无创通气回路中防水装置对人机同步的影响

Hu Jieying, Zheng Ze-guang, Yang Feng, Zhong Lihong, Chen Rongchang
{"title":"无创通气回路中防水装置对人机同步的影响","authors":"Hu Jieying, Zheng Ze-guang, Yang Feng, Zhong Lihong, Chen Rongchang","doi":"10.14302/ISSN.2642-9241.JRD-17-1683","DOIUrl":null,"url":null,"abstract":"Objective: Investigate the effect of connecting a waterproof device at the front end of the piezometric tube on pressure transmission and patient-machine synchronization during the noninvasive ventilation. Method: In test 1, the waterproof device was connected to the piezometric tube and put into a closed container, the pressure inside the container was varied to observe the corresponding pressure change in the piezometric tube. In test 2, a waterproof device was connected in front of the piezometric tube during noninvasive ventilation.12 subjects were received noninvasive ventilator so that dynamic changes of the pressure inside the mask (Pmask) and piezometric tube (Ptube) could be measured. Results: In test 1, when the pressure in the container was gradually increased to 50 cmH2O and then decreased to 0, the pressure inside the piezometric tube changed synchronously with the pressure inside the container, with no statistically significant difference between the pressures (0.009 ± 0.138) cmH2O. In test 2, there was no significant increase in triggering time, pressure, and power after connecting the waterproof device at the front end of the piezometric tube. There was no significant difference in the platform pressure and baseline pressure as measured by Pmask, before and after connecting the waterproof device. Finally, there was no significant difference in the platform pressure and baseline pressure between Pmask and Ptube after connecting the waterproof device. Conclusion: Connecting the waterproof device at the front end of the piezometric tube can prevent condensate from entering the tube without affecting pressure transmission . DOI : COMING SOON Corresponding author: Zheng Zeguang , First Affiliated Hospital of Guangzhou Medical University (State Key Laboratory of Respiratory Disease), Guangzhou Institute of Respiratory Disease, Guangzhou, Guangdong 510120, China","PeriodicalId":83134,"journal":{"name":"The Journal of respiratory diseases","volume":"1 1","pages":"1-7"},"PeriodicalIF":0.0000,"publicationDate":"2017-09-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Effect of A Waterproof Device in the Noninvasive Ventilation Circuit On Patient-machine Synchronization\",\"authors\":\"Hu Jieying, Zheng Ze-guang, Yang Feng, Zhong Lihong, Chen Rongchang\",\"doi\":\"10.14302/ISSN.2642-9241.JRD-17-1683\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: Investigate the effect of connecting a waterproof device at the front end of the piezometric tube on pressure transmission and patient-machine synchronization during the noninvasive ventilation. Method: In test 1, the waterproof device was connected to the piezometric tube and put into a closed container, the pressure inside the container was varied to observe the corresponding pressure change in the piezometric tube. In test 2, a waterproof device was connected in front of the piezometric tube during noninvasive ventilation.12 subjects were received noninvasive ventilator so that dynamic changes of the pressure inside the mask (Pmask) and piezometric tube (Ptube) could be measured. Results: In test 1, when the pressure in the container was gradually increased to 50 cmH2O and then decreased to 0, the pressure inside the piezometric tube changed synchronously with the pressure inside the container, with no statistically significant difference between the pressures (0.009 ± 0.138) cmH2O. In test 2, there was no significant increase in triggering time, pressure, and power after connecting the waterproof device at the front end of the piezometric tube. There was no significant difference in the platform pressure and baseline pressure as measured by Pmask, before and after connecting the waterproof device. Finally, there was no significant difference in the platform pressure and baseline pressure between Pmask and Ptube after connecting the waterproof device. Conclusion: Connecting the waterproof device at the front end of the piezometric tube can prevent condensate from entering the tube without affecting pressure transmission . DOI : COMING SOON Corresponding author: Zheng Zeguang , First Affiliated Hospital of Guangzhou Medical University (State Key Laboratory of Respiratory Disease), Guangzhou Institute of Respiratory Disease, Guangzhou, Guangdong 510120, China\",\"PeriodicalId\":83134,\"journal\":{\"name\":\"The Journal of respiratory diseases\",\"volume\":\"1 1\",\"pages\":\"1-7\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2017-09-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The Journal of respiratory diseases\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.14302/ISSN.2642-9241.JRD-17-1683\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The Journal of respiratory diseases","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14302/ISSN.2642-9241.JRD-17-1683","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:探讨无创通气过程中在测压管前端连接防水装置对压力传递和病人-机器同步的影响。方法:试验一将防水装置与测压管连接,放入密闭容器中,改变容器内压力,观察测压管内相应压力的变化。在试验2中,无创通气时在测压管前连接防水装置。12例患者接受无创呼吸机,测量面罩(Pmask)和测压管(Ptube)内压力的动态变化。结果:在试验1中,当容器内压力逐渐增加到50 cmH2O,再降低到0时,测压管内压力与容器内压力同步变化,两者之间的差异无统计学意义(0.009±0.138)cmH2O。在试验2中,在测压管前端连接防水装置后,触发时间、压力、功率均无明显增加。连接防水装置前后,Pmask测得的平台压力与基线压力无显著差异。最后,连接防水装置后,Pmask与Ptube的平台压力和基线压力无显著差异。结论:在测压管前端连接防水装置,既能防止冷凝水进入管内,又不影响压力传递。通讯作者:郑泽光,广州医科大学第一附属医院(呼吸疾病国家重点实验室),广州呼吸疾病研究所,广东广州510120
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Effect of A Waterproof Device in the Noninvasive Ventilation Circuit On Patient-machine Synchronization
Objective: Investigate the effect of connecting a waterproof device at the front end of the piezometric tube on pressure transmission and patient-machine synchronization during the noninvasive ventilation. Method: In test 1, the waterproof device was connected to the piezometric tube and put into a closed container, the pressure inside the container was varied to observe the corresponding pressure change in the piezometric tube. In test 2, a waterproof device was connected in front of the piezometric tube during noninvasive ventilation.12 subjects were received noninvasive ventilator so that dynamic changes of the pressure inside the mask (Pmask) and piezometric tube (Ptube) could be measured. Results: In test 1, when the pressure in the container was gradually increased to 50 cmH2O and then decreased to 0, the pressure inside the piezometric tube changed synchronously with the pressure inside the container, with no statistically significant difference between the pressures (0.009 ± 0.138) cmH2O. In test 2, there was no significant increase in triggering time, pressure, and power after connecting the waterproof device at the front end of the piezometric tube. There was no significant difference in the platform pressure and baseline pressure as measured by Pmask, before and after connecting the waterproof device. Finally, there was no significant difference in the platform pressure and baseline pressure between Pmask and Ptube after connecting the waterproof device. Conclusion: Connecting the waterproof device at the front end of the piezometric tube can prevent condensate from entering the tube without affecting pressure transmission . DOI : COMING SOON Corresponding author: Zheng Zeguang , First Affiliated Hospital of Guangzhou Medical University (State Key Laboratory of Respiratory Disease), Guangzhou Institute of Respiratory Disease, Guangzhou, Guangdong 510120, China
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Perceived Improvements of Quality of Life (QoL) among Patients with Idiopathic Pulmonary Fibrosis (IPF) in Response to a 6-Week Rehabilitation Program Mature Teratoma Revealed by an Encysted Pleural Effusion Patients’ Experiences and Opinions on Pulmonary Rehabilitation and Use of It as a Tool of Palliative Care on Idiopathic Pulmonary Fibrosis (IPF) Unusual Cause of Stridor in an Adult Man, Selective Immunoglobulin A Deficiency Portuguese Society of Intensive Care Score for Predicting SARS-CoV-2 Infection Applied to Inpatients with Pneumonia: A Reliable Tool?
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1