{"title":"中性醋酸糊精腹膜透析液的疗效和生物相容性。","authors":"Satoshi Shimada,&nbsp;Takefumi Mori,&nbsp;Kenji Koizumi,&nbsp;Shinichi Sato,&nbsp;Ikuko Oba-Yabana,&nbsp;Yusuke Ohsaki,&nbsp;Emiko Sato,&nbsp;Eri Naganuma,&nbsp;Naho Kurasawa,&nbsp;Mihoko Tsuchikawa,&nbsp;Sadayoshi Ito","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Neutral icodextrin peritoneal dialysis (PD) fluid (n-ICO) has become available for use in Japan. However, removal of water and solutes remains to be elucidated in detail. The present study was designed to determine removal of water, electrolytes, and small, middle, and large molecules in a period of 16 hours. In addition, biocompatibility with respect to peritoneal mesothelial cells was determined.Three stable patients undergoing PD at Tohoku University Hospital were administered n-ICO. Water removal was monitored every 2 hours. Sodium, urea nitrogen [molecular weight (MW): 28 Da], uric acid (MW: 168 Da), β<sub>2</sub>-microglobulin [β2M (MW: 11,800 Da)], α<sub>1</sub>-microglobulin [α1M (MW: 33,000 Da)], albumin (MW: 66,000 Da), and immunoglobulin G (MW: 160,000 Da) were measured in plasma and dialysate.Primary human peritoneal mesothelial cells were collected from 6 patients. Equal numbers of cells were seeded into 96-well culture plates and cultured for 12 hours. Culture medium was then replaced with dialysate, and 24-hour cell proliferation was determined by WST-1 assay.Water removal was sustained for 16 hours with n-ICO. The Na concentration in effluent did not change over that time. Small molecules such as urea nitrogen and uric acid were rapidly removed. Thus, their dialysate-to-plasma concentration ratio (D/P) approached 1.0 (equilibrium) in 2 - 4 hours. The D/P values for the larger molecules β2M and α1M were 0.4 and less than 0.1 respectively at 16 hours. However, larger molecules were removed in a time-dependent manner.Cell proliferation with n-ICO PD fluid was not different from that with lactate-buffered glucose PD fluid, but was increased from that with acidic icodextrin PD fluid (a-ICO).Water and solute removal with the new n-ICO is not much different from that with a-ICO. However, biocompatibility as reflected by cell proliferation was superior under n-ICO than under a-ICO and equal to proliferation under glucose PD fluid.</p>","PeriodicalId":7361,"journal":{"name":"Advances in peritoneal dialysis. Conference on Peritoneal Dialysis","volume":"32 ","pages":"46-50"},"PeriodicalIF":0.0000,"publicationDate":"2016-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Efficacy and Biocompatibility of Neutral Icodextrin Peritoneal Dialysis Fluid.\",\"authors\":\"Satoshi Shimada,&nbsp;Takefumi Mori,&nbsp;Kenji Koizumi,&nbsp;Shinichi Sato,&nbsp;Ikuko Oba-Yabana,&nbsp;Yusuke Ohsaki,&nbsp;Emiko Sato,&nbsp;Eri Naganuma,&nbsp;Naho Kurasawa,&nbsp;Mihoko Tsuchikawa,&nbsp;Sadayoshi Ito\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>Neutral icodextrin peritoneal dialysis (PD) fluid (n-ICO) has become available for use in Japan. However, removal of water and solutes remains to be elucidated in detail. The present study was designed to determine removal of water, electrolytes, and small, middle, and large molecules in a period of 16 hours. In addition, biocompatibility with respect to peritoneal mesothelial cells was determined.Three stable patients undergoing PD at Tohoku University Hospital were administered n-ICO. Water removal was monitored every 2 hours. Sodium, urea nitrogen [molecular weight (MW): 28 Da], uric acid (MW: 168 Da), β<sub>2</sub>-microglobulin [β2M (MW: 11,800 Da)], α<sub>1</sub>-microglobulin [α1M (MW: 33,000 Da)], albumin (MW: 66,000 Da), and immunoglobulin G (MW: 160,000 Da) were measured in plasma and dialysate.Primary human peritoneal mesothelial cells were collected from 6 patients. Equal numbers of cells were seeded into 96-well culture plates and cultured for 12 hours. Culture medium was then replaced with dialysate, and 24-hour cell proliferation was determined by WST-1 assay.Water removal was sustained for 16 hours with n-ICO. The Na concentration in effluent did not change over that time. Small molecules such as urea nitrogen and uric acid were rapidly removed. Thus, their dialysate-to-plasma concentration ratio (D/P) approached 1.0 (equilibrium) in 2 - 4 hours. The D/P values for the larger molecules β2M and α1M were 0.4 and less than 0.1 respectively at 16 hours. However, larger molecules were removed in a time-dependent manner.Cell proliferation with n-ICO PD fluid was not different from that with lactate-buffered glucose PD fluid, but was increased from that with acidic icodextrin PD fluid (a-ICO).Water and solute removal with the new n-ICO is not much different from that with a-ICO. However, biocompatibility as reflected by cell proliferation was superior under n-ICO than under a-ICO and equal to proliferation under glucose PD fluid.</p>\",\"PeriodicalId\":7361,\"journal\":{\"name\":\"Advances in peritoneal dialysis. Conference on Peritoneal Dialysis\",\"volume\":\"32 \",\"pages\":\"46-50\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2016-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Advances in peritoneal dialysis. Conference on Peritoneal Dialysis\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advances in peritoneal dialysis. Conference on Peritoneal Dialysis","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

中性醋酸糊精腹膜透析(PD)液(n-ICO)已在日本可用。然而,水和溶质的去除仍有待详细阐明。本研究旨在测定16小时内水、电解质和小、中、大分子的去除情况。此外,测定了与腹膜间皮细胞的生物相容性。在东北大学医院接受PD治疗的3例病情稳定的患者采用n-ICO治疗。每2小时监测一次除水情况。测定血浆和透析液中的钠、尿素氮[分子量(MW): 28 Da]、尿酸(MW: 168 Da)、β2-微球蛋白[β2M (MW: 11,800 Da)]、α1-微球蛋白[α1M (MW: 33,000 Da)]、白蛋白(MW: 66,000 Da)和免疫球蛋白G (MW: 160,000 Da)。从6例患者身上采集了原代人腹膜间皮细胞。将等量细胞接种到96孔培养板中,培养12小时。用透析液替代培养基,用WST-1法测定24小时细胞增殖情况。n-ICO持续除水16小时。在这段时间内,污水中的钠浓度没有变化。尿素氮和尿酸等小分子被迅速去除。因此,他们的透析液与血浆浓度比(D/P)在2 - 4小时内接近1.0(平衡)。16 h时,大分子β2M和α1M的D/P值分别为0.4和小于0.1。然而,较大的分子以时间依赖的方式被去除。n-ICO PD液的细胞增殖与乳酸缓冲葡萄糖PD液的细胞增殖无显著差异,但与酸性碘糊精PD液(a-ICO)相比,细胞增殖明显增加。新型n-ICO对水和溶质的去除效果与a-ICO相差不大。但生物相容性表现为n-ICO处理下细胞增殖优于a-ICO处理,与葡萄糖PD液处理下细胞增殖相同。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Efficacy and Biocompatibility of Neutral Icodextrin Peritoneal Dialysis Fluid.

Neutral icodextrin peritoneal dialysis (PD) fluid (n-ICO) has become available for use in Japan. However, removal of water and solutes remains to be elucidated in detail. The present study was designed to determine removal of water, electrolytes, and small, middle, and large molecules in a period of 16 hours. In addition, biocompatibility with respect to peritoneal mesothelial cells was determined.Three stable patients undergoing PD at Tohoku University Hospital were administered n-ICO. Water removal was monitored every 2 hours. Sodium, urea nitrogen [molecular weight (MW): 28 Da], uric acid (MW: 168 Da), β2-microglobulin [β2M (MW: 11,800 Da)], α1-microglobulin [α1M (MW: 33,000 Da)], albumin (MW: 66,000 Da), and immunoglobulin G (MW: 160,000 Da) were measured in plasma and dialysate.Primary human peritoneal mesothelial cells were collected from 6 patients. Equal numbers of cells were seeded into 96-well culture plates and cultured for 12 hours. Culture medium was then replaced with dialysate, and 24-hour cell proliferation was determined by WST-1 assay.Water removal was sustained for 16 hours with n-ICO. The Na concentration in effluent did not change over that time. Small molecules such as urea nitrogen and uric acid were rapidly removed. Thus, their dialysate-to-plasma concentration ratio (D/P) approached 1.0 (equilibrium) in 2 - 4 hours. The D/P values for the larger molecules β2M and α1M were 0.4 and less than 0.1 respectively at 16 hours. However, larger molecules were removed in a time-dependent manner.Cell proliferation with n-ICO PD fluid was not different from that with lactate-buffered glucose PD fluid, but was increased from that with acidic icodextrin PD fluid (a-ICO).Water and solute removal with the new n-ICO is not much different from that with a-ICO. However, biocompatibility as reflected by cell proliferation was superior under n-ICO than under a-ICO and equal to proliferation under glucose PD fluid.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Changes in Peritoneal Transport and Peritoneal Damage in Japanese Patients Undergoing Peritoneal Dialysis Using Neutral-pH Dialysate: A Retrospective Cohort Study at Two Centers. Maintaining Peritoneal Dialysis Adequacy: The Process of Incremental Prescription. Comparison of Various Scaling Parameters and Energy Expenditure in Peritoneal Dialysis Patients. Prevention of Peritoneal Dialysis Drop-Out. Accuracy of a Newly-Introduced Oscillometric Device for the Estimation of Arterial Stiffness Indices in Patients on Peritoneal Dialysis: A Preliminary Validation Study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1