capillillys 3 Octa®:HbA1c定量分析性能评估

Ganouni Mohamed Rayen, Chaira Maha, Bouslama Jamila, Neffati Fadoua, Nejjar Mohamed Fadhel
{"title":"capillillys 3 Octa®:HbA1c定量分析性能评估","authors":"Ganouni Mohamed Rayen, Chaira Maha, Bouslama Jamila, Neffati Fadoua, Nejjar Mohamed Fadhel","doi":"10.18502/ijdo.v14i2.9453","DOIUrl":null,"url":null,"abstract":"Objective: Determination of HbA1c level is a precious indicator for therapeutic follow-up of patients with type 1 or type 2 diabetes. Our study aimed to evaluate the analytical characteristics of Capillarys 3 Octa® HbA1c measurement by capillary electrophoresis. \nMaterials and Methods: Our study involved 265 venous whole blood specimens repeatability, intermediate fidelity, accuracy, linearity, and correlation with the Arkray HA-8180 analyzer which uses HPLC as a dosing method. We studied interferences such as hematocrit, triglycerides, total bilirubin, labile fraction, and hemoglobin abnormalities. \nResults: The linearity correlation was between 4.4% and 20.3%. There was a strong correlation with HPLC (r > 0.99, P< 0.0001). No interference from hematocrit (20-93%) (P: 0.888), triglycerides (until 25 mmol/L) (P: 0.388), total bilirubin (< 587 µmol/L) (P: 0.993), and labile fraction was observed. No problem related to inter-sample contamination was observed. The sensitivity was zero for homozygous sickle cell disease and S/C composite hemoglobinosis. However, sensitivity was high for heterozygous forms (69% for A/S and 60% for A/C). The analyzer was able to separate and quantify HbA2 fraction, allowing ß-thalassemia accidental detection. \nConclusion: Capillarys 3 Octa® based on capillary electrophoresis proved to be precise and a linear instrument for HbA1c measurement. Several clinical interferences and Hb variants had no effect on the results. The results of this evaluation suggest that this analyzer is suitable for routine use in clinical chemistry laboratories.","PeriodicalId":33205,"journal":{"name":"Iranian Journal of Diabetes and Obesity","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-05-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Capillarys 3 Octa®: Analytical Performance Assessment for HbA1c Quantification\",\"authors\":\"Ganouni Mohamed Rayen, Chaira Maha, Bouslama Jamila, Neffati Fadoua, Nejjar Mohamed Fadhel\",\"doi\":\"10.18502/ijdo.v14i2.9453\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: Determination of HbA1c level is a precious indicator for therapeutic follow-up of patients with type 1 or type 2 diabetes. Our study aimed to evaluate the analytical characteristics of Capillarys 3 Octa® HbA1c measurement by capillary electrophoresis. \\nMaterials and Methods: Our study involved 265 venous whole blood specimens repeatability, intermediate fidelity, accuracy, linearity, and correlation with the Arkray HA-8180 analyzer which uses HPLC as a dosing method. We studied interferences such as hematocrit, triglycerides, total bilirubin, labile fraction, and hemoglobin abnormalities. \\nResults: The linearity correlation was between 4.4% and 20.3%. There was a strong correlation with HPLC (r > 0.99, P< 0.0001). No interference from hematocrit (20-93%) (P: 0.888), triglycerides (until 25 mmol/L) (P: 0.388), total bilirubin (< 587 µmol/L) (P: 0.993), and labile fraction was observed. No problem related to inter-sample contamination was observed. The sensitivity was zero for homozygous sickle cell disease and S/C composite hemoglobinosis. However, sensitivity was high for heterozygous forms (69% for A/S and 60% for A/C). The analyzer was able to separate and quantify HbA2 fraction, allowing ß-thalassemia accidental detection. \\nConclusion: Capillarys 3 Octa® based on capillary electrophoresis proved to be precise and a linear instrument for HbA1c measurement. Several clinical interferences and Hb variants had no effect on the results. The results of this evaluation suggest that this analyzer is suitable for routine use in clinical chemistry laboratories.\",\"PeriodicalId\":33205,\"journal\":{\"name\":\"Iranian Journal of Diabetes and Obesity\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-05-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Iranian Journal of Diabetes and Obesity\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.18502/ijdo.v14i2.9453\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Diabetes and Obesity","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18502/ijdo.v14i2.9453","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:HbA1c水平的测定是1型或2型糖尿病患者治疗随访的重要指标。我们的研究旨在评估毛细管电泳法测定Capillarys 3 Octa®HbA1c的分析特性。材料和方法:我们的研究涉及265份静脉全血样本的可重复性、中间保真度、准确性、线性以及与Arkray HA-8180分析仪的相关性,该分析仪使用HPLC作为给药方法。我们研究了红细胞压积、甘油三酯、总胆红素、不稳定分数和血红蛋白异常等干扰因素。结果:线性相关度在4.4%-20.3%之间,与高效液相色谱法有很强的相关性(r>0.99,P<0.0001),血细胞压积(20-93%)(P<0.888)、甘油三酯(至25mmol/L)(P>0.388)、总胆红素(<587µmol/L)(P<0.993)和不稳定分数均无干扰。未观察到与样品间污染有关的问题。纯合子镰状细胞病和S/C复合性血红蛋白病的敏感性为零。然而,杂合型的敏感性很高(A/S为69%,A/C为60%)。该分析仪能够分离和量化HbA2分数,从而实现地中海贫血的意外检测。结论:基于毛细管电泳的Capillarys 3 Octa®是一种精确、线性的HbA1c测定仪器。一些临床干扰和Hb变异对结果没有影响。评估结果表明,该分析仪适用于临床化学实验室的常规使用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Capillarys 3 Octa®: Analytical Performance Assessment for HbA1c Quantification
Objective: Determination of HbA1c level is a precious indicator for therapeutic follow-up of patients with type 1 or type 2 diabetes. Our study aimed to evaluate the analytical characteristics of Capillarys 3 Octa® HbA1c measurement by capillary electrophoresis. Materials and Methods: Our study involved 265 venous whole blood specimens repeatability, intermediate fidelity, accuracy, linearity, and correlation with the Arkray HA-8180 analyzer which uses HPLC as a dosing method. We studied interferences such as hematocrit, triglycerides, total bilirubin, labile fraction, and hemoglobin abnormalities. Results: The linearity correlation was between 4.4% and 20.3%. There was a strong correlation with HPLC (r > 0.99, P< 0.0001). No interference from hematocrit (20-93%) (P: 0.888), triglycerides (until 25 mmol/L) (P: 0.388), total bilirubin (< 587 µmol/L) (P: 0.993), and labile fraction was observed. No problem related to inter-sample contamination was observed. The sensitivity was zero for homozygous sickle cell disease and S/C composite hemoglobinosis. However, sensitivity was high for heterozygous forms (69% for A/S and 60% for A/C). The analyzer was able to separate and quantify HbA2 fraction, allowing ß-thalassemia accidental detection. Conclusion: Capillarys 3 Octa® based on capillary electrophoresis proved to be precise and a linear instrument for HbA1c measurement. Several clinical interferences and Hb variants had no effect on the results. The results of this evaluation suggest that this analyzer is suitable for routine use in clinical chemistry laboratories.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
25
审稿时长
26 weeks
期刊最新文献
Vildagliptin-Associated Bullous Pemphigoid: A Case Report A Comparison of Renal Effects between Empagliflozin and Linagliptin in Diabetic Patients with Chronic Kidney Disease: A Randomized Clinical Trial Genetic and Epigenetic Etiologies of Type 1 Diabetes Mellitus Predictors of Mortality among Intensive Care Unit (ICU) Hospitalized Diabetic Patients with COVID-19 Antibody Engineering to Enhancement of Ranibizumab Binding Affinity for the Prevention and Treatment of Diabetic Retinopathy
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1