{"title":"数字化手术指南和槽设计钻头在种植体部位准备过程中的热变化:使用牛肋骨模型的离体研究。","authors":"Yoon-Sil Choi,&nbsp;Jae-Woon Oh,&nbsp;Young Lee,&nbsp;Dong-Woon Lee","doi":"10.5051/jpis.2106040302","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>In this study, we aimed to evaluate the degree of heat generation when a novel drill design with an irrigation slot was used with metal sleeve-free (MF) and metal sleeve-incorporated (MI) surgical guides in an environment similar to that of the actual oral cavity.</p><p><strong>Methods: </strong>A typodont with a missing mandibular right first molar and 21 bovine rib blocks were used. Three-dimensional-printed MF and MI surgical guides, designed for the placement of internal tapered implant fixtures, were used with slot and non-slot drills. The following groups were compared: group 1, MI surgical guide with slot drill; group 2, MI surgical guide with a non-slot drill; and group 3, MF surgical guide with a slot drill. A constant-temperature water bath at 36°C was used. The drilling was performed in 6 stages, and the initial, highest, and lowest temperatures of the cortical bone were measured at each stage using a non-contact infrared thermometer.</p><p><strong>Results: </strong>There were no temperature increases above the initial temperature in any drilling procedure. The only significant difference between the non-slot and slot groups was observed with the use of the first drill in the MI group, with a higher temperature in the non-slot group (<i>P</i>=0.012). When the heat generation during the first and the second drilling was compared in the non-slot group, the heat generation during the first drilling was significantly higher (<i>P<</i>0.001), and there was no significant difference in heat generation between the drills in the slot group.</p><p><strong>Conclusions: </strong>Within the limitations of this study, implant-site preparation with the surgical guide showed no critical increase in the temperature of the cortical bone, regardless of whether there was a slot in the drill. In particular, the slotted drill had a cooling effect during the initial drilling.</p>","PeriodicalId":48795,"journal":{"name":"Journal of Periodontal and Implant Science","volume":"52 5","pages":"411-421"},"PeriodicalIF":2.2000,"publicationDate":"2022-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/18/7c/jpis-52-411.PMC9614178.pdf","citationCount":"1","resultStr":"{\"title\":\"Thermal changes during implant site preparation with a digital surgical guide and slot design drill: an <i>ex vivo</i> study using a bovine rib model.\",\"authors\":\"Yoon-Sil Choi,&nbsp;Jae-Woon Oh,&nbsp;Young Lee,&nbsp;Dong-Woon Lee\",\"doi\":\"10.5051/jpis.2106040302\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>In this study, we aimed to evaluate the degree of heat generation when a novel drill design with an irrigation slot was used with metal sleeve-free (MF) and metal sleeve-incorporated (MI) surgical guides in an environment similar to that of the actual oral cavity.</p><p><strong>Methods: </strong>A typodont with a missing mandibular right first molar and 21 bovine rib blocks were used. Three-dimensional-printed MF and MI surgical guides, designed for the placement of internal tapered implant fixtures, were used with slot and non-slot drills. The following groups were compared: group 1, MI surgical guide with slot drill; group 2, MI surgical guide with a non-slot drill; and group 3, MF surgical guide with a slot drill. A constant-temperature water bath at 36°C was used. The drilling was performed in 6 stages, and the initial, highest, and lowest temperatures of the cortical bone were measured at each stage using a non-contact infrared thermometer.</p><p><strong>Results: </strong>There were no temperature increases above the initial temperature in any drilling procedure. The only significant difference between the non-slot and slot groups was observed with the use of the first drill in the MI group, with a higher temperature in the non-slot group (<i>P</i>=0.012). When the heat generation during the first and the second drilling was compared in the non-slot group, the heat generation during the first drilling was significantly higher (<i>P<</i>0.001), and there was no significant difference in heat generation between the drills in the slot group.</p><p><strong>Conclusions: </strong>Within the limitations of this study, implant-site preparation with the surgical guide showed no critical increase in the temperature of the cortical bone, regardless of whether there was a slot in the drill. In particular, the slotted drill had a cooling effect during the initial drilling.</p>\",\"PeriodicalId\":48795,\"journal\":{\"name\":\"Journal of Periodontal and Implant Science\",\"volume\":\"52 5\",\"pages\":\"411-421\"},\"PeriodicalIF\":2.2000,\"publicationDate\":\"2022-10-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/18/7c/jpis-52-411.PMC9614178.pdf\",\"citationCount\":\"1\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Periodontal and Implant Science\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.5051/jpis.2106040302\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"DENTISTRY, ORAL SURGERY & MEDICINE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Periodontal and Implant Science","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.5051/jpis.2106040302","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 1

摘要

目的:在本研究中,我们旨在评估一种带有冲洗槽的新型钻头与无金属套管(MF)和含金属套管(MI)手术导尿管在与实际口腔相似的环境中一起使用时的产热程度。方法:采用下颌右第一磨牙缺失1例,牛肋块21例。三维打印MF和MI手术指南,设计用于放置内部锥形种植固定装置,与槽和非槽钻头一起使用。比较以下两组:组1:带槽钻的心肌梗死手术指南;第2组,MI手术指南,无槽钻头;第三组为带槽钻的MF手术导尿管。采用36℃恒温水浴。钻孔分6个阶段进行,使用非接触式红外温度计测量每个阶段皮质骨的初始、最高和最低温度。结果:在任何钻孔过程中,温度均未高于初始温度。非槽组和槽组之间的唯一显著差异是在MI组使用第一个钻头时观察到,非槽组的温度更高(P=0.012)。非槽组第一次钻孔和第二次钻孔发热量比较,第一次钻孔发热量显著高于第一次钻孔(P0.001),槽组各钻头发热量无显著差异。结论:在本研究的限制范围内,在手术引导下进行种植体准备时,无论钻孔中是否有凹槽,皮质骨的温度都没有明显升高。特别是,槽钻在初始钻井过程中具有冷却效果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

摘要图片

摘要图片

摘要图片

查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Thermal changes during implant site preparation with a digital surgical guide and slot design drill: an ex vivo study using a bovine rib model.

Purpose: In this study, we aimed to evaluate the degree of heat generation when a novel drill design with an irrigation slot was used with metal sleeve-free (MF) and metal sleeve-incorporated (MI) surgical guides in an environment similar to that of the actual oral cavity.

Methods: A typodont with a missing mandibular right first molar and 21 bovine rib blocks were used. Three-dimensional-printed MF and MI surgical guides, designed for the placement of internal tapered implant fixtures, were used with slot and non-slot drills. The following groups were compared: group 1, MI surgical guide with slot drill; group 2, MI surgical guide with a non-slot drill; and group 3, MF surgical guide with a slot drill. A constant-temperature water bath at 36°C was used. The drilling was performed in 6 stages, and the initial, highest, and lowest temperatures of the cortical bone were measured at each stage using a non-contact infrared thermometer.

Results: There were no temperature increases above the initial temperature in any drilling procedure. The only significant difference between the non-slot and slot groups was observed with the use of the first drill in the MI group, with a higher temperature in the non-slot group (P=0.012). When the heat generation during the first and the second drilling was compared in the non-slot group, the heat generation during the first drilling was significantly higher (P<0.001), and there was no significant difference in heat generation between the drills in the slot group.

Conclusions: Within the limitations of this study, implant-site preparation with the surgical guide showed no critical increase in the temperature of the cortical bone, regardless of whether there was a slot in the drill. In particular, the slotted drill had a cooling effect during the initial drilling.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Periodontal and Implant Science
Journal of Periodontal and Implant Science DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
3.30
自引率
5.30%
发文量
38
期刊介绍: Journal of Periodontal & Implant Science (JPIS) is a peer-reviewed and open-access journal providing up-to-date information relevant to professionalism of periodontology and dental implantology. JPIS is dedicated to global and extensive publication which includes evidence-based original articles, and fundamental reviews in order to cover a variety of interests in the field of periodontal as well as implant science.
期刊最新文献
Does defect configuration affect the outcomes of alveolar ridge preservation? An experimental in vivo study. Comprehensive treatment protocol for peri-implantitis: an up-to date narrative review of the literature. Guided bone regeneration of calcium phosphate-coated and strontium ranelate-doped titanium mesh in a rat calvarial defect model. Soft-tissue volume augmentation using a connective tissue graft and a volume-stable collagen matrix with polydeoxyribonucleotide for immediate implant placement: a pilot study in a dog model. The activin/BMP-2 chimera AB204 promotes periodontal tissue regeneration in a buccal dehiscence model: a pilot study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1