Ying-Ming Wang, Cheng-Show Chen, Nien-Chun Chung, Xuan-De Ye, Kang Liu
{"title":"小儿胸椎手术完成后体位改变时呼吸回路突然完全阻塞。","authors":"Ying-Ming Wang, Cheng-Show Chen, Nien-Chun Chung, Xuan-De Ye, Kang Liu","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Positioning patients during the perioperative period is a common event to anesthesiologists. A variety of complications may arise during this cumbrous moment. We describe a rare circumstance that we came across immediately after changing the posture of a pediatric patient at the end of thoracic spinal surgery. A total occlusion of the combined heat-moisture exchanger (HME) and bacterial/viral filter was responsible for breathing circuit obstruction. The use of the combined HME and bacterial/viral filter is not entirely riskless during general anesthesia.</p>","PeriodicalId":79312,"journal":{"name":"Acta anaesthesiologica Sinica","volume":"41 3","pages":"145-8"},"PeriodicalIF":0.0000,"publicationDate":"2003-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Sudden complete obstruction of breathing circuit during postural change upon completion of thoracic spinal surgery in a pediatric patient.\",\"authors\":\"Ying-Ming Wang, Cheng-Show Chen, Nien-Chun Chung, Xuan-De Ye, Kang Liu\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Positioning patients during the perioperative period is a common event to anesthesiologists. A variety of complications may arise during this cumbrous moment. We describe a rare circumstance that we came across immediately after changing the posture of a pediatric patient at the end of thoracic spinal surgery. A total occlusion of the combined heat-moisture exchanger (HME) and bacterial/viral filter was responsible for breathing circuit obstruction. The use of the combined HME and bacterial/viral filter is not entirely riskless during general anesthesia.</p>\",\"PeriodicalId\":79312,\"journal\":{\"name\":\"Acta anaesthesiologica Sinica\",\"volume\":\"41 3\",\"pages\":\"145-8\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2003-09-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta anaesthesiologica Sinica\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta anaesthesiologica Sinica","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Sudden complete obstruction of breathing circuit during postural change upon completion of thoracic spinal surgery in a pediatric patient.
Positioning patients during the perioperative period is a common event to anesthesiologists. A variety of complications may arise during this cumbrous moment. We describe a rare circumstance that we came across immediately after changing the posture of a pediatric patient at the end of thoracic spinal surgery. A total occlusion of the combined heat-moisture exchanger (HME) and bacterial/viral filter was responsible for breathing circuit obstruction. The use of the combined HME and bacterial/viral filter is not entirely riskless during general anesthesia.