{"title":"影响部分定制基台保留效果的因素的比较评估,该基台专为在美容区应用单植入物而设计。","authors":"Jan C Kwan, Norman H Kwan","doi":"10.11607/jomi.10598","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To compare the effect of taper, length, angle, and number of vertical axial walls on the retentive strength of a cemented crown on a partially customized hybrid abutment in the esthetic zone.</p><p><strong>Materials and methods: </strong>A total of 35 metal copings were used and divided into two groups. One group had 30 copings cemented to their corresponding 8-degree tapered abutment with lengths from 3 to 8 mm, increasing in 1-mm increments (5 copings per length). The other group had the remaining 5 copings and consisted of a standardized metal coping that matched a hexagonal abutment with 3-mm vertical axial walls and the sequential removal of 1, 2, and 3 contiguous vertical axial walls. Dislodgment tests were performed for all copings in both groups. Maximum retentive forces were measured in retentive strength (kgF), with a conversion factor of 9.807 N &#61; 1 kgF.</p><p><strong>Results: </strong>At each tapered abutment length, the retentive strength increased proportionally and was significantly different, ranging from 31.67 ± 4.10 kgF to 67.68 ± 11.22 kgF (F [5,24] &#61; 20.46, P < .001). An unmodified hexagonal abutment demonstrated the highest retentive strength (70.15 ± 12.97 kgF). The sequential removal of 1, 2, and 3 contiguous vertical axial walls of the hexagonal abutment resulted in retentive strength values of 59.89 ± 10.06 kgF, 57.01 ± 9.62 kgF, and 55.99 ± 9.35 kgF, respectively, with no significant difference (P > .05) in strength.</p><p><strong>Conclusions: </strong>A partially customized abutment with vertical axial walls on one side and a profile reduction on the opposite side can provide comparable retention to cemented copings at one-third the length and at one-sixth the surface area of an 8-mm abutment with an 8-degree taper.</p>","PeriodicalId":94230,"journal":{"name":"The International journal of oral & maxillofacial implants","volume":"0 0","pages":"575-584"},"PeriodicalIF":0.0000,"publicationDate":"2024-08-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A Comparative Evaluation of Factors That Affect the Retentive Efficacy of a Partially Customized Abutment Specifically Designed for Single Implant Application in the Esthetic Zone.\",\"authors\":\"Jan C Kwan, Norman H Kwan\",\"doi\":\"10.11607/jomi.10598\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>To compare the effect of taper, length, angle, and number of vertical axial walls on the retentive strength of a cemented crown on a partially customized hybrid abutment in the esthetic zone.</p><p><strong>Materials and methods: </strong>A total of 35 metal copings were used and divided into two groups. One group had 30 copings cemented to their corresponding 8-degree tapered abutment with lengths from 3 to 8 mm, increasing in 1-mm increments (5 copings per length). The other group had the remaining 5 copings and consisted of a standardized metal coping that matched a hexagonal abutment with 3-mm vertical axial walls and the sequential removal of 1, 2, and 3 contiguous vertical axial walls. Dislodgment tests were performed for all copings in both groups. Maximum retentive forces were measured in retentive strength (kgF), with a conversion factor of 9.807 N &#61; 1 kgF.</p><p><strong>Results: </strong>At each tapered abutment length, the retentive strength increased proportionally and was significantly different, ranging from 31.67 ± 4.10 kgF to 67.68 ± 11.22 kgF (F [5,24] &#61; 20.46, P < .001). An unmodified hexagonal abutment demonstrated the highest retentive strength (70.15 ± 12.97 kgF). The sequential removal of 1, 2, and 3 contiguous vertical axial walls of the hexagonal abutment resulted in retentive strength values of 59.89 ± 10.06 kgF, 57.01 ± 9.62 kgF, and 55.99 ± 9.35 kgF, respectively, with no significant difference (P > .05) in strength.</p><p><strong>Conclusions: </strong>A partially customized abutment with vertical axial walls on one side and a profile reduction on the opposite side can provide comparable retention to cemented copings at one-third the length and at one-sixth the surface area of an 8-mm abutment with an 8-degree taper.</p>\",\"PeriodicalId\":94230,\"journal\":{\"name\":\"The International journal of oral & maxillofacial implants\",\"volume\":\"0 0\",\"pages\":\"575-584\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-08-29\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"The International journal of oral & maxillofacial implants\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.11607/jomi.10598\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"The International journal of oral & maxillofacial implants","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11607/jomi.10598","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究的目的是比较锥形、长度、角度和垂直轴向壁的数量等因素对美学区部分定制混合基牙上胶合冠的保持强度的影响。材料和方法:本研究共使用了35个金属牙冠,分为一组,其中30个牙冠与相应的8°锥形基牙粘接,长度从3到8 mm,增量为1 mm。其余五个由标准化金属顶盖组成,该顶盖与具有3mm垂直轴向壁的六边形桥台相匹配,并依次移除1、2和3个连续的垂直轴向壁。两组的所有复制品均进行了丢弃测试。最大保持力以kgF为单位,转换系数为9.807N至1kgF。结果:在每个锥形基台长度处,保持力成比例增加,从31.67±SD 4.10kgF到67.68±SD 11.22kgF有显著差异,未经修饰的六边形基牙的保持力最高,为70.15±SD 12.97kgF。连续去除六边形基牙的1、2和3个连续垂直轴向壁分别为59.89±SD 10.06 kgF、57.01±SD 9.62 kgF和55.99±SD 9.35 kgF,保持强度无显著差异(p>0.05)。结论:一侧为垂直轴壁,另一侧为轮廓缩小的部分定制基牙,在8°锥度的8 mm高基牙的1/3长度和1/6表面积下,可以提供相当的水泥盖固位。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
查看原文
分享 分享
微信好友 朋友圈 QQ好友 复制链接
本刊更多论文
A Comparative Evaluation of Factors That Affect the Retentive Efficacy of a Partially Customized Abutment Specifically Designed for Single Implant Application in the Esthetic Zone.

Purpose: To compare the effect of taper, length, angle, and number of vertical axial walls on the retentive strength of a cemented crown on a partially customized hybrid abutment in the esthetic zone.

Materials and methods: A total of 35 metal copings were used and divided into two groups. One group had 30 copings cemented to their corresponding 8-degree tapered abutment with lengths from 3 to 8 mm, increasing in 1-mm increments (5 copings per length). The other group had the remaining 5 copings and consisted of a standardized metal coping that matched a hexagonal abutment with 3-mm vertical axial walls and the sequential removal of 1, 2, and 3 contiguous vertical axial walls. Dislodgment tests were performed for all copings in both groups. Maximum retentive forces were measured in retentive strength (kgF), with a conversion factor of 9.807 N = 1 kgF.

Results: At each tapered abutment length, the retentive strength increased proportionally and was significantly different, ranging from 31.67 ± 4.10 kgF to 67.68 ± 11.22 kgF (F [5,24] = 20.46, P < .001). An unmodified hexagonal abutment demonstrated the highest retentive strength (70.15 ± 12.97 kgF). The sequential removal of 1, 2, and 3 contiguous vertical axial walls of the hexagonal abutment resulted in retentive strength values of 59.89 ± 10.06 kgF, 57.01 ± 9.62 kgF, and 55.99 ± 9.35 kgF, respectively, with no significant difference (P > .05) in strength.

Conclusions: A partially customized abutment with vertical axial walls on one side and a profile reduction on the opposite side can provide comparable retention to cemented copings at one-third the length and at one-sixth the surface area of an 8-mm abutment with an 8-degree taper.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
期刊最新文献
Clinical and Radiographic Outcome Short-span Fixed Partial Dentures Supported by 2 Immediately Placed Implants in the Anterior Mandible: A Long-term Retrospective Analysis. Comparison of Three-Dimensional Accuracy of Conventional Cast Metal to Milled Zirconia Implant Full-Arch Fixed Dental Prosthesis Frameworks. Reconstructive Peri-implant Site Development throughout an Implant Life Cycle. A Retrospective Study. Strain Assessment Around Two Short Implants Retaining Mandibular Overdenture with Different Attachment Systems: In- Vitro Study. The Effect of Basis Images and a Metal Artifact Reduction Algorithm on Artifact Expression and Trabecular Bone Architecture in the Vicinity of a Zirconium Dental Implant Placed in Different Mandibular Regions: An Ex Vivo Study.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
现在去查看 取消
×
提示
确定
0
微信
客服QQ
Book学术公众号 扫码关注我们
反馈
×
意见反馈
请填写您的意见或建议
请填写您的手机或邮箱
已复制链接
已复制链接
快去分享给好友吧!
我知道了
×
扫码分享
扫码分享
Book学术官方微信
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术
文献互助 智能选刊 最新文献 互助须知 联系我们:info@booksci.cn
Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。
Copyright © 2023 Book学术 All rights reserved.
ghs 京公网安备 11010802042870号 京ICP备2023020795号-1