{"title":"PRIMO蒙特卡罗软件作为外部放射治疗计划系统的调试工具。","authors":"Juan-Francisco Calvo-Ortega, Marcelino Hermida-López","doi":"10.5603/RPOR.a2023.0060","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The purpose was to validate the PRIMO Monte Carlo software to be used during the commissioning of a treatment planning system (TPS).</p><p><strong>Materials and methods: </strong>The Acuros XB v. 16.1 algorithm of the Eclipse was configured for 6 MV and 6 MV flattening-filter-free (FFF) photon beams, from a TrueBeam linac equipped with a high-definition 120-leaf multileaf collimator (MLC). PRIMO v. 0.3.64.1814 software was used with the phase space files provided by Varian and benchmarked against the reference dosimetry dataset published by the Imaging and Radiation Oncology Core-Houston (IROC-H). Thirty Eclipse clinical intensity-modulated radiation therapy (IMRT)/volumetric modulated arc therapy (VMAT) plans were verified in three ways: 1) using the PTW Octavius 4D (O4D) system; 2) the Varian Portal Dosimetry system and 3) the PRIMO software. Clinical validation of PRIMO was completed by comparing the simulated dose distributions on the O4D phantom against dose measurements for these 30 clinical plans. Agreement evaluations were performed using a 3% global/2 mm gamma index analysis.</p><p><strong>Results: </strong>PRIMO simulations agreed with the benchmark IROC-H data within 2.0% for both energies. Gamma passing rates (GPRs) from the 30 clinical plan verifications were (6 MV/6MV FFF): 99.4% ± 0.5%/99.9% ± 0.1%, 99.8% ± 0.4%/98.9% ± 1.4%, 99.7% ± 0.4%/99.7% ± 0.4%, for the 1), 2) and 3) verification methods, respectively. Agreement between PRIMO simulations on the O4D phantom and 3D dose measurements resulted in GPRs of 97.9% ± 2.4%/99.7% ± 0.4%.</p><p><strong>Conclusion: </strong>The PRIMO software is a valuable tool for dosimetric verification of clinical plans during the commissioning of the primary TPS.</p>","PeriodicalId":47283,"journal":{"name":"Reports of Practical Oncology and Radiotherapy","volume":"28 4","pages":"529-540"},"PeriodicalIF":1.2000,"publicationDate":"2023-08-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/71/42/rpor-28-4-529.PMC10547427.pdf","citationCount":"0","resultStr":"{\"title\":\"PRIMO Monte Carlo software as a tool for commissioning of an external beam radiotherapy treatment planning system.\",\"authors\":\"Juan-Francisco Calvo-Ortega, Marcelino Hermida-López\",\"doi\":\"10.5603/RPOR.a2023.0060\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The purpose was to validate the PRIMO Monte Carlo software to be used during the commissioning of a treatment planning system (TPS).</p><p><strong>Materials and methods: </strong>The Acuros XB v. 16.1 algorithm of the Eclipse was configured for 6 MV and 6 MV flattening-filter-free (FFF) photon beams, from a TrueBeam linac equipped with a high-definition 120-leaf multileaf collimator (MLC). PRIMO v. 0.3.64.1814 software was used with the phase space files provided by Varian and benchmarked against the reference dosimetry dataset published by the Imaging and Radiation Oncology Core-Houston (IROC-H). Thirty Eclipse clinical intensity-modulated radiation therapy (IMRT)/volumetric modulated arc therapy (VMAT) plans were verified in three ways: 1) using the PTW Octavius 4D (O4D) system; 2) the Varian Portal Dosimetry system and 3) the PRIMO software. Clinical validation of PRIMO was completed by comparing the simulated dose distributions on the O4D phantom against dose measurements for these 30 clinical plans. Agreement evaluations were performed using a 3% global/2 mm gamma index analysis.</p><p><strong>Results: </strong>PRIMO simulations agreed with the benchmark IROC-H data within 2.0% for both energies. Gamma passing rates (GPRs) from the 30 clinical plan verifications were (6 MV/6MV FFF): 99.4% ± 0.5%/99.9% ± 0.1%, 99.8% ± 0.4%/98.9% ± 1.4%, 99.7% ± 0.4%/99.7% ± 0.4%, for the 1), 2) and 3) verification methods, respectively. Agreement between PRIMO simulations on the O4D phantom and 3D dose measurements resulted in GPRs of 97.9% ± 2.4%/99.7% ± 0.4%.</p><p><strong>Conclusion: </strong>The PRIMO software is a valuable tool for dosimetric verification of clinical plans during the commissioning of the primary TPS.</p>\",\"PeriodicalId\":47283,\"journal\":{\"name\":\"Reports of Practical Oncology and Radiotherapy\",\"volume\":\"28 4\",\"pages\":\"529-540\"},\"PeriodicalIF\":1.2000,\"publicationDate\":\"2023-08-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://ftp.ncbi.nlm.nih.gov/pub/pmc/oa_pdf/71/42/rpor-28-4-529.PMC10547427.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Reports of Practical Oncology and Radiotherapy\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.5603/RPOR.a2023.0060\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2023/1/1 0:00:00\",\"PubModel\":\"eCollection\",\"JCR\":\"Q4\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Reports of Practical Oncology and Radiotherapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5603/RPOR.a2023.0060","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2023/1/1 0:00:00","PubModel":"eCollection","JCR":"Q4","JCRName":"ONCOLOGY","Score":null,"Total":0}
PRIMO Monte Carlo software as a tool for commissioning of an external beam radiotherapy treatment planning system.
Background: The purpose was to validate the PRIMO Monte Carlo software to be used during the commissioning of a treatment planning system (TPS).
Materials and methods: The Acuros XB v. 16.1 algorithm of the Eclipse was configured for 6 MV and 6 MV flattening-filter-free (FFF) photon beams, from a TrueBeam linac equipped with a high-definition 120-leaf multileaf collimator (MLC). PRIMO v. 0.3.64.1814 software was used with the phase space files provided by Varian and benchmarked against the reference dosimetry dataset published by the Imaging and Radiation Oncology Core-Houston (IROC-H). Thirty Eclipse clinical intensity-modulated radiation therapy (IMRT)/volumetric modulated arc therapy (VMAT) plans were verified in three ways: 1) using the PTW Octavius 4D (O4D) system; 2) the Varian Portal Dosimetry system and 3) the PRIMO software. Clinical validation of PRIMO was completed by comparing the simulated dose distributions on the O4D phantom against dose measurements for these 30 clinical plans. Agreement evaluations were performed using a 3% global/2 mm gamma index analysis.
Results: PRIMO simulations agreed with the benchmark IROC-H data within 2.0% for both energies. Gamma passing rates (GPRs) from the 30 clinical plan verifications were (6 MV/6MV FFF): 99.4% ± 0.5%/99.9% ± 0.1%, 99.8% ± 0.4%/98.9% ± 1.4%, 99.7% ± 0.4%/99.7% ± 0.4%, for the 1), 2) and 3) verification methods, respectively. Agreement between PRIMO simulations on the O4D phantom and 3D dose measurements resulted in GPRs of 97.9% ± 2.4%/99.7% ± 0.4%.
Conclusion: The PRIMO software is a valuable tool for dosimetric verification of clinical plans during the commissioning of the primary TPS.
期刊介绍:
Reports of Practical Oncology and Radiotherapy is an interdisciplinary bimonthly journal, publishing original contributions in clinical oncology and radiotherapy, as well as in radiotherapy physics, techniques and radiotherapy equipment. Reports of Practical Oncology and Radiotherapy is a journal of the Polish Society of Radiation Oncology, the Czech Society of Radiation Oncology, the Hungarian Society for Radiation Oncology, the Slovenian Society for Radiotherapy and Oncology, the Polish Study Group of Head and Neck Cancer, the Guild of Bulgarian Radiotherapists and the Greater Poland Cancer Centre, affiliated with the Spanish Society of Radiotherapy and Oncology, the Italian Association of Radiotherapy and the Portuguese Society of Radiotherapy - Oncology.