{"title":"85兆电子伏特临床质子束水热测量","authors":"H. Palmans ,&nbsp;J. Seuntjens ,&nbsp;F. Verhaegen ,&nbsp;J.M. Denis ,&nbsp;S. Vynckier ,&nbsp;H. Thierens","doi":"10.1016/0924-4212(96)84907-3","DOIUrl":null,"url":null,"abstract":"<div><p>The importance of water calorimetry in clinieal proton beam dosimetry has increased for various reasons: there has been an increasing interest in proton therapy from the radiotherapy world; existing dosimetry protocols for clinical proton beams recommend calorimetry as the primary dosimetry method (ie, AAPM and ECHED); water has recently become the reference material for dose specification. For the water calorimetric dose evaluation, the water calorimeter developed in Ghent was used. Ionometry was performed following the ECHED protocol. The study resulted in a calorimetric to ionometric dose ratio of 0.974<!--> <!-->±<!--> <!-->0.009. The discrepancy should in our opinion be attributed to the (<em>W/e</em>)<sub>p,air</sub> value of 35.2 J/C implemented in the protocol, possibly next to ion chamber dependent effects for which indications are found in the ionometry measurements.</p></div><div><p>Le nombre de centres de protonthérapie a considérablement augmenté dans le monde et les résultats cliniques sont prometteurs. Plusieurs protocoles dosimétriques existant (par exemple AAPM et ECHED) recommandent le calorimètre comme dosimètre de référence. De plus, il est couramment recommandé de prendre l’eau comme milieu de référence pour la spécification de la dose. Ceci explique l’intérêt porté au calorimètre à eau pour la dosimetrie clinique en protonthérapie. Les doses ont été mesurées, dans un faisceau de protons, d’une part au moyen du calorimètre à eau mis au point à l’Université de Gand et d’autre part au moyen d’une chambre d’ionisation selon le protocole de l’ECHED. Le rapport des doses mesurées par calorimetrie et ionométrie est de 0,974<!--> <!-->±<!--> <!-->0,009. Cette discordance pourrait être liée au choix de la valeur 35,2 J/C admise pour la grandeur (W/e)<sub>p air</sub> prévue dans le protocole. Elle pourrait également provenir de facteurs liés au type de chambre d’ionisation.</p></div>","PeriodicalId":79321,"journal":{"name":"Bulletin du cancer. Radiotherapie : journal de la Societe francaise du cancer : organe de la societe francaise de radiotherapie oncologique","volume":"83 ","pages":"Pages 176s-178s"},"PeriodicalIF":0.0000,"publicationDate":"1996-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/0924-4212(96)84907-3","citationCount":"14","resultStr":"{\"title\":\"Water calorimetry measurements in a 85 MeV clinical proton beam\",\"authors\":\"H. Palmans ,&nbsp;J. Seuntjens ,&nbsp;F. Verhaegen ,&nbsp;J.M. Denis ,&nbsp;S. Vynckier ,&nbsp;H. Thierens\",\"doi\":\"10.1016/0924-4212(96)84907-3\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>The importance of water calorimetry in clinieal proton beam dosimetry has increased for various reasons: there has been an increasing interest in proton therapy from the radiotherapy world; existing dosimetry protocols for clinical proton beams recommend calorimetry as the primary dosimetry method (ie, AAPM and ECHED); water has recently become the reference material for dose specification. For the water calorimetric dose evaluation, the water calorimeter developed in Ghent was used. Ionometry was performed following the ECHED protocol. The study resulted in a calorimetric to ionometric dose ratio of 0.974<!--> <!-->±<!--> <!-->0.009. The discrepancy should in our opinion be attributed to the (<em>W/e</em>)<sub>p,air</sub> value of 35.2 J/C implemented in the protocol, possibly next to ion chamber dependent effects for which indications are found in the ionometry measurements.</p></div><div><p>Le nombre de centres de protonthérapie a considérablement augmenté dans le monde et les résultats cliniques sont prometteurs. Plusieurs protocoles dosimétriques existant (par exemple AAPM et ECHED) recommandent le calorimètre comme dosimètre de référence. De plus, il est couramment recommandé de prendre l’eau comme milieu de référence pour la spécification de la dose. Ceci explique l’intérêt porté au calorimètre à eau pour la dosimetrie clinique en protonthérapie. Les doses ont été mesurées, dans un faisceau de protons, d’une part au moyen du calorimètre à eau mis au point à l’Université de Gand et d’autre part au moyen d’une chambre d’ionisation selon le protocole de l’ECHED. Le rapport des doses mesurées par calorimetrie et ionométrie est de 0,974<!--> <!-->±<!--> <!-->0,009. Cette discordance pourrait être liée au choix de la valeur 35,2 J/C admise pour la grandeur (W/e)<sub>p air</sub> prévue dans le protocole. Elle pourrait également provenir de facteurs liés au type de chambre d’ionisation.</p></div>\",\"PeriodicalId\":79321,\"journal\":{\"name\":\"Bulletin du cancer. Radiotherapie : journal de la Societe francaise du cancer : organe de la societe francaise de radiotherapie oncologique\",\"volume\":\"83 \",\"pages\":\"Pages 176s-178s\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1996-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/0924-4212(96)84907-3\",\"citationCount\":\"14\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bulletin du cancer. Radiotherapie : journal de la Societe francaise du cancer : organe de la societe francaise de radiotherapie oncologique\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/0924421296849073\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bulletin du cancer. Radiotherapie : journal de la Societe francaise du cancer : organe de la societe francaise de radiotherapie oncologique","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/0924421296849073","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 14

摘要

由于各种原因,水热法在临床质子束剂量测定中的重要性有所增加:放射治疗界对质子治疗的兴趣日益增加;现有临床质子束剂量测定方案推荐量热法作为主要剂量测定方法(即AAPM和ECHED);近年来,水已成为剂量标准的参考物质。水热剂量评价采用根特研制的水热计。离子化测量按照ECHED方案进行。研究结果表明,量热剂量比为0.974±0.009。我们认为,这种差异应归因于协议中实施的(W/e)p,空气值为35.2 J/C,可能紧挨着离子室依赖效应,在离子学测量中发现了迹象。在世界范围内和在世界范围内,都有关于交换交换和交换交换中心的规定。现有的多种协议(例如AAPM和ECHED)推荐的卡路里摄入量为1 / 3。德+ il est couramment recommande麦克风威尼斯这样的环境De参考倒拉规范De la剂量。Ceci excique 1 'intérêt卡洛丽亚卡洛丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡丽亚卡。Les dose ont 2013.2013.10, dans unfaisceau de质子,d 'une part au moyen du calorimitre, eau mis au point, l ' universitede de grande, d ' tre part au moyen d ' ne chamber d ' ionation selle protocol de l 'ECHED。Le rapport des剂量测量与热量计相比,在0,974±0,009之间进行了测量。Cette discorance pourrait être liassei au choix de la valeur 35,2 J/C admit pour la grandeur (W/e)p air pracemue dans le protocole。在电离室类型中,有一种因素是可变的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Water calorimetry measurements in a 85 MeV clinical proton beam

The importance of water calorimetry in clinieal proton beam dosimetry has increased for various reasons: there has been an increasing interest in proton therapy from the radiotherapy world; existing dosimetry protocols for clinical proton beams recommend calorimetry as the primary dosimetry method (ie, AAPM and ECHED); water has recently become the reference material for dose specification. For the water calorimetric dose evaluation, the water calorimeter developed in Ghent was used. Ionometry was performed following the ECHED protocol. The study resulted in a calorimetric to ionometric dose ratio of 0.974 ± 0.009. The discrepancy should in our opinion be attributed to the (W/e)p,air value of 35.2 J/C implemented in the protocol, possibly next to ion chamber dependent effects for which indications are found in the ionometry measurements.

Le nombre de centres de protonthérapie a considérablement augmenté dans le monde et les résultats cliniques sont prometteurs. Plusieurs protocoles dosimétriques existant (par exemple AAPM et ECHED) recommandent le calorimètre comme dosimètre de référence. De plus, il est couramment recommandé de prendre l’eau comme milieu de référence pour la spécification de la dose. Ceci explique l’intérêt porté au calorimètre à eau pour la dosimetrie clinique en protonthérapie. Les doses ont été mesurées, dans un faisceau de protons, d’une part au moyen du calorimètre à eau mis au point à l’Université de Gand et d’autre part au moyen d’une chambre d’ionisation selon le protocole de l’ECHED. Le rapport des doses mesurées par calorimetrie et ionométrie est de 0,974 ± 0,009. Cette discordance pourrait être liée au choix de la valeur 35,2 J/C admise pour la grandeur (W/e)p air prévue dans le protocole. Elle pourrait également provenir de facteurs liés au type de chambre d’ionisation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
The proton therapy system for MGH's NPTC: equipment description and progress report. [Apropos of a recent trip to Quebec]. [Incidence and risk factors of second primary cancers in patients treated for primary pharyngeal cancer]. [Prevention of second primary cancer with vitamin supplementation in patients treated for head and neck cancers]. [Surgery of the hypopharynx after radiotherapy].
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1