{"title":"柔性内窥镜自动洗涤/消毒器常规微生物控制的新型测试模型。","authors":"H K Geiss, C Keller, A Schu, A Wiebel","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Concerns exist about effective methods for endoscope cleaning, disinfection and sterilization. The aim of this study was to establish a model for controlling the function of automatic washer/disinfectors for flexible endoscopes under routine conditions. The team developed a dummy endoscope channel system representing two complete gastroscopes and a colonoscope, with channels joined together and fitted with adaptors to allow the models to fit into different washer/disinfectors. The models were artificially contaminated with various combinations of four test organisms and coagulating blood and run through test cycles in different washer/disinfectors. A total of 24 test runs were performed. The reduction of a given microbial burden by at least five log steps is an established measure of efficacy of a decontamination procedure. When the two challenge tests (45 channels) and the six positive controls are excluded, only 13/309 channels (4.2%) failed the proof of efficacy. In contrast, the failure rate in the challenge test was 65.3%. The difference of the final bioburden is statistically significant. The data suggest this test model could be regarded as a first step in rational and reliable biological control of flexible endoscope reprocessing.</p>","PeriodicalId":79670,"journal":{"name":"Journal of healthcare materiel management","volume":"12 8","pages":"36-42"},"PeriodicalIF":0.0000,"publicationDate":"1994-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A novel test model for routine microbiological control of automated washer/disinfectors for flexible endoscopes.\",\"authors\":\"H K Geiss, C Keller, A Schu, A Wiebel\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Concerns exist about effective methods for endoscope cleaning, disinfection and sterilization. The aim of this study was to establish a model for controlling the function of automatic washer/disinfectors for flexible endoscopes under routine conditions. The team developed a dummy endoscope channel system representing two complete gastroscopes and a colonoscope, with channels joined together and fitted with adaptors to allow the models to fit into different washer/disinfectors. The models were artificially contaminated with various combinations of four test organisms and coagulating blood and run through test cycles in different washer/disinfectors. A total of 24 test runs were performed. The reduction of a given microbial burden by at least five log steps is an established measure of efficacy of a decontamination procedure. When the two challenge tests (45 channels) and the six positive controls are excluded, only 13/309 channels (4.2%) failed the proof of efficacy. In contrast, the failure rate in the challenge test was 65.3%. The difference of the final bioburden is statistically significant. The data suggest this test model could be regarded as a first step in rational and reliable biological control of flexible endoscope reprocessing.</p>\",\"PeriodicalId\":79670,\"journal\":{\"name\":\"Journal of healthcare materiel management\",\"volume\":\"12 8\",\"pages\":\"36-42\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1994-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of healthcare materiel management\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of healthcare materiel management","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
A novel test model for routine microbiological control of automated washer/disinfectors for flexible endoscopes.
Concerns exist about effective methods for endoscope cleaning, disinfection and sterilization. The aim of this study was to establish a model for controlling the function of automatic washer/disinfectors for flexible endoscopes under routine conditions. The team developed a dummy endoscope channel system representing two complete gastroscopes and a colonoscope, with channels joined together and fitted with adaptors to allow the models to fit into different washer/disinfectors. The models were artificially contaminated with various combinations of four test organisms and coagulating blood and run through test cycles in different washer/disinfectors. A total of 24 test runs were performed. The reduction of a given microbial burden by at least five log steps is an established measure of efficacy of a decontamination procedure. When the two challenge tests (45 channels) and the six positive controls are excluded, only 13/309 channels (4.2%) failed the proof of efficacy. In contrast, the failure rate in the challenge test was 65.3%. The difference of the final bioburden is statistically significant. The data suggest this test model could be regarded as a first step in rational and reliable biological control of flexible endoscope reprocessing.