CT重建滤波选择对放射治疗计划中霍斯菲尔德单元的影响

IF 0.3 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING Journal of Radiotherapy in Practice Pub Date : 2023-06-19 DOI:10.1017/S1460396923000249
O. Nhila, M. Talbi, M. El Mansouri, M. A. Youssoufi, M. Erraoudi, E. Chakir, Mohamed Azougagh
{"title":"CT重建滤波选择对放射治疗计划中霍斯菲尔德单元的影响","authors":"O. Nhila, M. Talbi, M. El Mansouri, M. A. Youssoufi, M. Erraoudi, E. Chakir, Mohamed Azougagh","doi":"10.1017/S1460396923000249","DOIUrl":null,"url":null,"abstract":"Abstract Introduction: This work aims to evaluate the effect of Hitachi 16-slice scanner reconstruction filters on Hounsfield unit (HU) variations. In the literature, there is a lack of information from a wide variety of scanners in this regard. In addition, not all studies have investigated the effect of reconstruction filters on HU in an exhaustive way. Methods: The computerised imaging reference system electron density phantom (model 062M) was scanned with different substitute materials of different density from Hitachi 16-slice computed tomography. The raw images were obtained with four tube voltage settings: 80 kVp, 100 kVp, 120 kVp and 140 kVp. The raw images for each energy level were then reconstructed using different reconstruction filters. Results: The HU values of dense bone were significantly different when changing the reconstruction filters without beam hardening correction (BHC). Nevertheless, when selecting the BHC, this variation decreases heavily for 80 kVp and decreases slightly for 140 kVp, but it remains outside the tolerance of ±50 HU. However, for 100 kVp and 120 kVp, the differences in HU values become within the tolerances indicated for dense bone. Conclusions: Changing image reconstruction filters during a dosimetric scan had a significant effect on HU in dense bone. Therefore, it is recommended to evaluate this effect during the commissioning phase. As a result, this study provides a methodology to comprehensively investigate the effect of reconstruction filters on HU.","PeriodicalId":44597,"journal":{"name":"Journal of Radiotherapy in Practice","volume":null,"pages":null},"PeriodicalIF":0.3000,"publicationDate":"2023-06-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The effect of CT reconstruction filter selection on Hounsfield units in radiotherapy treatment planning\",\"authors\":\"O. Nhila, M. Talbi, M. El Mansouri, M. A. Youssoufi, M. Erraoudi, E. Chakir, Mohamed Azougagh\",\"doi\":\"10.1017/S1460396923000249\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Introduction: This work aims to evaluate the effect of Hitachi 16-slice scanner reconstruction filters on Hounsfield unit (HU) variations. In the literature, there is a lack of information from a wide variety of scanners in this regard. In addition, not all studies have investigated the effect of reconstruction filters on HU in an exhaustive way. Methods: The computerised imaging reference system electron density phantom (model 062M) was scanned with different substitute materials of different density from Hitachi 16-slice computed tomography. The raw images were obtained with four tube voltage settings: 80 kVp, 100 kVp, 120 kVp and 140 kVp. The raw images for each energy level were then reconstructed using different reconstruction filters. Results: The HU values of dense bone were significantly different when changing the reconstruction filters without beam hardening correction (BHC). Nevertheless, when selecting the BHC, this variation decreases heavily for 80 kVp and decreases slightly for 140 kVp, but it remains outside the tolerance of ±50 HU. However, for 100 kVp and 120 kVp, the differences in HU values become within the tolerances indicated for dense bone. Conclusions: Changing image reconstruction filters during a dosimetric scan had a significant effect on HU in dense bone. Therefore, it is recommended to evaluate this effect during the commissioning phase. As a result, this study provides a methodology to comprehensively investigate the effect of reconstruction filters on HU.\",\"PeriodicalId\":44597,\"journal\":{\"name\":\"Journal of Radiotherapy in Practice\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.3000,\"publicationDate\":\"2023-06-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Radiotherapy in Practice\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1017/S1460396923000249\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiotherapy in Practice","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1017/S1460396923000249","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

摘要简介:本工作旨在评估日立16层扫描仪重建滤波器对Hounsfield单位(HU)变化的影响。在文献中,缺乏来自各种扫描仪的这方面的信息。此外,并不是所有的研究都以详尽的方式研究了重建滤波器对HU的影响。方法:用日立16层计算机断层扫描的不同密度的替代材料对062M型计算机成像参考系统电子密度体模进行扫描。原始图像是用四种管电压设置获得的:80kVp、100kVp、120kVp和140kVp。然后使用不同的重建滤波器来重建每个能级的原始图像。结果:在不进行束硬化校正(BHC)的情况下,改变重建滤波器时,致密骨的HU值有显著差异。然而,当选择BHC时,80 kVp时,这种变化会大幅下降,140 kVp时略有下降,但仍在±50 HU的公差范围之外。然而,对于100 kVp和120 kVp,HU值的差异会在致密骨的公差范围内。结论:在剂量扫描过程中改变图像重建滤波器对致密骨的HU有显著影响。因此,建议在调试阶段评估这种影响。因此,本研究为全面研究重建滤波器对HU的影响提供了一种方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
The effect of CT reconstruction filter selection on Hounsfield units in radiotherapy treatment planning
Abstract Introduction: This work aims to evaluate the effect of Hitachi 16-slice scanner reconstruction filters on Hounsfield unit (HU) variations. In the literature, there is a lack of information from a wide variety of scanners in this regard. In addition, not all studies have investigated the effect of reconstruction filters on HU in an exhaustive way. Methods: The computerised imaging reference system electron density phantom (model 062M) was scanned with different substitute materials of different density from Hitachi 16-slice computed tomography. The raw images were obtained with four tube voltage settings: 80 kVp, 100 kVp, 120 kVp and 140 kVp. The raw images for each energy level were then reconstructed using different reconstruction filters. Results: The HU values of dense bone were significantly different when changing the reconstruction filters without beam hardening correction (BHC). Nevertheless, when selecting the BHC, this variation decreases heavily for 80 kVp and decreases slightly for 140 kVp, but it remains outside the tolerance of ±50 HU. However, for 100 kVp and 120 kVp, the differences in HU values become within the tolerances indicated for dense bone. Conclusions: Changing image reconstruction filters during a dosimetric scan had a significant effect on HU in dense bone. Therefore, it is recommended to evaluate this effect during the commissioning phase. As a result, this study provides a methodology to comprehensively investigate the effect of reconstruction filters on HU.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Radiotherapy in Practice
Journal of Radiotherapy in Practice RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
0.80
自引率
0.00%
发文量
36
期刊介绍: Journal of Radiotherapy in Practice is a peer-reviewed journal covering all of the current modalities specific to clinical oncology and radiotherapy. The journal aims to publish research from a wide range of styles and encourage debate and the exchange of information and opinion from within the field of radiotherapy practice and clinical oncology. The journal also aims to encourage technical evaluations and case studies as well as equipment reviews that will be of interest to an international radiotherapy audience.
期刊最新文献
Wobbling nature of gamma passing rate as a function of calibration field sizes in patient-specific quality assurance Secondary fragmentation and relative biological effectiveness (RBE) study using Bridge SOI microdosimeter: Monte Carlo simulation Keratin-based topical cream for radiation dermatitis during head and neck radiotherapy: a randomised, open-label pilot study. Single catheter 3d volume based hybrid inverse planning optimization in IVBT can improve organ sparing – CORRIGENDUM Dosimetric case study of 3-D FiF vs. VMAT techniques in the treatment of H/N tumour
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1