{"title":"聚维酮心脏起搏导线涂层材料的安全性评价及生物分布。","authors":"M Jay, G A Digenis","doi":"10.3109/10731198309118811","DOIUrl":null,"url":null,"abstract":"<p><p>Cardiac pacing leads coated with povidone-[131I] were implanted in dogs and the leaching of radioactivity from the leads was monitored by external scintigraphy. The activity which had dissipated from the pacing leads was not as [131I]-iodide, but as povidone-[131I]. Only 50% (mean) of the activity remained on the pacing leads after two weeks while a significant amount of radioactivity was eliminated via urine and feces. The liver was a major site of accumulation of retained activity which had leached off the pacing leads. There was no evidence of large pieces of povidone-[131I] in the lungs of the dogs, all of which appeared healthy at the time of sacrifice. The results of this study support the conclusions of a long-term study indicating that povidone is a safe and suitable coating material for pacing leads.</p>","PeriodicalId":75597,"journal":{"name":"Biomaterials, medical devices, and artificial organs","volume":"11 4","pages":"253-8"},"PeriodicalIF":0.0000,"publicationDate":"1983-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.3109/10731198309118811","citationCount":"2","resultStr":"{\"title\":\"Safety assessment and biodistribution of povidone as a coating material for cardiac pacing leads.\",\"authors\":\"M Jay, G A Digenis\",\"doi\":\"10.3109/10731198309118811\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Cardiac pacing leads coated with povidone-[131I] were implanted in dogs and the leaching of radioactivity from the leads was monitored by external scintigraphy. The activity which had dissipated from the pacing leads was not as [131I]-iodide, but as povidone-[131I]. Only 50% (mean) of the activity remained on the pacing leads after two weeks while a significant amount of radioactivity was eliminated via urine and feces. The liver was a major site of accumulation of retained activity which had leached off the pacing leads. There was no evidence of large pieces of povidone-[131I] in the lungs of the dogs, all of which appeared healthy at the time of sacrifice. The results of this study support the conclusions of a long-term study indicating that povidone is a safe and suitable coating material for pacing leads.</p>\",\"PeriodicalId\":75597,\"journal\":{\"name\":\"Biomaterials, medical devices, and artificial organs\",\"volume\":\"11 4\",\"pages\":\"253-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1983-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.3109/10731198309118811\",\"citationCount\":\"2\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Biomaterials, medical devices, and artificial organs\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.3109/10731198309118811\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Biomaterials, medical devices, and artificial organs","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3109/10731198309118811","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Safety assessment and biodistribution of povidone as a coating material for cardiac pacing leads.
Cardiac pacing leads coated with povidone-[131I] were implanted in dogs and the leaching of radioactivity from the leads was monitored by external scintigraphy. The activity which had dissipated from the pacing leads was not as [131I]-iodide, but as povidone-[131I]. Only 50% (mean) of the activity remained on the pacing leads after two weeks while a significant amount of radioactivity was eliminated via urine and feces. The liver was a major site of accumulation of retained activity which had leached off the pacing leads. There was no evidence of large pieces of povidone-[131I] in the lungs of the dogs, all of which appeared healthy at the time of sacrifice. The results of this study support the conclusions of a long-term study indicating that povidone is a safe and suitable coating material for pacing leads.