{"title":"一种新型抗核抗体免疫荧光分析自动化系统在日常临床实践中的可靠性。","authors":"Mohammed Alsuwaidi, Margit Dollinger, Martin Fleck, Boris Ehrenstein","doi":"10.1155/2016/6019268","DOIUrl":null,"url":null,"abstract":"<p><p>Automated interpretation (AI) systems for antinuclear antibody (ANA) analysis have been introduced based on assessment of indirect immunofluorescence (IIF) patterns. The diagnostic performance of a novel automated IIF reading system was compared with visual interpretation (VI) of IIF in daily clinical practice to evaluate the reduction of workload. ANA-IIF tests of consecutive serum samples from patients with suspected connective tissue disease were carried out using HEp-2 cells according to routine clinical care. AI was performed using a visual analyser (Zenit G-Sight, Menarini, Germany). Agreement rates between ANA results by AI and VI were calculated. Of the 336 samples investigated, VI yielded 205 (61%) negative, 42 (13%) ambiguous, and 89 (26%) positive results, whereas 82 (24%) were determined to be negative, 176 (52%) ambiguous, and 78 (24%) positive by AI. AI displayed a diagnostic accuracy of 175/336 samples (52%) with a kappa coefficient of 0.34 compared to VI being the gold standard. Solely relying on AI, with VI only performed for all ambiguous samples by AI, would have missed 1 of 89 (1%) positive results by VI and misclassified 2 of 205 (1%) negative results by VI as positive. The use of AI in daily clinical practice resulted only in a moderate reduction of the VI workload (82 of 336 samples: 24%). </p>","PeriodicalId":51715,"journal":{"name":"International Journal of Rheumatology","volume":null,"pages":null},"PeriodicalIF":2.3000,"publicationDate":"2016-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1155/2016/6019268","citationCount":"7","resultStr":"{\"title\":\"The Reliability of a Novel Automated System for ANA Immunofluorescence Analysis in Daily Clinical Practice.\",\"authors\":\"Mohammed Alsuwaidi, Margit Dollinger, Martin Fleck, Boris Ehrenstein\",\"doi\":\"10.1155/2016/6019268\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Automated interpretation (AI) systems for antinuclear antibody (ANA) analysis have been introduced based on assessment of indirect immunofluorescence (IIF) patterns. The diagnostic performance of a novel automated IIF reading system was compared with visual interpretation (VI) of IIF in daily clinical practice to evaluate the reduction of workload. ANA-IIF tests of consecutive serum samples from patients with suspected connective tissue disease were carried out using HEp-2 cells according to routine clinical care. AI was performed using a visual analyser (Zenit G-Sight, Menarini, Germany). Agreement rates between ANA results by AI and VI were calculated. Of the 336 samples investigated, VI yielded 205 (61%) negative, 42 (13%) ambiguous, and 89 (26%) positive results, whereas 82 (24%) were determined to be negative, 176 (52%) ambiguous, and 78 (24%) positive by AI. AI displayed a diagnostic accuracy of 175/336 samples (52%) with a kappa coefficient of 0.34 compared to VI being the gold standard. Solely relying on AI, with VI only performed for all ambiguous samples by AI, would have missed 1 of 89 (1%) positive results by VI and misclassified 2 of 205 (1%) negative results by VI as positive. The use of AI in daily clinical practice resulted only in a moderate reduction of the VI workload (82 of 336 samples: 24%). </p>\",\"PeriodicalId\":51715,\"journal\":{\"name\":\"International Journal of Rheumatology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2016-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://sci-hub-pdf.com/10.1155/2016/6019268\",\"citationCount\":\"7\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Rheumatology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1155/2016/6019268\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2016/5/9 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"RHEUMATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Rheumatology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1155/2016/6019268","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2016/5/9 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"RHEUMATOLOGY","Score":null,"Total":0}
The Reliability of a Novel Automated System for ANA Immunofluorescence Analysis in Daily Clinical Practice.
Automated interpretation (AI) systems for antinuclear antibody (ANA) analysis have been introduced based on assessment of indirect immunofluorescence (IIF) patterns. The diagnostic performance of a novel automated IIF reading system was compared with visual interpretation (VI) of IIF in daily clinical practice to evaluate the reduction of workload. ANA-IIF tests of consecutive serum samples from patients with suspected connective tissue disease were carried out using HEp-2 cells according to routine clinical care. AI was performed using a visual analyser (Zenit G-Sight, Menarini, Germany). Agreement rates between ANA results by AI and VI were calculated. Of the 336 samples investigated, VI yielded 205 (61%) negative, 42 (13%) ambiguous, and 89 (26%) positive results, whereas 82 (24%) were determined to be negative, 176 (52%) ambiguous, and 78 (24%) positive by AI. AI displayed a diagnostic accuracy of 175/336 samples (52%) with a kappa coefficient of 0.34 compared to VI being the gold standard. Solely relying on AI, with VI only performed for all ambiguous samples by AI, would have missed 1 of 89 (1%) positive results by VI and misclassified 2 of 205 (1%) negative results by VI as positive. The use of AI in daily clinical practice resulted only in a moderate reduction of the VI workload (82 of 336 samples: 24%).