Eileen Brobbin, Paolo Deluca, Simon Coulton, Colin Drummond
{"title":"实验室环境中透皮酒精监测装置的准确性。","authors":"Eileen Brobbin, Paolo Deluca, Simon Coulton, Colin Drummond","doi":"10.1093/alcalc/agad068","DOIUrl":null,"url":null,"abstract":"<p><p>The development of transdermal alcohol sensors (TASs) presents a new method to monitor alcohol consumption with the ability to objectively measure data 24/7. We aimed to evaluate the accuracy of two TASs (BACtrack Skyn and Smart Start BARE) in a laboratory setting. Thirty-two adults received a dose of ethanol 0.56 g/kg body weight as a 20% solution while wearing the two TASs and provided Breath Alcohol Concentration (BrAC) measurements for 3.5 h postalcohol consumption. Pearson's correlations and repeated measures analysis of variance tests were conducted on the peak, time-to-peak, and area under the curve data. Bland-Altman plots were derived. A time series analysis and cross-correlations were conducted to adjust for time lag. Both TASs were able to detect alcohol and increase within 20 min. BrAC peaked significantly quicker than Skyn and BARE. BrAC and Skyn peaks were negatively significantly correlated (r = -0.381, P = .035, n = 31), while Skyn and BARE peaks were positively significantly correlated (r = 0.380, P = .038, n = 30). Repeated measures analysis of variance found a significant difference between BrAC, Skyn, and BARE (F(1.946, 852.301) = 459.873, P < .001)). A time series analysis found when BrAC-Skyn and BrAC-BARE were adjusted for the delay to peak, and there was still a significant difference. Failure rates: 1.7% (Skyn) and 4.8% (BARE). Some evidence was obtained for TAS validity as both consistently detected alcohol. Failure rates and time lag show improvements in older device generations. However, neither TAS presented strong equivalence to the breathalyser even when the lag time was adjusted. With further testing and technology advancements, TAS could be a potential alcohol monitoring tool. Two of the newest TAS devices were worn in laboratory conditions for one afternoon to compare their accuracy of alcohol monitoring to a breathalyser. Findings suggest that the two TASs (BACtrack Skyn and SmartStart BARE) recorded significantly similar data postalcohol consumption, but not with the breathalyser.</p>","PeriodicalId":7407,"journal":{"name":"Alcohol and alcoholism","volume":null,"pages":null},"PeriodicalIF":2.1000,"publicationDate":"2024-01-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Accuracy of transdermal alcohol monitoring devices in a laboratory setting.\",\"authors\":\"Eileen Brobbin, Paolo Deluca, Simon Coulton, Colin Drummond\",\"doi\":\"10.1093/alcalc/agad068\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The development of transdermal alcohol sensors (TASs) presents a new method to monitor alcohol consumption with the ability to objectively measure data 24/7. We aimed to evaluate the accuracy of two TASs (BACtrack Skyn and Smart Start BARE) in a laboratory setting. Thirty-two adults received a dose of ethanol 0.56 g/kg body weight as a 20% solution while wearing the two TASs and provided Breath Alcohol Concentration (BrAC) measurements for 3.5 h postalcohol consumption. Pearson's correlations and repeated measures analysis of variance tests were conducted on the peak, time-to-peak, and area under the curve data. Bland-Altman plots were derived. A time series analysis and cross-correlations were conducted to adjust for time lag. Both TASs were able to detect alcohol and increase within 20 min. BrAC peaked significantly quicker than Skyn and BARE. BrAC and Skyn peaks were negatively significantly correlated (r = -0.381, P = .035, n = 31), while Skyn and BARE peaks were positively significantly correlated (r = 0.380, P = .038, n = 30). Repeated measures analysis of variance found a significant difference between BrAC, Skyn, and BARE (F(1.946, 852.301) = 459.873, P < .001)). A time series analysis found when BrAC-Skyn and BrAC-BARE were adjusted for the delay to peak, and there was still a significant difference. Failure rates: 1.7% (Skyn) and 4.8% (BARE). Some evidence was obtained for TAS validity as both consistently detected alcohol. Failure rates and time lag show improvements in older device generations. However, neither TAS presented strong equivalence to the breathalyser even when the lag time was adjusted. With further testing and technology advancements, TAS could be a potential alcohol monitoring tool. Two of the newest TAS devices were worn in laboratory conditions for one afternoon to compare their accuracy of alcohol monitoring to a breathalyser. Findings suggest that the two TASs (BACtrack Skyn and SmartStart BARE) recorded significantly similar data postalcohol consumption, but not with the breathalyser.</p>\",\"PeriodicalId\":7407,\"journal\":{\"name\":\"Alcohol and alcoholism\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2024-01-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Alcohol and alcoholism\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1093/alcalc/agad068\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"SUBSTANCE ABUSE\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Alcohol and alcoholism","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1093/alcalc/agad068","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"SUBSTANCE ABUSE","Score":null,"Total":0}
Accuracy of transdermal alcohol monitoring devices in a laboratory setting.
The development of transdermal alcohol sensors (TASs) presents a new method to monitor alcohol consumption with the ability to objectively measure data 24/7. We aimed to evaluate the accuracy of two TASs (BACtrack Skyn and Smart Start BARE) in a laboratory setting. Thirty-two adults received a dose of ethanol 0.56 g/kg body weight as a 20% solution while wearing the two TASs and provided Breath Alcohol Concentration (BrAC) measurements for 3.5 h postalcohol consumption. Pearson's correlations and repeated measures analysis of variance tests were conducted on the peak, time-to-peak, and area under the curve data. Bland-Altman plots were derived. A time series analysis and cross-correlations were conducted to adjust for time lag. Both TASs were able to detect alcohol and increase within 20 min. BrAC peaked significantly quicker than Skyn and BARE. BrAC and Skyn peaks were negatively significantly correlated (r = -0.381, P = .035, n = 31), while Skyn and BARE peaks were positively significantly correlated (r = 0.380, P = .038, n = 30). Repeated measures analysis of variance found a significant difference between BrAC, Skyn, and BARE (F(1.946, 852.301) = 459.873, P < .001)). A time series analysis found when BrAC-Skyn and BrAC-BARE were adjusted for the delay to peak, and there was still a significant difference. Failure rates: 1.7% (Skyn) and 4.8% (BARE). Some evidence was obtained for TAS validity as both consistently detected alcohol. Failure rates and time lag show improvements in older device generations. However, neither TAS presented strong equivalence to the breathalyser even when the lag time was adjusted. With further testing and technology advancements, TAS could be a potential alcohol monitoring tool. Two of the newest TAS devices were worn in laboratory conditions for one afternoon to compare their accuracy of alcohol monitoring to a breathalyser. Findings suggest that the two TASs (BACtrack Skyn and SmartStart BARE) recorded significantly similar data postalcohol consumption, but not with the breathalyser.
期刊介绍:
About the Journal
Alcohol and Alcoholism publishes papers on the biomedical, psychological, and sociological aspects of alcoholism and alcohol research, provided that they make a new and significant contribution to knowledge in the field.
Papers include new results obtained experimentally, descriptions of new experimental (including clinical) methods of importance to the field of alcohol research and treatment, or new interpretations of existing results.
Theoretical contributions are considered equally with papers dealing with experimental work provided that such theoretical contributions are not of a largely speculative or philosophical nature.