宫颈癌近距离放射治疗的剂量传递方法

D. Kazlouski, A. A. Suslava, I. Dubik, Y. Kazlouskaya, V. Suslava, N. Artemova, I. H. Tarutin
{"title":"宫颈癌近距离放射治疗的剂量传递方法","authors":"D. Kazlouski, A. A. Suslava, I. Dubik, Y. Kazlouskaya, V. Suslava, N. Artemova, I. H. Tarutin","doi":"10.35596/1729-7648-2023-21-1-58-65","DOIUrl":null,"url":null,"abstract":"A method is proposed for delivering an absorbed dose to a target during brachytherapy for cervical cancer by using a new applicator, which consists of an annular part and two uterine canals. The design of the applicator makes it possible to implement it to the required depth, taking into account the anatomical features, to widely cover the cervix and corporis of a womb with a 100 % isodose of irradiation, which makes it possible to deliver the planned dose. The patient admitted to the brachytherapy department received the first treatment fraction using the Interstitial Ring applicator, the second and subsequent fractions were performed using the proposed applicator (6 fractions in total). In comparison with the standard applicator Interstitial Ring, the use of the proposed applicator made it possible to increase the coverage of the target with the prescribed dose from 69.2 to 95.5 %. This led to an increase in the absorbed dose per target for a course of radiation therapy from the planned 70.4 to 85.4 Gy. Twelve treatment plans for 3 patients were analyzed using different dose fractionation schemes. In all three cases, it was possible to increase the coverage of the target with the prescribed dose, which led to an increase in the total absorbed dose per target, which was more than 85 Gy. The proposed method has been successfully applied in the N. N. Alexandrov National Cancer Centre during brachytherapy for cancer of the cervix and corporis of a womb.","PeriodicalId":33565,"journal":{"name":"Doklady Belorusskogo gosudarstvennogo universiteta informatiki i radioelektroniki","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-03-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Method of Dose Delivery in Brachytherapy for Cervical Cancer\",\"authors\":\"D. Kazlouski, A. A. Suslava, I. Dubik, Y. Kazlouskaya, V. Suslava, N. Artemova, I. H. Tarutin\",\"doi\":\"10.35596/1729-7648-2023-21-1-58-65\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"A method is proposed for delivering an absorbed dose to a target during brachytherapy for cervical cancer by using a new applicator, which consists of an annular part and two uterine canals. The design of the applicator makes it possible to implement it to the required depth, taking into account the anatomical features, to widely cover the cervix and corporis of a womb with a 100 % isodose of irradiation, which makes it possible to deliver the planned dose. The patient admitted to the brachytherapy department received the first treatment fraction using the Interstitial Ring applicator, the second and subsequent fractions were performed using the proposed applicator (6 fractions in total). In comparison with the standard applicator Interstitial Ring, the use of the proposed applicator made it possible to increase the coverage of the target with the prescribed dose from 69.2 to 95.5 %. This led to an increase in the absorbed dose per target for a course of radiation therapy from the planned 70.4 to 85.4 Gy. Twelve treatment plans for 3 patients were analyzed using different dose fractionation schemes. In all three cases, it was possible to increase the coverage of the target with the prescribed dose, which led to an increase in the total absorbed dose per target, which was more than 85 Gy. The proposed method has been successfully applied in the N. N. Alexandrov National Cancer Centre during brachytherapy for cancer of the cervix and corporis of a womb.\",\"PeriodicalId\":33565,\"journal\":{\"name\":\"Doklady Belorusskogo gosudarstvennogo universiteta informatiki i radioelektroniki\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-03-02\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Doklady Belorusskogo gosudarstvennogo universiteta informatiki i radioelektroniki\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.35596/1729-7648-2023-21-1-58-65\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Doklady Belorusskogo gosudarstvennogo universiteta informatiki i radioelektroniki","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.35596/1729-7648-2023-21-1-58-65","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

本文提出了一种在宫颈癌近距离治疗过程中,通过使用一种由环形部分和两个子宫管组成的新型涂抹器,将吸收剂量输送到靶点的方法。应用器的设计使其能够实现所需的深度,考虑到解剖特征,以100%等剂量的辐射广泛覆盖子宫的子宫颈和子宫,从而可以提供计划剂量。入住近距离放疗科的患者使用间质环涂抹器接受第一次治疗分数,第二次及后续分数使用建议的涂抹器进行(总共6个分数)。与标准施药器间隙环相比,使用建议的施药器可以将规定剂量的目标覆盖率从69.2提高到95.5%。这导致一个放射治疗过程中每个靶的吸收剂量从计划的70.4戈瑞增加到85.4戈瑞。分析3例患者的12种治疗方案,采用不同的剂量分割方案。在所有三种情况下,都有可能用规定剂量增加靶物的覆盖范围,这导致每个靶物的总吸收剂量增加,超过85戈瑞。所提出的方法已成功地应用于N. N.亚历山德罗夫国家癌症中心在近距离治疗宫颈癌和子宫癌。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
Method of Dose Delivery in Brachytherapy for Cervical Cancer
A method is proposed for delivering an absorbed dose to a target during brachytherapy for cervical cancer by using a new applicator, which consists of an annular part and two uterine canals. The design of the applicator makes it possible to implement it to the required depth, taking into account the anatomical features, to widely cover the cervix and corporis of a womb with a 100 % isodose of irradiation, which makes it possible to deliver the planned dose. The patient admitted to the brachytherapy department received the first treatment fraction using the Interstitial Ring applicator, the second and subsequent fractions were performed using the proposed applicator (6 fractions in total). In comparison with the standard applicator Interstitial Ring, the use of the proposed applicator made it possible to increase the coverage of the target with the prescribed dose from 69.2 to 95.5 %. This led to an increase in the absorbed dose per target for a course of radiation therapy from the planned 70.4 to 85.4 Gy. Twelve treatment plans for 3 patients were analyzed using different dose fractionation schemes. In all three cases, it was possible to increase the coverage of the target with the prescribed dose, which led to an increase in the total absorbed dose per target, which was more than 85 Gy. The proposed method has been successfully applied in the N. N. Alexandrov National Cancer Centre during brachytherapy for cancer of the cervix and corporis of a womb.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
87
审稿时长
8 weeks
期刊最新文献
Recognition of Aerodynamic Objects on Spectral Portraits Taking into Account Design Features of Turbojets Skeleting of Low-Contrast Noisy Halftone Images Electrically Tunable Four-Mirror Gyrotron with Crossed Fields Assessment of the Contribution of Radiations of User Equipment to the Anthropogenic Electromagnetic Background Created by Mobile (Cellular) Communications Ontological Representation of Business Processes in an Educational Institution
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1