{"title":"射线照相和电子工作长度比较。三种电子顶点定位器的体内研究。","authors":"V Di Taranto, A Libonati, E Montemurro, G Gallusi","doi":"10.23812/21-3supp1-10","DOIUrl":null,"url":null,"abstract":"<p><p>The aim of this study is to compare the working length measured with Root ZX (EAL1), Propex II (EAL2) and Endo Analyzer Model 8005 (EAL3) with radiographic measurements during endodontic treatment. Fifty single canal teeth scheduled for endodontic treatment were selected for the study. After endodontic access preparation, pulp chamber was irrigated with 5mL of 5% sodium hypochlorite (NaOCl). The canal was scouted with a size 10 K-file introduced slightly over the apex to verify the canal patency. The working length (WL) was measured using all three EAL devices and radiographically. All measured working lengths were recorded and compared using ANOVA for repeated measures considering the multiple comparison of paired data Least Significant Difference (LSD). The WL mean values for EAL1 were 19.1 ± 1.7 mm; for EAL2 were 19.4 ± 1.7; for EAL3 were 19.4 ± 1.6 and for RG were 19.2 ± 1.7. The Pearson correlation coefficients indicated that the following correlation between electronic devices and radiographic WL: EAL1- 0.986, EAL2- 0.953 and EAL3- 0.931. Considering the mean values compared to radiographic measures, Root ZX gave better results than the other tested EAL devices. Endo Analyzer Model 8005 showed the worst performance.</p>","PeriodicalId":15084,"journal":{"name":"Journal of biological regulators and homeostatic agents","volume":"35 3 Suppl. 1","pages":"77-85"},"PeriodicalIF":0.8000,"publicationDate":"2021-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Radiographic and electronic working length comparison. An <i>in vivo</i> study of three Electronic Apex Locator devices.\",\"authors\":\"V Di Taranto, A Libonati, E Montemurro, G Gallusi\",\"doi\":\"10.23812/21-3supp1-10\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The aim of this study is to compare the working length measured with Root ZX (EAL1), Propex II (EAL2) and Endo Analyzer Model 8005 (EAL3) with radiographic measurements during endodontic treatment. Fifty single canal teeth scheduled for endodontic treatment were selected for the study. After endodontic access preparation, pulp chamber was irrigated with 5mL of 5% sodium hypochlorite (NaOCl). The canal was scouted with a size 10 K-file introduced slightly over the apex to verify the canal patency. The working length (WL) was measured using all three EAL devices and radiographically. All measured working lengths were recorded and compared using ANOVA for repeated measures considering the multiple comparison of paired data Least Significant Difference (LSD). The WL mean values for EAL1 were 19.1 ± 1.7 mm; for EAL2 were 19.4 ± 1.7; for EAL3 were 19.4 ± 1.6 and for RG were 19.2 ± 1.7. The Pearson correlation coefficients indicated that the following correlation between electronic devices and radiographic WL: EAL1- 0.986, EAL2- 0.953 and EAL3- 0.931. Considering the mean values compared to radiographic measures, Root ZX gave better results than the other tested EAL devices. Endo Analyzer Model 8005 showed the worst performance.</p>\",\"PeriodicalId\":15084,\"journal\":{\"name\":\"Journal of biological regulators and homeostatic agents\",\"volume\":\"35 3 Suppl. 1\",\"pages\":\"77-85\"},\"PeriodicalIF\":0.8000,\"publicationDate\":\"2021-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of biological regulators and homeostatic agents\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.23812/21-3supp1-10\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"ENDOCRINOLOGY & METABOLISM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of biological regulators and homeostatic agents","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.23812/21-3supp1-10","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0
摘要
本研究的目的是比较根管治疗期间用Root ZX (EAL1)、Propex II (EAL2)和Endo Analyzer Model 8005 (EAL3)测量的工作长度与x线测量的工作长度。选择50颗单管牙进行牙髓治疗。根管通路准备完成后,用5mL 5%次氯酸钠(NaOCl)冲洗牙髓腔。用10k大小的锉刀在根管顶端稍上方检查根管是否通畅。工作长度(WL)测量使用所有三个EAL装置和放射照相。所有测量的工作长度都被记录下来,并考虑成对数据的多重比较,使用方差分析对重复测量进行比较,最小显著差异(LSD)。EAL1的WL平均值为19.1±1.7 mm;EAL2为19.4±1.7;EAL3为19.4±1.6,RG为19.2±1.7。Pearson相关系数显示,电子设备与x线影像WL的相关系数分别为EAL1- 0.986、EAL2- 0.953、EAL3- 0.931。考虑到与放射测量相比较的平均值,Root ZX给出了比其他测试EAL设备更好的结果。远藤分析仪8005型表现最差。
Radiographic and electronic working length comparison. An in vivo study of three Electronic Apex Locator devices.
The aim of this study is to compare the working length measured with Root ZX (EAL1), Propex II (EAL2) and Endo Analyzer Model 8005 (EAL3) with radiographic measurements during endodontic treatment. Fifty single canal teeth scheduled for endodontic treatment were selected for the study. After endodontic access preparation, pulp chamber was irrigated with 5mL of 5% sodium hypochlorite (NaOCl). The canal was scouted with a size 10 K-file introduced slightly over the apex to verify the canal patency. The working length (WL) was measured using all three EAL devices and radiographically. All measured working lengths were recorded and compared using ANOVA for repeated measures considering the multiple comparison of paired data Least Significant Difference (LSD). The WL mean values for EAL1 were 19.1 ± 1.7 mm; for EAL2 were 19.4 ± 1.7; for EAL3 were 19.4 ± 1.6 and for RG were 19.2 ± 1.7. The Pearson correlation coefficients indicated that the following correlation between electronic devices and radiographic WL: EAL1- 0.986, EAL2- 0.953 and EAL3- 0.931. Considering the mean values compared to radiographic measures, Root ZX gave better results than the other tested EAL devices. Endo Analyzer Model 8005 showed the worst performance.
期刊介绍:
Journal of Biological Regulators & Homeostatic Agents (IF 1.397) is a peer-reviewed journal published every 2 months. The journal publishes original papers describing research in the fields of experimental and clinical medicine, molecular biology, biochemistry, regulatory molecules, cellular immunology and pharmacology.