaO2 levels between 70 and 55%.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Physical fitness in the elderly. Anesthesia delivery systems. Effects of using potassium adsorption filters on saline-filled and saline-removed methods for the removal of potassium from red blood cell solutions Plasma viscosity and whole blood viscosity as diagnostic tools of blood abnormalities by using simple syringe method A technique for measuring spatial distribution of electromagnetic fields created by radio-frequency coils in MRI scanners
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1