{"title":"用聚乙烯和全陶瓷插入物在金属背衬的髋臼杯内对髋臼进行载荷。","authors":"Frederick J Kummer,&nbsp;Kazuho Iesaka,&nbsp;John Rogers,&nbsp;Paul E Di Cesare","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Both static and dynamic loads were applied to metal-backed acetabular cups with ceramic or polyethylene inserts and the resulting load transmissions at a simulated bone interface were determined. Perfect fit and under-sized and over-sized cavities were prepared in simulated bone substrates, lined with Fuji pressure sensitive film, and acetabular cups inserted with physiological loads. The magnitude and location of contact forces between the cup and bone were measured. These cups were then subjected to a controlled impact load and the intensity and frequency of the loads transmitted to the substrate were determined. The results suggest that a polyethylene backing for ceramics is not necessary as there were no major differences in the static and dynamic stresses transmitted to the cup-bone interface with all polyethylene or ceramic inserts.</p>","PeriodicalId":77050,"journal":{"name":"Bulletin (Hospital for Joint Diseases (New York, N.Y.))","volume":"61 3-4","pages":"132-4"},"PeriodicalIF":0.0000,"publicationDate":"2003-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Loading of the acetabulum by polyethylene and all-ceramic inserts in metal-backed acetabular cups.\",\"authors\":\"Frederick J Kummer,&nbsp;Kazuho Iesaka,&nbsp;John Rogers,&nbsp;Paul E Di Cesare\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Both static and dynamic loads were applied to metal-backed acetabular cups with ceramic or polyethylene inserts and the resulting load transmissions at a simulated bone interface were determined. Perfect fit and under-sized and over-sized cavities were prepared in simulated bone substrates, lined with Fuji pressure sensitive film, and acetabular cups inserted with physiological loads. The magnitude and location of contact forces between the cup and bone were measured. These cups were then subjected to a controlled impact load and the intensity and frequency of the loads transmitted to the substrate were determined. The results suggest that a polyethylene backing for ceramics is not necessary as there were no major differences in the static and dynamic stresses transmitted to the cup-bone interface with all polyethylene or ceramic inserts.</p>\",\"PeriodicalId\":77050,\"journal\":{\"name\":\"Bulletin (Hospital for Joint Diseases (New York, N.Y.))\",\"volume\":\"61 3-4\",\"pages\":\"132-4\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2003-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Bulletin (Hospital for Joint Diseases (New York, N.Y.))\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Bulletin (Hospital for Joint Diseases (New York, N.Y.))","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

将静态载荷和动态载荷分别施加于带有陶瓷或聚乙烯嵌套的金属支撑髋臼杯,并确定在模拟骨界面处产生的载荷传递。在模拟骨基质上制备完美贴合、过小和过大的腔体,内衬富士压敏膜,并插入有生理负荷的髋臼杯。测量了杯和骨之间的接触力的大小和位置。然后,这些杯子受到控制的冲击载荷,并确定传递到基材的载荷的强度和频率。结果表明,聚乙烯陶瓷衬垫是不必要的,因为所有聚乙烯或陶瓷衬垫传递到杯骨界面的静态和动态应力没有大的差异。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Loading of the acetabulum by polyethylene and all-ceramic inserts in metal-backed acetabular cups.

Both static and dynamic loads were applied to metal-backed acetabular cups with ceramic or polyethylene inserts and the resulting load transmissions at a simulated bone interface were determined. Perfect fit and under-sized and over-sized cavities were prepared in simulated bone substrates, lined with Fuji pressure sensitive film, and acetabular cups inserted with physiological loads. The magnitude and location of contact forces between the cup and bone were measured. These cups were then subjected to a controlled impact load and the intensity and frequency of the loads transmitted to the substrate were determined. The results suggest that a polyethylene backing for ceramics is not necessary as there were no major differences in the static and dynamic stresses transmitted to the cup-bone interface with all polyethylene or ceramic inserts.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Acetabular Fractures in the Elderly The effect of interference screw diameter on soft tissue graft fixation. The interaction between the whipstitch sutures of multi-strand ACL grafts and interference screw fixation. Slipped capital femoral epiphysis in identical twins: is there an HLA predisposition? Report of a case and review of the literature. Loose bodies in a sublabral recess: diagnosis and treatment.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1