{"title":"自动免疫化学分析仪Auto ICS的评价。","authors":"M C Smith","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The Beckman Instruments' Automated Immunochemistry System, the Auto ICS is a fully automated version of the ICS. This evaluation examines several analytical and cost parameters for five representative assays: IgG, IgA, IgM, phenytoin, and phenobarbital. Correlation with more established methods was excellent, with the coefficients of correlation being 0.9880, 0.9948, 0.9862, 0.9812, and 0.9618 respectively. Intra- and interassay precision was on a par with other immunoassays, with CVs varying from 1.8% to 9.5%. Linearity of each assay was verified. The instrument can be cost-effective, if the simultaneous assay feature is used on batches of at least six samples. The instrument requires very little operator maintenance.</p>","PeriodicalId":76595,"journal":{"name":"The American journal of medical technology","volume":"49 12","pages":"885-92"},"PeriodicalIF":0.0000,"publicationDate":"1983-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of an automated immunochemistry analyzer, the Auto ICS.\",\"authors\":\"M C Smith\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The Beckman Instruments' Automated Immunochemistry System, the Auto ICS is a fully automated version of the ICS. This evaluation examines several analytical and cost parameters for five representative assays: IgG, IgA, IgM, phenytoin, and phenobarbital. Correlation with more established methods was excellent, with the coefficients of correlation being 0.9880, 0.9948, 0.9862, 0.9812, and 0.9618 respectively. Intra- and interassay precision was on a par with other immunoassays, with CVs varying from 1.8% to 9.5%. Linearity of each assay was verified. The instrument can be cost-effective, if the simultaneous assay feature is used on batches of at least six samples. The instrument requires very little operator maintenance.</p>\",\"PeriodicalId\":76595,\"journal\":{\"name\":\"The American journal of medical technology\",\"volume\":\"49 12\",\"pages\":\"885-92\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1983-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The American journal of medical technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The American journal of medical technology","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Evaluation of an automated immunochemistry analyzer, the Auto ICS.
The Beckman Instruments' Automated Immunochemistry System, the Auto ICS is a fully automated version of the ICS. This evaluation examines several analytical and cost parameters for five representative assays: IgG, IgA, IgM, phenytoin, and phenobarbital. Correlation with more established methods was excellent, with the coefficients of correlation being 0.9880, 0.9948, 0.9862, 0.9812, and 0.9618 respectively. Intra- and interassay precision was on a par with other immunoassays, with CVs varying from 1.8% to 9.5%. Linearity of each assay was verified. The instrument can be cost-effective, if the simultaneous assay feature is used on batches of at least six samples. The instrument requires very little operator maintenance.