本文采用一种新的重合计算机模拟模型,评价了重合对流式细胞术测定粒细胞-血小板复合物的影响

János Fent, Péter Bihari , József Fűrész, János Hamar , Susan Lakatos
{"title":"本文采用一种新的重合计算机模拟模型,评价了重合对流式细胞术测定粒细胞-血小板复合物的影响","authors":"János Fent,&nbsp;Péter Bihari ,&nbsp;József Fűrész,&nbsp;János Hamar ,&nbsp;Susan Lakatos","doi":"10.1016/j.jbbm.2007.07.008","DOIUrl":null,"url":null,"abstract":"<div><p>Cell complexes composed of two different cells labeled with different fluorophores can be detected as double positive events in the flow cytometer. Double positivity can originate not only from real complexes but from non-interacting coinciding cells as well. Coincidence has a high impact on the determination of the amount of platelet–granulocyte complexes since platelet concentration is in the orders of magnitude higher than that of the granulocytes. A computer model has been developed to simulate coincidence in the flow cytometer to reveal the contribution of coincidence to the overestimation of the total amount of platelet–granulocyte complexes. Mixtures of non-interacting fluorescent beads as well as EDTA-anticoagulated blood samples were analyzed in the flow cytometer.</p><p>An excellent fit was found between computer simulated and measured data pairs. Bead mixture in the flow cytometer and simulation of that resulted in 37.3<!--> <!-->±<!--> <!-->1.3 and 35.7<!--> <!-->±<!--> <!-->0.6% double positivity, respectively. 30.2<!--> <!-->±<!--> <!-->4.3% double positivity was measured for EDTA-anticoagulated blood samples while simulation of that resulted in 28.3<!--> <!-->±<!--> <!-->0.6%. Double positivity attributed to platelet–granulocyte complexes in slightly diluted blood samples might originate in coincidence and not from true complexes.</p></div>","PeriodicalId":15257,"journal":{"name":"Journal of biochemical and biophysical methods","volume":"70 6","pages":"Pages 1086-1090"},"PeriodicalIF":0.0000,"publicationDate":"2008-04-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/j.jbbm.2007.07.008","citationCount":"11","resultStr":"{\"title\":\"Impact of coincidence on granulocyte–platelet complex determination by flow cytometry is evaluated by a novel computer simulation model of coincidence\",\"authors\":\"János Fent,&nbsp;Péter Bihari ,&nbsp;József Fűrész,&nbsp;János Hamar ,&nbsp;Susan Lakatos\",\"doi\":\"10.1016/j.jbbm.2007.07.008\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><p>Cell complexes composed of two different cells labeled with different fluorophores can be detected as double positive events in the flow cytometer. Double positivity can originate not only from real complexes but from non-interacting coinciding cells as well. Coincidence has a high impact on the determination of the amount of platelet–granulocyte complexes since platelet concentration is in the orders of magnitude higher than that of the granulocytes. A computer model has been developed to simulate coincidence in the flow cytometer to reveal the contribution of coincidence to the overestimation of the total amount of platelet–granulocyte complexes. Mixtures of non-interacting fluorescent beads as well as EDTA-anticoagulated blood samples were analyzed in the flow cytometer.</p><p>An excellent fit was found between computer simulated and measured data pairs. Bead mixture in the flow cytometer and simulation of that resulted in 37.3<!--> <!-->±<!--> <!-->1.3 and 35.7<!--> <!-->±<!--> <!-->0.6% double positivity, respectively. 30.2<!--> <!-->±<!--> <!-->4.3% double positivity was measured for EDTA-anticoagulated blood samples while simulation of that resulted in 28.3<!--> <!-->±<!--> <!-->0.6%. Double positivity attributed to platelet–granulocyte complexes in slightly diluted blood samples might originate in coincidence and not from true complexes.</p></div>\",\"PeriodicalId\":15257,\"journal\":{\"name\":\"Journal of biochemical and biophysical methods\",\"volume\":\"70 6\",\"pages\":\"Pages 1086-1090\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2008-04-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1016/j.jbbm.2007.07.008\",\"citationCount\":\"11\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of biochemical and biophysical methods\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0165022X07001510\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of biochemical and biophysical methods","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0165022X07001510","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11

摘要

用不同荧光团标记的两种不同细胞组成的细胞复合物在流式细胞仪上可以检测到双阳性事件。双重正性不仅可以来源于真实的复合体,也可以来源于不相互作用的重合细胞。巧合对血小板-粒细胞复合物数量的测定有很大影响,因为血小板浓度比粒细胞浓度高几个数量级。建立了一个计算机模型来模拟流式细胞仪中的巧合,以揭示巧合对高估血小板-粒细胞复合物总量的贡献。在流式细胞仪上分析非相互作用荧光珠和edta抗凝血样品的混合物。在计算机模拟和测量数据对之间发现了很好的拟合。流式细胞仪和模拟实验结果显示,双阳性分别为37.3±1.3和35.7±0.6%。edta抗凝血双阳性为30.2±4.3%,模拟双阳性为28.3±0.6%。在轻度稀释的血液样本中,血小板-粒细胞复合物的双重阳性可能是巧合,而不是真正的复合物。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Impact of coincidence on granulocyte–platelet complex determination by flow cytometry is evaluated by a novel computer simulation model of coincidence

Cell complexes composed of two different cells labeled with different fluorophores can be detected as double positive events in the flow cytometer. Double positivity can originate not only from real complexes but from non-interacting coinciding cells as well. Coincidence has a high impact on the determination of the amount of platelet–granulocyte complexes since platelet concentration is in the orders of magnitude higher than that of the granulocytes. A computer model has been developed to simulate coincidence in the flow cytometer to reveal the contribution of coincidence to the overestimation of the total amount of platelet–granulocyte complexes. Mixtures of non-interacting fluorescent beads as well as EDTA-anticoagulated blood samples were analyzed in the flow cytometer.

An excellent fit was found between computer simulated and measured data pairs. Bead mixture in the flow cytometer and simulation of that resulted in 37.3 ± 1.3 and 35.7 ± 0.6% double positivity, respectively. 30.2 ± 4.3% double positivity was measured for EDTA-anticoagulated blood samples while simulation of that resulted in 28.3 ± 0.6%. Double positivity attributed to platelet–granulocyte complexes in slightly diluted blood samples might originate in coincidence and not from true complexes.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
An Experience Sampling Study of Physical Activity and Positive Affect: Investigating the Role of Situational Motivation and Perceived Intensity Across Time. Editorial Board Fluorescent method for detection of cleaved collagens using O-phthaldialdehyde (OPA) A rapid and non leaky way for preparation of the sharp intracellular recording microelectrodes Quantification of penicillin G during labor and delivery by capillary electrophoresis
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1